Meta-analysis of methods used to assess implant stability.

Murat Cavit Cehreli1, Durdu Karasoy, Kivanc Akca, Steven E Eckert.   

Abstract

PURPOSE: To evaluate correlations between Periotest, cutting torque or insertion torque, and/or the reverse torque test with resonance frequency analysis measurements.
MATERIALS AND METHODS: The combined search terms dental implant, primary stability, resonance frequency analysis, removal torque, vibration analysis, torque, Periotest, cutting torque, and insertion torque and specific inclusion and exclusion criteria were used to retrieve eligible articles published between 1998 and 2008 (up to April 1) from PubMed, EMBASE, OVID, The Cochrane Library databases, and 10 journals by hand-searching. The pooled P value was calculated with the Fisher method. The correlation coefficients (r) were combined using fixed and random effect models. Heterogeneity of the studies was tested by performing the homogeneity test.
RESULTS: Fifty articles were identified, but 3 studies were excluded because of limitations in study design. Studies that fulfilled the inclusion criteria fell into a variety of categories (11 human cadaver, 16 clinical, 15 animal, and 5 in vitro studies). Assessment of P values revealed that correlation between cutting torque or insertion torque and resonance frequency analysis was significant (6 studies; P = .0022). Correlation between Periotest and cutting torque or insertion torque was significant (1 study; P = .015), and correlations between resonance frequency analysis and reverse torque test values (1 study; P = .319; r = 0.405) and Periotest and resonance frequency analysis (P = .28) were insignificant. Nine articles provided r values, and cutting torque or insertion torque and resonance frequency analysis showed a statistically significant and direct relationship of 55.4% (P = .000). One study provided an r value of -0.149. A statistically significant and direct relationship of 87.6% was found between cutting torque or insertion torque and reverse torque (2 studies; P = .000).
CONCLUSION: This meta-analysis of 47 studies showed a statistically significant correlation between cutting torque or insertion torque and resonance frequency analysis.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20162105

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  14 in total

1.  Evaluation of dental implant insertion torque using a manual ratchet.

Authors:  M M Goswami; Mukul Kumar; Abhinav Vats; A S Bansal
Journal:  Med J Armed Forces India       Date:  2013-09-24

2.  Assessment of implant stability following sinus lift procedures with different grafting materials.

Authors:  Damir Jelušić; Ivan Puhar; Darije Plančak
Journal:  Acta Stomatol Croat       Date:  2014-03

3.  A clinico radiographic study of immediate loading implants in rehabilitation of mandibular ridges.

Authors:  M Viswambaran; Vimal Arora; S H Gupta; R K Dhiman; N Thiruvalluvan
Journal:  Med J Armed Forces India       Date:  2014-08-22

4.  Is trabecular bone related to primary stability of miniscrews?

Authors:  Mariana Marquezan; Inayá Lima; Ricardo Tadeu Lopes; Eduardo Franzotti Sant'Anna; Margareth Maria Gomes de Souza
Journal:  Angle Orthod       Date:  2013-11-18       Impact factor: 2.079

5.  Analysis of primary stability of dental implants inserted in different substrates using the pullout test and insertion torque.

Authors:  Nathalia Ferraz Oliscovicz; Antônio Carlos Shimano; Elcio Marcantonio Junior; César Penazzo Lepri; Andréa Candido Dos Reis
Journal:  Int J Dent       Date:  2013-01-22

6.  "Reverse torque of 30 Ncm applied to dental implants as test for osseointegration"-a human observational study.

Authors:  Sabrina G Simeone; María Rios; Jeannette Simonpietri
Journal:  Int J Implant Dent       Date:  2016-12-07

7.  Relation between the stability of dental implants and two biological markers during the healing period: a prospective clinical study.

Authors:  Choknapa Tirachaimongkol; Peraphan Pothacharoen; Peter A Reichart; Pathawee Khongkhunthian
Journal:  Int J Implant Dent       Date:  2016-12-08

8.  Implant Stability of Biological Hydroxyapatites Used in Dentistry.

Authors:  Maria Piedad Ramírez Fernández; Sergio A Gehrke; Patricia Mazón; Jose L Calvo-Guirado; Piedad N De Aza
Journal:  Materials (Basel)       Date:  2017-06-12       Impact factor: 3.623

9.  Effect of implant macro-design on primary stability: A prospective clinical study.

Authors:  Naroa Lozano-Carrascal; Oscar Salomó-Coll; Marta Gilabert-Cerdà; Nuria Farré-Pagés; Jordi Gargallo-Albiol; Federico Hernández-Alfaro
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2016-03-01

Review 10.  Does cortical thickness influence the primary stability of miniscrews?: A systematic review and meta-analysis.

Authors:  Mariana Marquezan; Cláudia Trindade Mattos; Eduardo Franzotti Sant'Anna; Margareth Maria Gomes de Souza; Lucianne Cople Maia
Journal:  Angle Orthod       Date:  2014-04-02       Impact factor: 2.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.