Inclusion of the dose from kilovoltage cone beam CT in the radiation therapy treatment plans.

Parham Alaei1, George Ding, Huaiqun Guan.   

Abstract

PURPOSE: Cone beam CT is increasingly being used for daily patient positioning verification during radiation therapy treatments. The daily use of CBCT could lead to accumulated patient doses higher than the older technique of weekly portal imaging. There have been several studies focusing on measurement or calculation of the patient dose from CBCT recently.
METHODS: This study investigates the feasibility of configuring a kV x-ray source in a commercial treatment planning system to calculate the dose to patient resulting from an IGRT procedure. The method proposed in this article can be used to calculate dose from CBCT imaging procedure and include that in the patient treatment plans.
RESULTS: The kilovoltage beam generated by the CBCT imager has been modeled using the planning system. The modeled profiles agree with the measured ones to within 5%. The modeled beam was used to calculate dose to phantom in the pelvic region and the calculations were compared to TLD measurements. The agreement between calculated and measured doses ranges from 0% to 19% in soft tissue with larger variations observed near and within the bone.
CONCLUSIONS: The modeling of the beam produces reasonable results and the dose calculation comparisons indicate the potential for computing kilovoltage CBCT doses using a treatment planning system. Further improvements in the dose calculation algorithm are necessary, especially for dose calculations in and near the bone.

Entities:  

Mesh:

Year:  2010        PMID: 20175487     DOI: 10.1118/1.3271582

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Optimization of an on-board imaging system for extremely rapid radiation therapy.

Authors:  Erica M Cherry Kemmerling; Meng Wu; He Yang; Peter G Maxim; Billy W Loo; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  Dosimetric properties and commissioning of cone-beam CT image beam line with a carbon target.

Authors:  Y Dzierma; F G Nuesken; N P Licht; C Ruebe
Journal:  Strahlenther Onkol       Date:  2013-05-30       Impact factor: 3.621

Review 3.  kV cone-beam CT-based IGRT: a clinical review.

Authors:  Judit Boda-Heggemann; Frank Lohr; Frederik Wenz; Michael Flentje; Matthias Guckenberger
Journal:  Strahlenther Onkol       Date:  2011-04-26       Impact factor: 3.621

4.  Monte Carlo calculation of imaging doses from diagnostic multidetector CT and kilovoltage cone-beam CT as part of prostate cancer treatment plans.

Authors:  Aiping Ding; Jianwei Gu; Alexei V Trofimov; X George Xu
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

5.  [Organizational models of inpatient emergency and acute care].

Authors:  A Kaltwasser; C Dodt
Journal:  Med Klin Intensivmed Notfmed       Date:  2014-10       Impact factor: 0.840

6.  Dosimetry challenges for implementing emerging technologies.

Authors:  Fang-Fang Yin; Mark Oldham; Jing Cai; Qiuwen Wu
Journal:  J Phys Conf Ser       Date:  2010

7.  Reducing scan angle using adaptive prior knowledge for a limited-angle intrafraction verification (LIVE) system for conformal arc radiotherapy.

Authors:  Yawei Zhang; Fang-Fang Yin; You Zhang; Lei Ren
Journal:  Phys Med Biol       Date:  2017-03-24       Impact factor: 3.609

8.  Image quality and dose distributions of three linac-based imaging modalities.

Authors:  Yvonne Dzierma; Evemarie Ames; Frank Nuesken; Jan Palm; Norbert Licht; Christian Rübe
Journal:  Strahlenther Onkol       Date:  2014-12-20       Impact factor: 3.621

9.  Image acquisition optimization of a limited-angle intrafraction verification (LIVE) system for lung radiotherapy.

Authors:  Yawei Zhang; Xinchen Deng; Fang-Fang Yin; Lei Ren
Journal:  Med Phys       Date:  2017-11-30       Impact factor: 4.071

10.  Accounting for patient size in the optimization of dose and image quality of pelvis cone beam CT protocols on the Varian OBI system.

Authors:  Tim J Wood; Craig S Moore; Carl J Horsfield; John R Saunderson; Andrew W Beavis
Journal:  Br J Radiol       Date:  2015-09-30       Impact factor: 3.039

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.