Retrospective evaluation of exposure index (EI) values from plain radiographs reveals important considerations for quality improvement.

Ursula Mothiram1, Patrick C Brennan1, John Robinson1, Sarah J Lewis1, Bernadette Moran2.   

Abstract

INTRODUCTION: Following X-ray exposure, radiographers receive immediate feedback on detector exposure in the form of the exposure index (EI).
PURPOSE: To identify whether radiographers are meeting manufacturer-recommended EI (MREI) ranges for routine chest, abdomen and pelvis X-ray examinations under a variety of conditions and to examine factors affecting the EI.
METHODS: Data on 5000 adult X-ray examinations including the following variables were collected: examination parameters, EI values, patient gender, date of birth, date and time of examination, grid usage and the presence of implant or prosthesis. Descriptive statistics were used to summarize each data set and the Mann-Whitney U test was used to determine significant differences, with P < 0.05 indicating significance for all tests.
RESULTS: Most examinations demonstrated EI values that were outside the MREI ranges, with significantly higher median EI values recorded for female patient radiographs than those for male patients for all manufacturers, indicating higher detector exposures for all units except for Philips digital radiography (DR), where increased EI values indicate lower exposure (P = 0.01). Median EI values for out of hours radiography were also significantly higher compared with normal working hours for all technologies (P ≤ 0.02). Significantly higher median EI values were demonstrated for Philips DR chest X-rays without as compared to those with the employment of a grid (P = 0.03), while significantly lower median EI values were recorded for Carestream Health computed radiography (CR) chest X-rays when an implant or prosthesis was present (P = 0.02).
CONCLUSIONS: Non-adherence to MREIs has been demonstrated with EI value discrepancies being dependent on patient gender, time/day of exposure, grid usage and the presence of an implant or prosthesis. Retrospective evaluation of EI databases is a valuable tool to assess the need of quality improvement in routine DR.

Entities:  

Keywords:  Computed radiography (CR); digital radiography (DR); exposure index (EI); plain radiographs; quality improvement

Year:  2013        PMID: 26229619      PMCID: PMC4175822          DOI: 10.1002/jmrs.25

Source DB:  PubMed          Journal:  J Med Radiat Sci        ISSN: 2051-3895


Introduction

Digital radiography (DR) was first introduced almost 30 years ago and it is now considered standard technology in most radiology centres.1 The transition to digital technologies has brought about a much wider dynamic range of radiographic exposure parameters than that of conventional film screen.1,2 This means that a diagnostic image can be generated over a much greater range of entrance doses,1 without an adverse effect on image quality.3–5 There is, however, a need to ensure that with an increased dynamic range, patient radiation doses are not excessive; therefore, an indication of exposure is required. With the current digital radiographic technology, this comes in the form of an exposure indicator that specifically offers feedback on the radiation level reaching the imaging detector.2,6 However, it is important to note that the exposure index (EI) is not an actual measure of patient dose.7 Currently, there are a variety of manufacturer-dependent EIs, which presents the opportunity for user confusion. To counter this, an international standardized EI has been developed simultaneously by the International Electrotechnical Commission (IEC) and the American Association of Physicists in Medicine (AAPM) in conjunction with the DR system manufacturers.8 The standardized EI has been designed to generate a linear relationship between the detector exposure and index value.8,9 This study demonstrates the usefulness of careful examination of EI data, indicating for example the influence of patient type, technology and radiographic practice on EI values. It examines the EIs recorded for routine adult X-ray examinations of the chest, abdomen and pelvis to assess adherence to manufacturers' EI recommendations and it also explores how the EI values are affected by parameters such as patient gender, time or date of exposure, the use of secondary radiation grid and the presence of implant or prosthesis.

