| Literature DB >> 26348197 |
Yu Cong1, Nirong Bao1, Jianning Zhao1, Guangping Mao1.
Abstract
BACKGROUND: The aim of this study was to compare the accuracy of cervical pedicle screw placement between a three-dimensional guidance system and manual manipulation.Entities:
Mesh:
Year: 2015 PMID: 26348197 PMCID: PMC4571536 DOI: 10.12659/MSM.894074
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Nails insertion results of locator guidance.
| C3 (%) | C4 (%) | C5 (%) | C6 (%) | C7 (%) | Total (%) | |
|---|---|---|---|---|---|---|
| Within pedicle | 16 (88.9) | 16 (88.9) | 16 (88.9) | 15 (83.3) | 18 (100) | 81 (90) |
| Outside pedicle | 2 (11.1) | 2 (11.1) | 2 (11.1) | 3 (16.7) | 0 (0.0) | 9 (10) |
| NRI | 0 (0.0) | 0 (0.0) | 1 (5.6) | 1 (5.6) | 0 (0.0) | 2 (2.2) |
| VAI | 0 (0.0) | 0 (0.0) | 1 (5.6) | 0 (0.0) | 0 (0.0) | 1 (1.1) |
| SCI | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
NRI – nerve root injury; VAI – vertebral artery injury; SCI – spinal cord injury.
The number of offset error nails of locator guidance in its cross-sectional CT Image.
| C3 (%) | C4 (%) | C5 (%) | C6 (%) | C7 (%) | Total (%) | |
|---|---|---|---|---|---|---|
| I | 11 (61.1) | 14 (77.8) | 15 (83.3) | 10 (55.6) | 13 (72.2) | 63 (70.0) |
| II | 6 (33.3) | 2 (11.1) | 3 (16.7) | 6 (33.3 | 2 (11.1) | 19 (21.1) |
| V | 1 (5.6) | 2 (11.1) | 0 (0.0) | 2 (11.1) | 3 (16.7) | 8 (8.9) |
I – vertical offset distance of the nail axis ≤1 mm and simultaneity an angular offset ≤5° from the pedicle axis.
Figure 1There were significant differences between group A and B in the axial CT images and between the axial and sagittal CT images of group A. There were no significant difference between group A and B in the axial CT images or between axial and sagittal CT images of group B.
Figure 2There were significant differences between group A and B in the axial and sagittal CT image and between axial and sagittal CT images of group A. There was no significant difference between axial and sagittal CT images of group B.
Nails insertion results of manual manipulation.
| C3 (%) | C4 (%) | C5 (%) | C6 (%) | C7 (%) | Total (%) | |
|---|---|---|---|---|---|---|
| Within pedicle | 8 (44.4) | 13 (72.2) | 5 (27.8) | 14 (77.8) | 10 (55.6) | 50 (55.6) |
| Outside pedicle | 10 (55.6) | 5 (27.8) | 13 (72.2) | 4 (22.2) | 8 (44.4) | 40 (44.4) |
| NRI | 4 (22.2) | 2 (11.1) | 8 (44.4) | 2 (11.1) | 3 (16.7) | 19 (21.1) |
| VAI | 1 (5.56) | 0 (0.0) | 2 (11.1) | 0 (0.0) | 1 (5.56) | 4 (4.4) |
| SCI | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
NRI – nerve root injury; VAI – vertebral artery injury; SCI – spinal cord injury.
The number of offset error nails of manual manipulation in its cross-sectional CT image.
| C3 (%) | C4 (%) | C5 (%) | C6 (%) | C7 (%) | Total (%) | |
|---|---|---|---|---|---|---|
| I | 4 (22.2) | 5 (27.8) | 8 (44.4) | 3 (16.7) | 5 (27.8) | 25 (27.8) |
| II | 7 (38.9) | 5 (27.8) | 8 (44.4) | 8 (44.4) | 10 (55.6) | 38 (42.2) |
| V | 12 (66.7) | 12 (66.7) | 10 (55.6) | 11 (61.1) | 9 (50.0) | 54 (60.0) |
I – vertical offset distance of the nail axis ≤1 mm and simultaneity an angular offset ≤5° from the pedicle axis; II – vertical offset distance of the nail axis >1mm from the pedicle axis; V – angle of the nail axis >5° with the pedicle axis.
Hitting accuracy of pedicle of two insertion techniques.
| HAP (%) | ASR (%) | RSR (%) | UR (%) | |
|---|---|---|---|---|
| Locator guidance | 90 | 60.49 | 17.28 | 22.22 |
| Manual manipulation | 55.6 | 25 | 25 | 50 |
HAP – hitting accuracy of pedicle; ASR – absolute stability rate of the hitting accuracy of the pedicle; RSR – relative stability rate of the hitting accuracy of the pedicle; UR – unstability rate of the hitting accuracy of the pedicle.