Limited-angle effect compensation for respiratory binned cardiac SPECT.

Wenyuan Qi1, Yongyi Yang1, Miles N Wernick1, P Hendrik Pretorius2, Michael A King2.   

Abstract

PURPOSE: In cardiac single photon emission computed tomography (SPECT), respiratory-binned study is used to combat the motion blur associated with respiratory motion. However, owing to the variability in respiratory patterns during data acquisition, the acquired data counts can vary significantly both among respiratory bins and among projection angles within individual bins. If not properly accounted for, such variation could lead to artifacts similar to limited-angle effect in image reconstruction. In this work, the authors aim to investigate several reconstruction strategies for compensating the limited-angle effect in respiratory binned data for the purpose of reducing the image artifacts.
METHODS: The authors first consider a model based correction approach, in which the variation in acquisition time is directly incorporated into the imaging model, such that the data statistics are accurately described among both the projection angles and respiratory bins. Afterward, the authors consider an approximation approach, in which the acquired data are rescaled to accommodate the variation in acquisition time among different projection angles while the imaging model is kept unchanged. In addition, the authors also consider the use of a smoothing prior in reconstruction for suppressing the artifacts associated with limited-angle effect. In our evaluation study, the authors first used Monte Carlo simulated imaging with 4D NCAT phantom wherein the ground truth is known for quantitative comparison. The authors evaluated the accuracy of the reconstructed myocardium using a number of metrics, including regional and overall accuracy of the myocardium, uniformity and spatial resolution of the left ventricle (LV) wall, and detectability of perfusion defect using a channelized Hotelling observer. As a preliminary demonstration, the authors also tested the different approaches on five sets of clinical acquisitions.
RESULTS: The quantitative evaluation results show that the three compensation methods could all, but to different extents, reduce the reconstruction artifacts over no compensation. In particular, the model based approach reduced the mean-squared-error of the reconstructed myocardium by as much as 40%. Compared to the approach of data rescaling, the model based approach further improved both the overall and regional accuracy of the myocardium; it also further improved the lesion detectability and the uniformity of the LV wall. When ML reconstruction was used, the model based approach was notably more effective for improving the LV wall; when MAP reconstruction was used, the smoothing prior could reduce the noise level and artifacts with little or no increase in bias, but at the cost of a slight resolution loss of the LV wall. The improvements in image quality by the different compensation methods were also observed in the clinical acquisitions.
CONCLUSIONS: Compensating for the uneven distribution of acquisition time among both projection angles and respiratory bins can effectively reduce the limited-angle artifacts in respiratory-binned cardiac SPECT reconstruction. Direct incorporation of the time variation into the imaging model together with a smoothing prior in reconstruction can lead to the most improvement in the accuracy of the reconstructed myocardium.

Mesh:

Year:  2016        PMID: 26745937      PMCID: PMC4701077          DOI: 10.1118/1.4938579

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  Development of respiratory gated myocardial SPECT system.

Authors:  K Cho; S Kumiata; S Okada; T Kumazaki
Journal:  J Nucl Cardiol       Date:  1999 Jan-Feb       Impact factor: 5.952

2.  Globally convergent image reconstruction for emission tomography using relaxed ordered subsets algorithms.

Authors:  Sangtae Ahn; Jeffrey A Fessler
Journal:  IEEE Trans Med Imaging       Date:  2003-05       Impact factor: 10.048

3.  Comparison of analytic and algebraic methods for motion-compensated cone-beam CT reconstruction of the thorax.

Authors:  Simon Rit; David Sarrut; Laurent Desbat
Journal:  IEEE Trans Med Imaging       Date:  2009-02-10       Impact factor: 10.048

4.  Respiratory motion correction in gated cardiac SPECT using quaternion-based, rigid-body registration.

Authors:  Jason G Parker; Bernard A Mair; David R Gilland
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

5.  A Monte Carlo program for the simulation of scintillation camera characteristics.

Authors:  M Ljungberg; S E Strand
Journal:  Comput Methods Programs Biomed       Date:  1989-08       Impact factor: 5.428

6.  Addition of a channel mechanism to the ideal-observer model.

Authors:  K J Myers; H H Barrett
Journal:  J Opt Soc Am A       Date:  1987-12       Impact factor: 2.129

7.  A comparison between amplitude sorting and phase-angle sorting using external respiratory measurement for 4D CT.

Authors:  Wei Lu; Parag J Parikh; James P Hubenschmidt; Jeffrey D Bradley; Daniel A Low
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

8.  Effect of patient motion on tomographic myocardial perfusion imaging.

Authors:  J A Cooper; P H Neumann; B K McCandless
Journal:  J Nucl Med       Date:  1992-08       Impact factor: 10.057

9.  A quantitative evaluation study of four-dimensional gated cardiac SPECT reconstruction.

Authors:  Mingwu Jin; Yongyi Yang; Xiaofeng Niu; Thibault Marin; Jovan G Brankov; Bing Feng; P Hendrik Pretorius; Michael A King; Miles N Wernick
Journal:  Phys Med Biol       Date:  2009-09-01       Impact factor: 3.609

10.  List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations.

Authors:  F Lamare; M J Ledesma Carbayo; T Cresson; G Kontaxakis; A Santos; C Cheze Le Rest; A J Reader; D Visvikis
Journal:  Phys Med Biol       Date:  2007-08-09       Impact factor: 3.609

View more
  6 in total

1.  Respiratory motion reduction with a dual gating approach in myocardial perfusion SPECT: Effect on left ventricular functional parameters.

Authors:  Matti J Kortelainen; Tuomas M Koivumäki; Marko J Vauhkonen; Marja K Hedman; Satu T J Kärkkäinen; Juanita Niño Quintero; Mikko A Hakulinen
Journal:  J Nucl Cardiol       Date:  2017-03-16       Impact factor: 5.952

2.  Improving perfusion defect detection with respiratory motion correction in cardiac SPECT at standard and reduced doses.

Authors:  Chao Song; Yongyi Yang; Albert Juan Ramon; Miles N Wernick; P Hendrik Pretorius; Karen L Johnson; Piotr J Slomka; Michael A King
Journal:  J Nucl Cardiol       Date:  2018-07-30       Impact factor: 5.952

3.  Novel SPECT Technologies and Approaches in Cardiac Imaging.

Authors:  Piotr Slomka; Guang-Uei Hung; Guido Germano; Daniel S Berman
Journal:  Cardiovasc Innov Appl       Date:  2016-12-01

4.  4-D Reconstruction With Respiratory Correction for Gated Myocardial Perfusion SPECT.

Authors:  Wenyuan Qi; Yongyi Yang; Chao Song; Miles N Wernick; P Hendrik Pretorius; Michael A King
Journal:  IEEE Trans Med Imaging       Date:  2017-04-04       Impact factor: 10.048

5.  New Trends in Quantitative Nuclear Cardiology Methods.

Authors:  Javier Gomez; Rami Doukky; Guido Germano; Piotr Slomka
Journal:  Curr Cardiovasc Imaging Rep       Date:  2018-01-19

6.  Time-modified OSEM algorithm for more robust assessment of left ventricular dyssynchrony with phase analysis in ECG-gated myocardial perfusion SPECT.

Authors:  Matti J Kortelainen; Tuomas M Koivumäki; Marko J Vauhkonen; Mikko A Hakulinen
Journal:  EJNMMI Phys       Date:  2019-12-27
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.