Literature DB >> 11034656

Cervical pedicle screws: comparative accuracy of two insertion techniques.

S C Ludwig1, J M Kowalski, C C Edwards, J G Heller.   

Abstract

STUDY
DESIGN: Independently assessed radiographic and anatomic comparison of device implantation methods.
OBJECTIVES: To compare the relative accuracy of two techniques of inserting cervical pedicle screws. SUMMARY OF BACKGROUND DATA: In an attempt to define the anatomic risks of cervical pedicle screw insertion, image-guided stereotactic technology was shown to be superior to some other methods in vitro.- Meanwhile, in vivo experience with Abumi's technique of screw insertion has had few clinically relevant instances of screw malposition. There has been no direct comparison between current image-guided technology and Abumi's fluoroscopically assisted technique.
METHODS: The pedicles (C3-C7) of human cadaveric cervical spines were instrumented with 3.5-mm screws with either of two techniques. Cortical integrity and potential neurovascular injury were independently assessed by computed tomographic (CT) scans and anatomic dissection. A cortical breach was considered "critical" if the screw encroached on any vital structure. If any part of the screw violated the cortex of the pedicle but no vital structure was at risk for injury, the breach was classified as "noncritical."
RESULTS: In Group I (StealthStation; Sofamor-Danek, Memphis, TN), 82% of screws were placed in the pedicle, and 18% had a critical breach. In Group II (Abumi technique), 88% of screws were placed in the pedicle, and 12% had a critical breach. No statistically significant differences were demonstrated between each group (P = 0.59). Regarding pedicle dimensions and safety of insertion, a critical pedicle diameter of 4.5 mm was determined to be the size below which a critical breach was likely, but above which there was a significantly greater likelihood for safe screw placement. The most common structure injured in each group was the vertebral artery.
CONCLUSIONS: The use of a computer-assisted image guidance system did not enhance safety or accuracy in placing pedicle screws compared with Abumi's technique. Both techniques have a noteworthy risk of injuring a critical structure if inserted into the pedicles with a diameter of less than 4.5 mm. Under laboratory conditions, pedicles with a diameter of more than 4.5 mm have a significantly greater likelihood of being safely instrumented by either technique. These data indicate that cervical pedicle screw placement is feasible, but it should be reserved for selected circumstances with clear indications and in the presence of suitable pedicle morphology.

Entities:  

Mesh:

Year:  2000        PMID: 11034656     DOI: 10.1097/00007632-200010150-00022

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  52 in total

1.  Computer-assisted posterior instrumentation of the cervical and cervico-thoracic spine.

Authors:  Marcus Richter; Thomas Mattes; Balkan Cakir
Journal:  Eur Spine J       Date:  2003-11-22       Impact factor: 3.134

2.  The use of pedicle screw-rod system for the posterior fixation in cervico-thoracic junction.

Authors:  Wonik Cho; Ahmed Shawky Eid; Ung-Kyu Chang
Journal:  J Korean Neurosurg Soc       Date:  2010-07-31

Review 3.  Pedicle screw insertion accuracy with different assisted methods: a systematic review and meta-analysis of comparative studies.

Authors:  Nai-Feng Tian; Qi-Shan Huang; Ping Zhou; Yang Zhou; Rui-Kai Wu; Yi Lou; Hua-Zi Xu
Journal:  Eur Spine J       Date:  2010-09-23       Impact factor: 3.134

4.  Pedicle screw fixation of the C7 vertebra using an anteroposterior fluoroscopic imaging technique.

Authors:  Shaunak Desai; Anil Sethi; Christopher C Ninh; Stephen Bartol; Rahul Vaidya
Journal:  Eur Spine J       Date:  2010-07-17       Impact factor: 3.134

5.  An evaluation of image-guided technologies in the placement of anterior thoracic vertebral body screws in spinal trauma: a cadaver study.

Authors:  Alexander R Vaccaro; Philip S Yuan; Harvey E Smith; Jonathon Hott; Rick Sasso; Stephen Papadopoulos
Journal:  J Spinal Cord Med       Date:  2005       Impact factor: 1.985

6.  Cervical pedicle screw placement: feasibility and accuracy of two new insertion techniques based on morphometric data.

Authors:  M Reinhold; F Magerl; M Rieger; M Blauth
Journal:  Eur Spine J       Date:  2006-04-21       Impact factor: 3.134

7.  CT evaluation of cervical pedicle in a Chinese population for surgical application of transpedicular screw placement.

Authors:  Zhu Ruofu; Yang Huilin; Hu Xiaoyun; He Xishun; Tang Tiansi; Chen Liang; Li Xigong
Journal:  Surg Radiol Anat       Date:  2008-03-20       Impact factor: 1.246

Review 8.  Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literature.

Authors:  M Reinhold; C Bach; L Audigé; R Bale; R Attal; M Blauth; F Magerl
Journal:  Eur Spine J       Date:  2008-01-22       Impact factor: 3.134

9.  Cervical anterior transpedicular screw fixation (ATPS)--Part II. Accuracy of manual insertion and pull-out strength of ATPS.

Authors:  Heiko Koller; Frank Acosta; Mark Tauber; Michael Fox; Hudelmaier Martin; Rosmarie Forstner; Peter Augat; Rainer Penzkofer; Christian Pirich; H Kässmann; Herbert Resch; Wolfgang Hitzl
Journal:  Eur Spine J       Date:  2008-01-26       Impact factor: 3.134

10.  Preoperative imaging of cervical pedicles: comparison of accuracy of oblique radiographs versus axial CT scans.

Authors:  James P Sieradzki; Eldin E Karaikovic; Eugene P Lautenschlager; Martin L Lazarus
Journal:  Eur Spine J       Date:  2008-07-26       Impact factor: 3.134

View more