Advanced Modeled Iterative Reconstruction in Low-Tube-Voltage Contrast-Enhanced Neck CT: Evaluation of Objective and Subjective Image Quality.

J-E Scholtz1, M Kaup2, K Hüsers2, M H Albrecht2, B Bodelle2, S C Metzger2, J M Kerl2, R W Bauer2, T Lehnert2, T J Vogl2, J L Wichmann3.   

Abstract

BACKGROUND AND
PURPOSE: Dose-saving techniques in neck CT cause increased image noise that can be counteracted by iterative reconstruction. Our aim was to evaluate the image quality of advanced modeled iterative reconstruction (ADMIRE) in contrast-enhanced low-tube-voltage neck CT.
MATERIALS AND METHODS: Sixty-one patients underwent 90-kV(peak) neck CT by using third-generation 192-section dual-source CT. Image series were reconstructed with standard filtered back-projection and ADMIRE strength levels 1, 3, and 5. Attenuation and noise of the sternocleidomastoid muscle, internal jugular vein, submandibular gland, tongue, subscapularis muscle, and cervical fat were measured. Signal-to-noise and contrast-to-noise ratios were calculated. Two radiologists assessed image noise, image contrast, delineation of smaller structures, and overall diagnostic acceptability. Interobserver agreement was calculated.
RESULTS: Image noise was significantly reduced by using ADMIRE compared with filtered back-projection with the lowest noise observed in ADMIRE 5 (filtered back-projection, 9.4 ± 2.4 Hounsfield units [HU]; ADMIRE 1, 8.3 ± 2.8 HU; ADMIRE 3, 6.7 ± 2.0 HU; ADMIRE 5, 5.4 ± 1.7 HU; all, P < .001). Sternocleidomastoid SNR and internal jugular vein-sternocleidomastoid contrast-to-noise ratios were significantly higher for ADMIRE with the best results in ADMIRE 5 (all, P < .001). Subjective image quality and image contrast of ADMIRE 3 and 5 were consistently rated better than those for filtered back-projection and ADMIRE 1 (all, P < .001). Image noise was rated highest for ADMIRE 5 (all, P < .005). Delineation of smaller structures was voted higher in all ADMIRE strength levels compared with filtered back-projection (P < .001). Global interobserver agreement was good (0.75).
CONCLUSIONS: Contrast-enhanced 90-kVp neck CT is feasible, and ADMIRE 5 shows superior objective image quality compared with filtered back-projection. ADMIRE 3 and 5 show the best subjective image quality.
© 2016 by American Journal of Neuroradiology.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26427836      PMCID: PMC7960220          DOI: 10.3174/ajnr.A4502

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  24 in total

Review 1.  Iterative reconstruction methods in X-ray CT.

Authors:  Marcel Beister; Daniel Kolditz; Willi A Kalender
Journal:  Phys Med       Date:  2012-02-10       Impact factor: 2.685

2.  Ultra-low dose dual-source high-pitch computed tomography of the paranasal sinus: diagnostic sensitivity and radiation dose.

Authors:  Boris Schulz; Stefan Potente; Stefan Zangos; Ingke Friedrichs; Ralf W Bauer; Matthias Kerl; Thomas J Vogl; Martin M Mack
Journal:  Acta Radiol       Date:  2012-04-19       Impact factor: 1.990

Review 3.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

Review 4.  Imaging of vascular lesions of the head and neck.

Authors:  Julius Griauzde; Ashok Srinivasan
Journal:  Radiol Clin North Am       Date:  2014-10-05       Impact factor: 2.303

5.  Diagnostic Performance of an Advanced Modeled Iterative Reconstruction Algorithm for Low-Contrast Detectability with a Third-Generation Dual-Source Multidetector CT Scanner: Potential for Radiation Dose Reduction in a Multireader Study.

Authors:  Justin Solomon; Achille Mileto; Juan Carlos Ramirez-Giraldo; Ehsan Samei
Journal:  Radiology       Date:  2015-03-04       Impact factor: 11.105

6.  Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose.