Materials and Methods

Equipment

The study was performed in one of the largest teaching hospitals in Ireland that operates in a full picture archiving and communication system (PACS) environment. The authors have chosen to keep the hospital name anonymous to ensure the anonymity of the radiographers working there. Chest, abdomen and pelvis examination details were extracted following image acquisition using Siemens Medical VX and MX Axiom Aristos DR, a Philips DR DigitalDiagnost system version 1.4 and a Carestream Health Direct View CR500 computed radiography (CR). The Philips and Siemens DR systems utilize flat panel detector technologies. Details of the grids used are (in the order grid ratio [R], line frequency [N] and focal distance [F]): Siemens: R = 15:1, N = 80 lines/cm, F = 115 cm; R = 15:1, N = 80 lines/cm, F = 180 cm Philips: R = 8:1, N = 36 lines/cm, F = 110 cm; R = 8:1, N = 36 lines/cm, F = 140 cm The allocation of specific examinations to specific technologies is demonstrated in Tables 2–4. All the pieces of equipment included in this study have been calibrated according to manufacturers' recommendations. In addition, for both Philips and Siemens equipment, the radiographers are also periodically prompted to perform a system calibration by the system software. The Carestream Health CR plates are calibrated at 6-monthly intervals during scheduled maintenance. In addition, the CR readers are calibrated at 6-monthly intervals and matched. The systems are also closely monitored by hospital medical physics staff.
Table 2

Descriptive statistics of exposure index value distributions for the manufacturer/examinations included in this study (for manufacturer-recommended exposure index value ranges see Table 1).

Exposure

Manufacturer/examinationNumber (n)MedianFirst quartile (Q1)Third quartile (Q3)% Over% Under
Siemens DR
 Chest posterior–anterior245322819526716
 Chest lateral107319278360131
 Chest supine1522619428900
 Chest lateral decubitus2193142243050
 Chest anterior–posterior4249146270025
 Abdomen37723320525601
 Pelvis239467369581670
Philips DR
 Chest posterior–anterior21340032040001
 Chest lateral1540032040000
 Chest anterior–posterior94002506301111
 Abdomen2663063063000
 Pelvis363050063000
Carestream Health CR
 Mobile chests1261204019302170793
 Departmental chests273202018802130724
 Abdomen11400140014000100
 Pelvis2165015101790050
Total5000284

DR, digital radiography; CR, computed radiography.

Table 4

Exposure index values for in hours and out of hours examinations (for manufacturer-recommended exposure index value ranges see Table 1).

Exposure

Manufacturer/examinationIn hours (I)/out of hours (O)Number (n)MedianFirst quartile (Q1)Third quartile (Q3)% Over% UnderP-value
Siemens DR
ChestI155322419225816<0.0001
O102824520428725
AbdomenI21022419425102<0.0001
O16724121526100
 PelvisI194469376568690ns
O45428343615620
Philips DR
ChestI52400320400000.01
O18540040040011
 AbdomenI763063063000ns
O1963063063000
 PelvisI263063063000nf
O150050050000
Carestream Health CR
Mobile chestsI5962020191021507730.0005
O665206019402190823
Departmental chestsI1391970184021206650.02
O134205019102140783
 AbdomenI11400140014000100nf
O000
 PelvisI11510151015100100nf
O117901790179000
Total5000

Rows with values in bold indicate significant differences in median exposure index values (P < 0.05), where medians were equal the mean was used to determine significance. nf, numbers too few; ns, not significant; DR, digital radiography; CR, computed radiography.

The manufacturer-recommended exposure index value ranges. DR, digital radiography; CR, computed radiography; MREI, manufacturer-recommended exposure index; EXI, exposure index; EI, exposure index. [Correction added on 27 November 2013, after first online publication: In the first row of Table 1, ‘EXI’ has been corrected from ‘EI’, and the corresponding legend ‘EXI, exposure index;’ has been inserted.]
Table 1

The manufacturer-recommended exposure index value ranges.

Manufacturer/modelMREI rangeExposure indicator
Siemens Medical VX and MX Axiom Aristos DR (Erlangen, Germany)150–400 (Schramm H, 2012, personal communication)EXI6
Philips DR DigitalDiagnost system version 1.4 (Hamburg, Germany)250–630 (Neitzel U, 2012, personal communication)EI6
Carestream Health Direct View CR500 CR (Rochester, New York)1700–190012,13EI6

DR, digital radiography; CR, computed radiography; MREI, manufacturer-recommended exposure index; EXI, exposure index; EI, exposure index. [Correction added on 27 November 2013, after first online publication: In the first row of Table 1, ‘EXI’ has been corrected from ‘EI’, and the corresponding legend ‘EXI, exposure index;’ has been inserted.]

