Measurements comparing the initial stability of five designs of dental implants: a human cadaver study.

D O'Sullivan1, L Sennerby, N Meredith.   

Abstract

BACKGROUND: A number of different dental implant designs are currently in clinical use. A successful outcome of implant placement is thought, at least in part, to be due to the primary stability of an implant after placement. Few data are available for comparing the primary stability characteristics of different implant designs.
PURPOSE: This investigation compared the primary stability of five types of endosseous dental implant of varying geometry and surface topography.
MATERIALS AND METHODS: Comparison was made between a standard threaded commercially pure titanium implant (Nobel Biocare AB, Göteborg, Sweden), the Mark II self-tapping implant (Nobel Biocare AB, Göteborg, Sweden), the Mark IV tapered self-tapping implant (Nobel Biocare AB, Göteborg, Sweden), the Astra Tioblast (AstraTech AB, Mölndahl, Sweden), and the 3i Osseotite (3I [Implant Innovations Incorporated], Palm Beach, Florida, USA). Fifty-two fixtures were placed in the maxillary bone of nine unembalmed human cadavers. Implant stability as a function of peak insertion torque and resonance frequency values was recorded for each fixture site after placement. Removal torque was also measured 1-hour postinsertion. Assessment of bone quality at each site was made.
RESULTS: All of the implants tested demonstrated good primary stability in type 2 and 3 bone. The Standard, Mark II, Osseotite, and Tioblast were less stable when placed into bone type 4. The Mark IV implants appeared to maintain a high primary stability even in Type 4 bone.
CONCLUSION: When looking across all bone qualities, the Mark IV implant develops a significantly higher insertion torque than the Standard, Mark II, and Osseotite implant types, and a significantly higher resonance frequency value than the Standard implant, indicating a higher interfacial stiffness at the implant-bone interface.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11359268     DOI: 10.1111/j.1708-8208.2000.tb00110.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  44 in total

1.  The effect of implant shape and bone preparation on primary stability.

Authors:  Sang-Hyun Moon; Heung-Sik Um; Jae-Kwan Lee; Beom-Seok Chang; Min-Ku Lee
Journal:  J Periodontal Implant Sci       Date:  2010-10-31       Impact factor: 2.614

2.  Evaluation of dental implant insertion torque using a manual ratchet.

Authors:  M M Goswami; Mukul Kumar; Abhinav Vats; A S Bansal
Journal:  Med J Armed Forces India       Date:  2013-09-24

3.  [Systematic qualitative histology of enossal implants with anodically oxidised surfaces].

Authors:  B Al-Nawas; K A Grötz; H Goetz; M Feil; H Duschner; W Wagner
Journal:  Mund Kiefer Gesichtschir       Date:  2006-07

4.  Clinical study on the primary stability of two dental implant systems with resonance frequency analysis.

Authors:  Annette Rabel; Steffen Gerhard Köhler; Andrea Maria Schmidt-Westhausen
Journal:  Clin Oral Investig       Date:  2007-03-31       Impact factor: 3.573

5.  Minimally invasive transmucosal insertion and immediate provisonalization of one-piece implant in partially edentulous posterior mandible.

Authors:  Narpatsingh Rajput; Siyad K P; Rathinavelu G; Chandrasekaran S C; Jumshad Mohammed
Journal:  J Clin Diagn Res       Date:  2013-09-10

6.  Implant stability in different techniques of surgical sites preparation: an in vitro study.

Authors:  C Rastelli; G Falisi; R Gatto; M Galli; E Saccone; M Severino; C DI Paolo
Journal:  Oral Implantol (Rome)       Date:  2014-12-27

7.  Histomorphometry and stability analysis of early loaded implants with two different surface conditions in beagle dogs.

Authors:  Dong-Seok Kim; Dae-Gon Kim; Chan-Jin Park; Lee-Ra Cho
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

8.  The effect of various thread designs on the initial stability of taper implants.

Authors:  Ju-Hee Park; Young-Jun Lim; Myung-Joo Kim; Ho-Beom Kwon
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

9.  Differences in crestal bone-to-implant contact following an under-drilling compared to an over-drilling protocol. A study in the rabbit tibia.

Authors:  Omer Cohen; Zeev Ormianer; Haim Tal; Daniel Rothamel; Miron Weinreb; Ofer Moses
Journal:  Clin Oral Investig       Date:  2016-03-02       Impact factor: 3.573

Review 10.  Evaluation of functional dynamics during osseointegration and regeneration associated with oral implants.

Authors:  Po-Chun Chang; Niklaus P Lang; William V Giannobile
Journal:  Clin Oral Implants Res       Date:  2010-01       Impact factor: 5.977

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.