Literature DB >> 32822249

Patient-based low dose cone beam CT acquisition settings for prostate image-guided radiotherapy treatments on a Varian TrueBeam linear accelerator.

Maria Antonietta Piliero1, Margherita Casiraghi1, Davide Giovanni Bosetti2, Simona Cima2, Letizia Deantonio2, Stefano Leva2, Francesco Martucci2, Marino Tettamanti2, Francesco Pupillo1, Luca Bellesi1, Antonella Richetti2, Stefano Presilla1.   

Abstract

OBJECTIVE: To evaluate the performance of low dose cone beam CT (CBCT) acquisition protocols for image-guided radiotherapy of prostate cancer.
METHODS: CBCT images of patients undergoing prostate cancer radiotherapy were acquired with the settings currently used in our department and two low dose settings at 50% and 63% lower exposure. Four experienced radiation oncologists and two radiation therapy technologists graded the images on five image quality characteristics. The scores were analysed through Visual Grading Regression, using the acquisition settings and the patient size as covariates.
RESULTS: The low dose acquisition settings have no impact on the image quality for patients with body profile length at hip level below 100 cm.
CONCLUSIONS: A reduction of about 60% of the dose is feasible for patients with size below 100 cm. The visibility of low contrast features can be compromised if using the low dose acquisition settings for patients with hip size above 100 cm. ADVANCES IN KNOWLEDGE: Low dose CBCT acquisition protocols for the pelvis, based on subjective evaluation of patient images.

Entities:  

Mesh:

Year:  2020        PMID: 32822249      PMCID: PMC8519649          DOI: 10.1259/bjr.20200412

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  25 in total

1.  Quantifying the potential for dose reduction with visual grading regression.

Authors:  O Smedby; M Fredrikson; J De Geer; L Borgen; M Sandborg
Journal:  Br J Radiol       Date:  2012-06-20       Impact factor: 3.039

2.  Cone beam CT dose reduction in prostate radiotherapy using Likert scale methods.

Authors:  Keith A Langmack; Louise A Newton; Suzanne Jordan; Ruth Smith
Journal:  Br J Radiol       Date:  2015-12-21       Impact factor: 3.039

3.  Visual grading regression: analysing data from visual grading experiments with regression models.

Authors:  O Smedby; M Fredrikson
Journal:  Br J Radiol       Date:  2010-03-11       Impact factor: 3.039

4.  Comprehensive Monte Carlo study of patient doses from cone-beam CT imaging in radiotherapy.

Authors:  T E Marchant; K D Joshi
Journal:  J Radiol Prot       Date:  2016-12-06       Impact factor: 1.394

5.  Image guidance doses delivered during radiotherapy: Quantification, management, and reduction: Report of the AAPM Therapy Physics Committee Task Group 180.

Authors:  George X Ding; Parham Alaei; Bruce Curran; Ryan Flynn; Michael Gossman; T Rock Mackie; Moyed Miften; Richard Morin; X George Xu; Timothy C Zhu
Journal:  Med Phys       Date:  2018-03-24       Impact factor: 4.071

6.  Reducing radiation exposure to patients from kV-CBCT imaging.

Authors:  George X Ding; Peter Munro; Jason Pawlowski; Arnold Malcolm; Charles W Coffey
Journal:  Radiother Oncol       Date:  2010-09-16       Impact factor: 6.280

7.  Accounting for patient size in the optimization of dose and image quality of pelvis cone beam CT protocols on the Varian OBI system.

Authors:  Tim J Wood; Craig S Moore; Carl J Horsfield; John R Saunderson; Andrew W Beavis
Journal:  Br J Radiol       Date:  2015-09-30       Impact factor: 3.039

8.  High Agreement and High Prevalence: The Paradox of Cohen's Kappa.

Authors:  Slavica Zec; Nicola Soriani; Rosanna Comoretto; Ileana Baldi
Journal:  Open Nurs J       Date:  2017-10-31

9.  Optimizing the target detectability of cone beam CT performed in image-guided radiation therapy for patients of different body sizes.

Authors:  Ching-Ching Yang; Pei-Chieh Yu; Jau-Ming Ruan; Yu-Cheng Chen
Journal:  J Appl Clin Med Phys       Date:  2018-03-08       Impact factor: 2.102

10.  Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction.

Authors:  Bharti Kataria; Jonas Nilsson Althén; Örjan Smedby; Anders Persson; Hannibal Sökjer; Michael Sandborg
Journal:  Eur Radiol       Date:  2018-01-24       Impact factor: 5.315

View more
  1 in total

Review 1.  Integrated MRI-guided radiotherapy - opportunities and challenges.

Authors:  Paul J Keall; Caterina Brighi; Carri Glide-Hurst; Gary Liney; Paul Z Y Liu; Suzanne Lydiard; Chiara Paganelli; Trang Pham; Shanshan Shan; Alison C Tree; Uulke A van der Heide; David E J Waddington; Brendan Whelan
Journal:  Nat Rev Clin Oncol       Date:  2022-04-19       Impact factor: 65.011

  1 in total