Descriptive statistics of exposure index value distributions for the manufacturer/examinations included in this study (for manufacturer-recommended exposure index value ranges see Table 1). DR, digital radiography; CR, computed radiography. Exposure index values for each patient gender and examination type included in this study (for manufacturer-recommended exposure index value ranges see Table 1). Rows with values in bold indicate significant differences in median exposure index values (P < 0.05), where medians were equal the mean was used to determine significance. M, male; F, female; nf, numbers too few; ns, not significant; DR, digital radiography; CR, computed radiography. Exposure index values for in hours and out of hours examinations (for manufacturer-recommended exposure index value ranges see Table 1). Rows with values in bold indicate significant differences in median exposure index values (P < 0.05), where medians were equal the mean was used to determine significance. nf, numbers too few; ns, not significant; DR, digital radiography; CR, computed radiography.

Data collection

The study was conducted from November 2010 to August 2011. EIs were retrospectively recorded on a total of 5000 routine examinations performed over a 35-day period between October and November 2010 in a busy teaching hospital. Recorded information consists of chest (n = 4352), abdomen (n = 404) and pelvis (n = 244) X-ray examinations. The examination details for each patient were retrieved from the hospital PACS. The specific examination types are shown in Tables 2–4. Data were gathered on adult male (2622) and female (2378) patients and individuals' details were anonymized to protect the privacy of patients. Ethics approval was granted by the Human Research Ethics Committee of the University of Sydney and the Ethics Committee and Risk and Legal departments within the hospital in which the study was based. In addition to the EI, the following details were recorded for each examination: patient gender, time and date of exposure, and the use of a secondary radiation grid. When an image demonstrated the presence of an implant or prosthesis, this was also recorded as the literature has indicated this can affect EI values.10 Only initial presentation examinations were recorded as additional specialized projections may not be indicative of normal practice. The data on the date and the time of taking the radiographic exposures were used to categorize the radiographs into two: those taken ‘in hours’ and ‘out of hours’. Normal working hours (or in hours) were established as 9 a.m. to 5 p.m. Monday to Friday and out of hours included public holidays, weekends and from 5 p.m. to 9 a.m. Monday to Friday.

Data analysis

The data were analysed in several ways. GraphPad Prism 5 software (La Jolla, CA) was used to perform statistical analysis. Descriptive statistics were used to summarize the characteristics of each data set relating to specific manufacturer and examination type.11 Exposure indices for each examination were compared to manufacturer-recommended EIs (MREIs) for specific equipment (see Table 1) and the percentage of values not adhering to recommendations was calculated. As the conditions for normality were not met the Mann–Whitney U test was then used to determine significant differences between EI values, for each examination type, for the following paired comparisons: male versus female patients; in hours versus out of hours; use versus non-use of secondary radiation grid (this was only possible for Philips DR chest examinations); presence versus absence of an implant or prosthesis (examples of implants and prostheses included pacemakers, orthopaedic joint replacements, pins and plates). A P-value of <0.05 was used to describe significant differences.

Results

Readers are reminded that for the Siemens and Carestream technologies higher EI values indicate a higher radiation exposure compared to lower EIs, whereas the reverse is relevant for the Philips equipment, where higher EI values indicate lower radiation exposure. MREI value ranges for Philips and Siemens DR systems are based on them operating at a sensitivity of 400.

Descriptive statistics

EI data relating to the examinations included in this study are outlined in Table 2. Most examinations demonstrated that EIs for at least some exposures fell outside the MREI, the exceptions being the following examinations: Siemens supine chest; Philips lateral chest; Philips abdomen; Philips pelvis (see Table 2). Overall 28% of examinations demonstrated EI values above and 4% below the MREI: here, Philips values less than 250 were included in the overexposure category and values greater than 630 were included in the underexposure category.

