Diagnostic Performance of an Advanced Modeled Iterative Reconstruction Algorithm for Low-Contrast Detectability with a Third-Generation Dual-Source Multidetector CT Scanner: Potential for Radiation Dose Reduction in a Multireader Study.

Justin Solomon1, Achille Mileto, Juan Carlos Ramirez-Giraldo, Ehsan Samei.   

Abstract

PURPOSE: To assess the effect of radiation dose reduction on low-contrast detectability by using an advanced modeled iterative reconstruction (ADMIRE; Siemens Healthcare, Forchheim, Germany) algorithm in a contrast-detail phantom with a third-generation dual-source multidetector computed tomography (CT) scanner.
MATERIALS AND METHODS: A proprietary phantom with a range of low-contrast cylindrical objects, representing five contrast levels (range, 5-20 HU) and three sizes (range, 2-6 mm) was fabricated with a three-dimensional printer and imaged with a third-generation dual-source CT scanner at various radiation dose index levels (range, 0.74-5.8 mGy). Image data sets were reconstructed by using different section thicknesses (range, 0.6-5.0 mm) and reconstruction algorithms (filtered back projection [FBP] and ADMIRE with a strength range of three to five). Eleven independent readers blinded to technique and reconstruction method assessed all data sets in two reading sessions by measuring detection accuracy with a two-alternative forced choice approach (first session) and by scoring the total number of visible object groups (second session). Dose reduction potentials based on both reading sessions were estimated. Results between FBP and ADMIRE were compared by using both paired t tests and analysis of variance tests at the 95% significance level.
RESULTS: During the first session, detection accuracy increased with increasing contrast, size, and dose index (diagnostic accuracy range, 50%-87%; interobserver variability, ±7%). When compared with FBP, ADMIRE improved detection accuracy by 5.2% on average across the investigated variables (P < .001). During the second session, a significantly increased number of visible objects was noted with increasing radiation dose index, section thickness, and ADMIRE strength over FBP (up to 80% more visible objects, P < .001). Radiation dose reduction potential ranged from 56% to 60% and from 4% to 80% during the two sessions, respectively.
CONCLUSION: Low-contrast detectability performance increased with increasing object size, object contrast, dose index, section thickness, and ADMIRE strength. Compared with FBP, ADMIRE allows a substantial radiation dose reduction while preserving low-contrast detectability. Online supplemental material is available for this article. RSNA, 2015

Mesh:

Substances:

Year:  2015        PMID: 25751228     DOI: 10.1148/radiol.15142005

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  43 in total

1.  Consistency of Renal Stone Volume Measurements Across CT Scanner Model and Reconstruction Algorithm Configurations.

Authors:  Alice E Huang; Juan C Montoya; Maria Shiung; Shuai Leng; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2017-04-12       Impact factor: 3.959

2.  Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT.

Authors:  Julian L Wichmann; Andrew D Hardie; U Joseph Schoepf; Lloyd M Felmly; Jonathan D Perry; Akos Varga-Szemes; Stefanie Mangold; Damiano Caruso; Christian Canstein; Thomas J Vogl; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2016-05-10       Impact factor: 5.315

Review 3.  Imaging Techniques to Aid IR Treatment of Musculoskeletal Malignancy.

Authors:  A Kyle Jones; Steven Yevich
Journal:  Semin Intervent Radiol       Date:  2019-03-29       Impact factor: 1.513

4.  Impact of model-based iterative reconstruction on low-contrast lesion detection and image quality in abdominal CT: a 12-reader-based comparative phantom study with filtered back projection at different tube voltages.

Authors:  André Euler; Bram Stieltjes; Zsolt Szucs-Farkas; Reto Eichenberger; Clemens Reisinger; Anna Hirschmann; Caroline Zaehringer; Achim Kircher; Matthias Streif; Sabine Bucher; David Buergler; Luigia D'Errico; Sebastién Kopp; Markus Wilhelm; Sebastian T Schindera
Journal:  Eur Radiol       Date:  2017-04-03       Impact factor: 5.315

5.  Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography.

Authors:  Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-09-22

Review 6.  Regularization strategies in statistical image reconstruction of low-dose x-ray CT: A review.

Authors:  Hao Zhang; Jing Wang; Dong Zeng; Xi Tao; Jianhua Ma
Journal:  Med Phys       Date:  2018-09-10       Impact factor: 4.071

7.  Diagnostic performance of reduced-dose CT with a hybrid iterative reconstruction algorithm for the detection of hypervascular liver lesions: a phantom study.

Authors:  Atsushi Nakamoto; Yoshikazu Tanaka; Hiroshi Juri; Go Nakai; Shushi Yoshikawa; Yoshifumi Narumi
Journal:  Eur Radiol       Date:  2016-12-12       Impact factor: 5.315

8.  Diagnostic yield of 90-kVp low-tube-voltage carotid and intracerebral CT-angiography: effects on radiation dose, image quality and diagnostic performance for the detection of carotid stenosis.

Authors:  Doris Leithner; Julian L Wichmann; Scherwin Mahmoudi; Simon S Martin; Moritz H Albrecht; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Br J Radiol       Date:  2018-03-08       Impact factor: 3.039

9.  Interchangeability between real and three-dimensional simulated lung tumors in computed tomography: an interalgorithm volumetry study.

Authors:  Marthony Robins; Justin Solomon; Jocelyn Hoye; Taylor Smith; Yuese Zheng; Lukas Ebner; Kingshuk Roy Choudhury; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2018-09-24

10.  Statistical model based iterative reconstruction in clinical CT systems. Part III. Task-based kV/mAs optimization for radiation dose reduction.

Authors:  Ke Li; Daniel Gomez-Cardona; Jiang Hsieh; Meghan G Lubner; Perry J Pickhardt; Guang-Hong Chen
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.