The effect of implant design on insertion torque and immediate micromotion.

Amilcar C Freitas1, Estevam A Bonfante, Gabriela Giro, Malvin N Janal, Paulo G Coelho.   

Abstract

OBJECTIVES: To evaluate the effect of insertion torque on micromotion to a lateral force in three different implant designs.
MATERIAL AND METHODS: Thirty-six implants with identical thread design, but different cutting groove design were divided in three groups: (1) non-fluted (no cutting groove, solid screw-form); (2) fluted (90° cut at the apex, tap design); and (3) Blossom(™) (Patent pending) (non-fluted with engineered trimmed thread design). The implants were screwed into polyurethane foam blocks and the insertion torque was recorded after each turn of 90° by a digital torque gauge. Controlled lateral loads of 10 N followed by increments of 5 up to 100 N were sequentially applied by a digital force gauge on a titanium abutment. Statistical comparison was performed with two-way mixed model ANOVA that evaluated implant design group, linear effects of turns and displacement loads, and their interaction.
RESULTS: While insertion torque increased as a function of number of turns for each design, the slope and final values increased (P<0.001) progressively from the Blossom to the fluted to the non-fluted design (M ± standard deviation [SD]=64.1 ± 26.8, 139.4 ± 17.2, and 205.23 ± 24.3 Ncm, respectively). While a linear relationship between horizontal displacement and lateral force was observed for each design, the slope and maximal displacement increased (P<0.001) progressively from the Blossom to the fluted to the non-fluted design (M ± SD=530 ± 57.7, 585.9 ± 82.4, and 782.33 ± 269.4 μm, respectively). There was negligible to moderate levels of association between insertion torque and lateral displacement in the Blossom, fluted and non-fluted design groups, respectively.
CONCLUSION: Insertion torque was reduced in implant macrodesigns that incorporated cutting edges, and lesser insertion torque was generally associated with decreased micromovement. However, insertion torque and micromotion were unrelated within implant designs, particularly for those designs showing the least insertion torque.
© 2011 John Wiley & Sons A/S.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21426405     DOI: 10.1111/j.1600-0501.2010.02142.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  8 in total

1.  Evaluation of dental implant insertion torque using a manual ratchet.

Authors:  M M Goswami; Mukul Kumar; Abhinav Vats; A S Bansal
Journal:  Med J Armed Forces India       Date:  2013-09-24

2.  Analysis of the influence of implant shape on primary stability using the correlation of multiple methods.

Authors:  Mariana Lima da Costa Valente; Denise Tornavoi de Castro; Antonio Carlos Shimano; César Penazzo Lepri; Andréa Cândido dos Reis
Journal:  Clin Oral Investig       Date:  2015-02-14       Impact factor: 3.573

3.  In-vitro development of a temporal abutment screw to protect osseointegration in immediate loaded implants.

Authors:  Herminio García-Roncero; Jordi Caballé-Serrano; Jordi Cano-Batalla; Josep Cabratosa-Termes; Oscar Figueras-Álvarez
Journal:  J Adv Prosthodont       Date:  2015-04-23       Impact factor: 1.904

Review 4.  Micromotions and combined damages at the dental implant/bone interface.

Authors:  Shan-Shan Gao; Ya-Rong Zhang; Zhuo-Li Zhu; Hai-Yang Yu
Journal:  Int J Oral Sci       Date:  2012-12-21       Impact factor: 6.344

Review 5.  Current trends to measure implant stability.

Authors:  Vasanthi Swami; Vasantha Vijayaraghavan; Vinit Swami
Journal:  J Indian Prosthodont Soc       Date:  2016 Apr-Jun

6.  Effects of implant tilting and the loading direction on the displacement and micromotion of immediately loaded implants: an in vitro experiment and finite element analysis.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Sadami Tsutsumi; Tadaaki Kirita
Journal:  J Periodontal Implant Sci       Date:  2017-08-28       Impact factor: 2.614

7.  Evaluation of Primary Stability of Cylindrical and Tapered Implants in Different Bone Types by Measuring Implant Displacement: An In vitro Study.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Tadaaki Kirita
Journal:  Contemp Clin Dent       Date:  2019 Jul-Sep

8.  Peri-Implant Repair Using a Modified Implant Macrogeometry in Diabetic Rats: Biomechanical and Molecular Analyses of Bone-Related Markers.

Authors:  Hugo Robertson Sant'Anna; Marcio Zaffalon Casati; Mounir Colares Mussi; Fabiano Ribeiro Cirano; Suzana Peres Pimentel; Fernanda Vieira Ribeiro; Mônica Grazieli Corrêa
Journal:  Materials (Basel)       Date:  2022-03-21       Impact factor: 3.623

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.