Authors:  Matthias S May; Manuel R Kramer; Achim Eller; Wolfgang Wuest; Michael Scharf; Michael Brand; Marc Saake; Bernhard Schmidt; Michael Uder; Michael M Lell
Journal:  Neuroradiology       Date:  2014-06-25       Impact factor: 2.804

7.  Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions.

Authors:  H Greess; H Wolf; U Baum; M Lell; M Pirkl; W Kalender; W A Bautz
Journal:  Eur Radiol       Date:  2000       Impact factor: 5.315

8.  Objective and subjective image quality of primary and recurrent squamous cell carcinoma on head and neck low-tube-voltage 80-kVp computed tomography.

Authors:  Jan-Erik Scholtz; Moritz Kaup; Johannes Kraft; Eva-Maria Nöske; Friedrich Scheerer; Boris Schulz; Iris Burck; Jens Wagenblast; J Matthias Kerl; Ralf W Bauer; Thomas Lehnert; Thomas J Vogl; Julian L Wichmann
Journal:  Neuroradiology       Date:  2015-03-26       Impact factor: 2.804

Review 9.  Imaging of acute head and neck infections.

Authors:  Aldo Gonzalez-Beicos; Diego Nunez
Journal:  Radiol Clin North Am       Date:  2012-01       Impact factor: 2.303

10.  Projected cancer risks from computed tomographic scans performed in the United States in 2007.

Authors:  Amy Berrington de González; Mahadevappa Mahesh; Kwang-Pyo Kim; Mythreyi Bhargavan; Rebecca Lewis; Fred Mettler; Charles Land
Journal:  Arch Intern Med       Date:  2009-12-14
View more
  6 in total

Review 1.  Regularization strategies in statistical image reconstruction of low-dose x-ray CT: A review.

Authors:  Hao Zhang; Jing Wang; Dong Zeng; Xi Tao; Jianhua Ma
Journal:  Med Phys       Date:  2018-09-10       Impact factor: 4.071

2.  Head and neck angiography at 70 kVp with a third-generation dual-source CT system in patients: comparison with 100 kVp.

Authors:  Yu Chen; Xiaobo Zhang; Huadan Xue; Yuanli Zhu; Yun Wang; Yumei Li; Zhuhua Zhang; Zhengyu Jin
Journal:  Neuroradiology       Date:  2017-08-24       Impact factor: 2.804

3.  Advantages of 70-kV CT Angiography for the Visualization of the Adamkiewicz Artery: Comparison with 120-kV Imaging.

Authors:  S Shimoyama; T Nishii; Y Watanabe; A K Kono; K Kagawa; S Takahashi; K Sugimura
Journal:  AJNR Am J Neuroradiol       Date:  2017-09-14       Impact factor: 3.825

4.  Diagnostic yield of 90-kVp low-tube-voltage carotid and intracerebral CT-angiography: effects on radiation dose, image quality and diagnostic performance for the detection of carotid stenosis.

Authors:  Doris Leithner; Julian L Wichmann; Scherwin Mahmoudi; Simon S Martin; Moritz H Albrecht; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Br J Radiol       Date:  2018-03-08       Impact factor: 3.039

5.  Pediatric head computed tomography with advanced modeled iterative reconstruction: focus on image quality and reduction of radiation dose.

Authors:  Hyun-Hae Cho; So Mi Lee; Sun Kyoung You
Journal:  Pediatr Radiol       Date:  2019-10-19

6.  Third-generation dual-source CT of the neck using automated tube voltage adaptation in combination with advanced modeled iterative reconstruction: evaluation of image quality and radiation dose.

Authors:  Jan-Erik Scholtz; Julian L Wichmann; Kristina Hüsers; Moritz H Albrecht; Martin Beeres; Ralf W Bauer; Thomas J Vogl; Boris Bodelle
Journal:  Eur Radiol       Date:  2015-11-11       Impact factor: 5.315

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.