Literature DB >> 11074671

Cervical spine pedicle screws: a biomechanical comparison of two insertion techniques.

J M Kowalski1, S C Ludwig, W C Hutton, J G Heller.   

Abstract

STUDY
DESIGN: Biomechanical testing of the pullout strengths of pedicle screws placed by two different techniques in adult human cadaveric cervical spines.
OBJECTIVES: To determine whether there is a significant difference in screw purchase of two commonly proposed methods of cervical pedicle screw insertion. SUMMARY OF BACKGROUND DATA: Wiring techniques remain the gold standard for posterior cervical fixation. However, absent or deficient posterior elements may dictate the use of alternative fixation techniques. Cervical pedicle screws have been shown to have significantly higher pullout strength than lateral mass screws.
METHODS: Fifty fresh disarticulated human vertebrae (C3-C7) were evaluated with computed tomography for anatomic disease and pedicle morphometry. The right and left pedicles were randomly assigned to either a standard method or the Abumi insertion method. In the latter technique the cortex and cancellous bone of lateral mass are removed with a high-speed burr, which provides a direct view of the pedicle introitus. The pedicle is then probed and tapped and a 3.5-mm cortical screw inserted. Each screw was subjected to a uniaxial load to failure.
RESULTS: There was no significant difference in the mean pullout resistance between the Abumi (696 N) and standard (636.5 N) insertion techniques (P = 0.41). There was no difference in pullout resistance between vertebral levels or within vertebral levels. Two (4%) minor pedicle wall violations were observed.
CONCLUSION: In selected circumstances pedicle screw instrumentation of the cervical spine may be used to manage complex deformities and patterns of instability. Surgeons need not be concerned about reduced screw purchase when deciding between the Abumi method and its alternatives.

Entities:  

Mesh:

Year:  2000        PMID: 11074671     DOI: 10.1097/00007632-200011150-00005

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  18 in total

1.  Factors affecting the interface of cervical spine facet screws placed in the technique by Roy-Camille et al.

Authors:  T R Pitzen; S Zenner; D Barbier; T Georg; W I Steudel
Journal:  Eur Spine J       Date:  2004-03-27       Impact factor: 3.134

2.  Cervical pedicle screw insertion using a gutter entry point at the transitional area between the lateral mass and lamina.

Authors:  Katsuhiro Tofuku; Hiroaki Koga; Setsuro Komiya
Journal:  Eur Spine J       Date:  2011-08-10       Impact factor: 3.134

3.  Cervical pedicle screw placement: feasibility and accuracy of two new insertion techniques based on morphometric data.

Authors:  M Reinhold; F Magerl; M Rieger; M Blauth
Journal:  Eur Spine J       Date:  2006-04-21       Impact factor: 3.134

Review 4.  Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literature.

Authors:  M Reinhold; C Bach; L Audigé; R Bale; R Attal; M Blauth; F Magerl
Journal:  Eur Spine J       Date:  2008-01-22       Impact factor: 3.134

5.  Cervical anterior transpedicular screw fixation (ATPS)--Part II. Accuracy of manual insertion and pull-out strength of ATPS.

Authors:  Heiko Koller; Frank Acosta; Mark Tauber; Michael Fox; Hudelmaier Martin; Rosmarie Forstner; Peter Augat; Rainer Penzkofer; Christian Pirich; H Kässmann; Herbert Resch; Wolfgang Hitzl
Journal:  Eur Spine J       Date:  2008-01-26       Impact factor: 3.134

6.  Clinical accuracy of cervical pedicle screw insertion using lateral fluoroscopy: a radiographic analysis of the learning curve.

Authors:  Hisashi Yoshimoto; Shigenobu Sato; Takahiko Hyakumachi; Yasushi Yanagibashi; Taiki Kanno; Takeshi Masuda
Journal:  Eur Spine J       Date:  2009-08-04       Impact factor: 3.134

7.  Cervical pedicle screw insertion using the technique with direct exposure of the pedicle by laminoforaminotomy.

Authors:  Dae-Jean Jo; Eun-Min Seo; Ki-Tack Kim; Sung-Min Kim; Sang-Hun Lee
Journal:  J Korean Neurosurg Soc       Date:  2012-11-30

8.  Percutaneous, Navigated Minimally Invasive Posterior Cervical Pedicle Screw Fixation.

Authors:  Domagoj Coric; Vincent J Rossi; John Peloza; Paul K Kim; Tim E Adamson
Journal:  Int J Spine Surg       Date:  2020-10-29

9.  Preoperative imaging of cervical pedicles: comparison of accuracy of oblique radiographs versus axial CT scans.

Authors:  James P Sieradzki; Eldin E Karaikovic; Eugene P Lautenschlager; Martin L Lazarus
Journal:  Eur Spine J       Date:  2008-07-26       Impact factor: 3.134

10.  Intra-osseous ultrasound for pedicle screw positioning in the subaxial cervical spine: an experimental study.

Authors:  Sven Rainer Kantelhardt; Hans Christoph Bock; Laila Siam; Jörg Larsen; Ralf Burger; Wolfgang Schillinger; Volker Bockermann; Veit Rohde; Alf Giese
Journal:  Acta Neurochir (Wien)       Date:  2009-07-14       Impact factor: 2.216

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.