Estimation of organ doses from kilovoltage cone-beam CT imaging used during radiotherapy patient position verification.

Daniel E Hyer1, David E Hintenlang.   

Abstract

PURPOSE: The purpose of this study was to develop a practical method for estimating organ doses from kilovoltage cone-beam CT (CBCT) that can be performed with readily available phantoms and dosimeters. The accuracy of organ dose estimates made using the ImPACT patient dose calculator was also evaluated.
METHODS: A 100 mm pencil chamber and standard CT dose index (CTDI) phantoms were used to measure the cone-beam dose index (CBDI). A weighted CBDI (CBDI(W)) was then calculated from these measurements to represent the average volumetric dose in the CTDI phantom. By comparing CBDI(W) to the previously published organ doses, organ dose conversion coefficients were developed. The measured CBDI values were also used as inputs for the ImPACT calculator to estimate organ doses. All CBDI dose measurements were performed on both the Elekta XVI and Varian OBI at three clinically relevant locations: Head, chest, and pelvis.
RESULTS: The head, chest, and pelvis protocols yielded CBDI(W) values of 0.98, 16.62, and 24.13 mGy for the XVI system and 5.17, 6.14, and 21.57 mGy for the OBI system, respectively. Organ doses estimated with the ImPACT CT dose calculator showed a large range of variation from the previously measured organ doses, demonstrating its limitations for use with CBCT.
CONCLUSIONS: The organ dose conversion coefficients developed in this work relate CBDI(W) values to organ doses previously measured using the same clinical protocols. Ultimately, these coefficients will allow for the quick estimation of organ doses from routine measurements performed using standard CTDI phantoms and pencil chambers.

Entities:  

Mesh:

Year:  2010        PMID: 20964180     DOI: 10.1118/1.3476459

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Accounting for patient size in the optimization of dose and image quality of pelvis cone beam CT protocols on the Varian OBI system.

Authors:  Tim J Wood; Craig S Moore; Carl J Horsfield; John R Saunderson; Andrew W Beavis
Journal:  Br J Radiol       Date:  2015-09-30       Impact factor: 3.039

2.  Evaluation of three presets for four-dimensional cone beam CT in lung radiotherapy verification by visual grading analysis.

Authors:  Sally A Kember; Vibeke N Hansen; Martin F Fast; Simeon Nill; Fiona McDonald; Merina Ahmed; Karen Thomas; Helen A McNair
Journal:  Br J Radiol       Date:  2016-04-25       Impact factor: 3.039

3.  Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies.

Authors:  Jarryd G Buckley; Dean Wilkinson; Alessandra Malaroda; Peter Metcalfe
Journal:  J Appl Clin Med Phys       Date:  2017-12-19       Impact factor: 2.102

4.  Phantom-based evaluation of dose exposure of ultrafast combined kV-MV-CBCT towards clinical implementation for IGRT of lung cancer.

Authors:  Anna Arns; Manuel Blessing; Jens Fleckenstein; Dzmitry Stsepankou; Judit Boda-Heggemann; Juergen Hesser; Frank Lohr; Frederik Wenz; Hansjoerg Wertz
Journal:  PLoS One       Date:  2017-11-10       Impact factor: 3.240

5.  eNAL++: a new and effective off-line correction protocol for rotational setup errors when using a robotic couch.

Authors:  Daan Martens; Mark Luesink; Henk Huizenga; Kasper L Pasma
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

6.  Difference in performance between 3D and 4D CBCT for lung imaging: a dose and image quality analysis.

Authors:  Sheeba Thengumpallil; Kathleen Smith; Pascal Monnin; Jean Bourhis; François Bochud; Raphaël Moeckli
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

7.  Radiation treatment planning and delivery strategies for a pregnant brain tumor patient.

Authors:  Zacariah E Labby; Brendan Barraclough; R Adam Bayliss; Abigail E Besemer; David A P Dunkerley; Steven P Howard
Journal:  J Appl Clin Med Phys       Date:  2018-07-30       Impact factor: 2.102

8.  On the improvement of CBCT image quality for soft tissue-based SRS localization.

Authors:  Weihua Mao; Stephen J Gardner; Karen Chin Snyder; Ning Winston Wen; Hualiang Zhong; Haisen Li; Paul Jackson; Mira Shah; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2018-10-07       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.