Statistically significant results

Patient gender

EI variations were evident between the patient genders (see Table 3), with female patients recording significantly higher median EI values than male patients for all the manufacturers for the following examinations: Siemens: chest posterior–anterior (PA) (P < 0.0001), lateral (P = 0.04) and supine examinations (P = 0.03); Carestream Health technology departmental chests (P = 0.001); Philips DR PA chest examinations (P = 0.01). Conversely male patients recorded higher values than female patients for abdomen examinations (P = 0.001) with the Siemens technology.
Table 3

Exposure index values for each patient gender and examination type included in this study (for manufacturer-recommended exposure index value ranges see Table 1).

Exposure

Manufacturer/examinationGenderNumber (n)MedianFirst quartile (Q1)Third quartile (Q3)% Over% UnderP-value
Siemens DR
Chest posterior–anteriorF120723920927513<0.0001
M124621618525518
Chest lateralF483322854092900.04
M5931326834322
Chest supineF8263218306000.03
M720118022700
 Chest lateral decubitusF2193142243050nf
M000
 Chest anterior–posteriorF226926727100nf
M2174118230050
AbdomenF204227199252020.001
M17324021226500
 PelvisF157479384596700ns
M82447357526620
Philips DR
Chest posterior–anteriorF90400400400000.01
M12340032040001
 Chest lateralF340040040000ns
M1240032047500
 Chest anterior–posteriorF5320225515200ns
M4630345758025
 AbdomenF1763063063000ns
M963063063000
 PelvisF256550063000nf
M163063063000
Carestream Health CR
 Mobile chestsF483206019302190813ns
M778203019202160783
Departmental chestsF1482055191021607730.001
M125197018552080665
 AbdomenF000nf
M11400140014000100
 PelvisF2165015101790050nf
M000
Total5000

Rows with values in bold indicate significant differences in median exposure index values (P < 0.05), where medians were equal the mean was used to determine significance. M, male; F, female; nf, numbers too few; ns, not significant; DR, digital radiography; CR, computed radiography.

In hours and out of hours

Median EI values out of hours compared to within hours were significantly higher for all technologies: Siemens: chest (P < 0.0001) and abdomen examinations (P < 0.0001); Carestream: mobile (P = 0.0005) and departmental (P = 0.02) chest examinations; Philips DR chest X-ray examinations (P = 0.01) (see Table 4).

Secondary radiation grid

Significantly higher median EI values were demonstrated for Philips DR chest X-rays without grid use (P = 0.03) compared to when a grid was employed (see Table 5).
Table 5

Exposure index values for Philips digital radiography chest X-rays and grid usage (for manufacturer-recommended exposure index value ranges see Table 1).

Exposure

Philips chest DRNumber (n)MedianFirst quartile (Q1)Third quartile (Q3)% Over% UnderP-value
Grid in22540032040000.40.03
Grid out124502886308.38.3
Total237

Rows with values in bold indicate significant differences in median exposure index values (P < 0.05). DR, digital radiography.

Exposure index values for Philips digital radiography chest X-rays and grid usage (for manufacturer-recommended exposure index value ranges see Table 1). Rows with values in bold indicate significant differences in median exposure index values (P < 0.05). DR, digital radiography.

Presence of implant or prosthesis

Lower median EI values were identified with Carestream chest X-rays when an implant or prosthesis was present compared to when an implant or prosthesis was absent (P = 0.02).

