Improving perfusion defect detection with respiratory motion correction in cardiac SPECT at standard and reduced doses.

Chao Song1, Yongyi Yang2, Albert Juan Ramon1, Miles N Wernick1, P Hendrik Pretorius3, Karen L Johnson3, Piotr J Slomka4, Michael A King3.   

Abstract

BACKGROUND: In cardiac SPECT perfusion imaging, respiratory motion can cause non-uniform blurring in the reconstructed myocardium. We investigate the potential benefit of respiratory correction with respiratory-binned acquisitions, both at standard dose and at reduced dose, for defect detection and for left ventricular (LV) wall resolution.
METHODS: We applied two reconstruction methods for respiratory motion correction: post-reconstruction motion correction (PMC) and motion-compensated reconstruction (MCR), and compared with reconstruction without motion correction (Non-MC). We quantified the presence of perfusion defects in reconstructed images by using the total perfusion deficit (TPD) scores and conducted receiver-operating-characteristic (ROC) studies using TPD. We quantified the LV spatial resolution by using the FWHM of its cross-sectional intensity profile.
RESULTS: The values in the area-under-the-ROC-curve (AUC) achieved by MCR, PMC, and Non-MC at standard dose were 0.835, 0.830, and 0.798, respectively. Similar AUC improvements were also obtained by MCR and PMC over Non-MC at 50%, 25%, and 12.5% of full dose. Improvements in LV resolution were also observed with motion correction.
CONCLUSIONS: Respiratory-binned acquisitions can improve perfusion-defect detection accuracy over traditional reconstruction both at standard dose and at reduced dose. Motion correction may contribute to achieving further dose reduction while maintaining the diagnostic accuracy of traditional acquisitions.

Entities:  

Keywords:  CAD; MPI; SPECT; image reconstruction

Year:  2018        PMID: 30062470     DOI: 10.1007/s12350-018-1374-9

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  31 in total

Review 1.  A survey of medical image registration.

Authors:  J B Maintz; M A Viergever
Journal:  Med Image Anal       Date:  1998-03       Impact factor: 8.545

2.  Automated quantification of myocardial perfusion SPECT using simplified normal limits.

Authors:  Piotr J Slomka; Hidetaka Nishina; Daniel S Berman; Cigdem Akincioglu; Aiden Abidov; John D Friedman; Sean W Hayes; Guido Germano
Journal:  J Nucl Cardiol       Date:  2005 Jan-Feb       Impact factor: 5.952

3.  New reconstruction algorithm allows shortened acquisition time for myocardial perfusion SPECT.

Authors:  Ines Valenta; Valerie Treyer; Lars Husmann; Oliver Gaemperli; Michael J Schindler; Bernhard A Herzog; Patrick Veit-Heibach; Ronny R Buechel; René Nkoulou; Aju P Pazhenkottil; Philipp A Kaufmann
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-11-18       Impact factor: 9.236

4.  Clinical evaluation of a block sequential regularized expectation maximization reconstruction algorithm in 18F-FDG PET/CT studies.

Authors:  Bert-Ram Sah; Paul Stolzmann; Gaspar Delso; Scott D Wollenweber; Martin Hüllner; Yahya A Hakami; Marcelo A Queiroz; Felipe de Galiza Barbosa; Gustav K von Schulthess; Carsten Pietsch; Patrick Veit-Haibach
Journal:  Nucl Med Commun       Date:  2017-01       Impact factor: 1.690

5.  4D reconstruction for low-dose cardiac gated SPECT.

Authors:  Mingwu Jin; Xiaofeng Niu; Wenyuan Qi; Yongyi Yang; Joyoni Dey; Michael A King; Seth Dahlberg; Miles N Wernick
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

6.  Motion-compensated image reconstruction vs postreconstruction correction in respiratory-binned SPECT with standard and reduced-dose acquisitions.

Authors:  Chao Song; Yongyi Yang; Wenyuan Qi; Miles N Wernick; P Hendrik Pretorius; Michael A King
Journal:  Med Phys       Date:  2018-05-13       Impact factor: 4.071

7.  Effect of patient motion on tomographic myocardial perfusion imaging.

Authors:  J A Cooper; P H Neumann; B K McCandless
Journal:  J Nucl Med       Date:  1992-08       Impact factor: 10.057

8.  4-D Reconstruction With Respiratory Correction for Gated Myocardial Perfusion SPECT.

Authors:  Wenyuan Qi; Yongyi Yang; Chao Song; Miles N Wernick; P Hendrik Pretorius; Michael A King
Journal:  IEEE Trans Med Imaging       Date:  2017-04-04       Impact factor: 10.048

9.  Human-observer receiver-operating-characteristic evaluation of attenuation, scatter, and resolution compensation strategies for (99m)Tc myocardial perfusion imaging.

Authors:  Manoj V Narayanan; Michael A King; P Hendrik Pretorius; Seth T Dahlberg; Frederick Spencer; Ellen Simon; Eric Ewald; Edward Healy; Kirk MacNaught; Jeffrey A Leppo
Journal:  J Nucl Med       Date:  2003-11       Impact factor: 10.057

10.  Investigation of dose reduction in cardiac perfusion SPECT via optimization and choice of the image reconstruction strategy.

Authors:  Albert Juan Ramon; Yongyi Yang; P Hendrik Pretorius; Piotr J Slomka; Karen L Johnson; Michael A King; Miles N Wernick
Journal:  J Nucl Cardiol       Date:  2017-05-23       Impact factor: 3.872

View more
  1 in total

1.  Deep learning, another important tool for improving acquisition efficiency in SPECT MPI Imaging.

Authors:  Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2020-05-17       Impact factor: 5.952

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.