Discussion

The purpose of this study was to identify whether radiographers are meeting the MREI ranges for routine chest, abdomen and pelvis examinations and to examine if patient gender, time and day of exposure, use of secondary radiation grid and presence of an implant or prosthesis impacted significantly upon EI values. Most examinations demonstrated EI values that were outside the MREI ranges. Carestream Health CR mobile and departmental chest X-rays demonstrated the highest overexposure rates of 79% and 72%, respectively, followed by Siemens DR pelvis X-ray examinations, with an overexposure rate of 67%. This apparent overexposure is largely in agreement with the literature, which indicates that transition from conventional film screen radiography to CR or DR can result in an increase in radiation dose to the patient.5,6 The lowest underexposure rates were demonstrated for Carestream Health CR abdomen and pelvis X-rays and Siemens DR lateral decubitus chest X-rays; however, due to the small sample sizes in these three categories these results should be viewed with caution. The high rate of overexposures for Carestream Health CR chest X-rays is a concern as it is one of the most frequently performed radiographic examinations,14 particularly because clinically acceptable images at EIs lower than the MREI have been shown previously.13,15 The overexposures attributed to Carestream Health CR mobile chest X-rays may be linked to the complexities associated with mobile chest X-rays where patient position, X-ray source image receptor distances (SID) and exposures are much more variable than in the departmental situation.16 There may also be increased pressure on radiographers not wanting to have to repeat the examination due to low-dose-related image mottle so that ward staff exposures and inconvenience are minimized. The recording of previous examination exposure factors with mobile chest radiography has been suggested to assist in reducing patients' radiation doses with film screen technologies.17 Here, a clear link between shorter focus to skin distances (FSD) and higher entrance skin doses (ESD) was demonstrated, however, when exposure factors were optimized for different FSDs and when FSDs were measured for each examination dose reductions were achieved.17 The potential to lower EIs through optimizing exposures for various FSDs using CR should be investigated. However, more difficult to explain is the excessive exposures evident with the department-based exposures and this requires further investigation. It is important to note that the CR examinations included in this study were not performed in conjunction with an automatic exposure control (AEC). Although observing non-compliant exposures is a useful quality assurance task, it is probably of greater value to identify factors relating to EI variations and in this study we identified four key agents – patient gender, time/date of exposure, grid usage and the presence of an implant or prosthesis. In terms of patient gender, for most examinations irrespective of the manufacturer, radiographs on female patients demonstrated higher EI values (higher exposures) when compared with male patients. For the Siemens and Carestream technology, this was statistically significant in four instances for female patients compared with only one example demonstrating significantly higher values with male patients. There is no clear justification for this higher exposure pattern in female patients. A previous study has suggested that inadequate exposure charts for female patients could result in higher EI values and recommended optimization of exposure charts to reduce detector dose.18 Exposure chart optimization could counteract this gender bias. It has be noted in the literature that variations in male and female doses are relevant for the monitoring of AEC systems: the smaller the variations in mean doses detected the more sensitive the AEC is to changes in the attenuation characteristics of different patients groups.19 Although referring to actual radiation dose, this would be worth investigating, as AEC performance could be a factor in the higher female EI values demonstrated in this study using DR. In this study, all the Siemens DR examinations were performed using the AEC. It should be noted that while chest PA examinations with the Philips technology also exhibited higher EI values for female patients compared with male patients, this actually indicates a lower exposure. Out of hours exposures were also associated with significantly higher EIs for a range of examinations, which has been previously reported and may be linked to reduced on-call staff numbers coupled with a strong reluctance among radiographers to have to repeat an exposure,15 when working a busy shift out of hours. The higher EIs could be related to staffing and the level of experience of radiographers performing shift work. Whatever the reason, this is an unacceptable inconsistency that needs to be addressed and highlights the importance of regular monitoring of EI values within departments to highlight when excessive out of hours exposures take place. A significantly higher median EI value (lower exposure) was demonstrated for Philips DR chest X-rays without a grid. It may be argued that with non-usage of a grid, the exposure levels would be less, but this ignores the fact that the EI is based on pixel values, so these should remain relatively constant regardless of whether a grid has been used. With the DigitalDiagnost system the automatic exposure device (AED) can remain in position when the grid is removed, thus allowing for phototimed exposures to be made with grid in or out.20 If AEDs are being used, and because these are positioned on the detector side of the grid,21 proper calibration and operation need to be confirmed.22 It is also possible that radiographers were using adjustments when using the AEC, which could account for the EI values seen here. Variations in the SIDs applied in conjunction with grid usage in this instance could also have had a bearing on the EIs recorded here. The reduction in EI values with Carestream Health CR chest examinations in the presence of an implant or prostheses is also notable and highlights that radiographers need to understand the basis on which EI values are calculated before judgements can be made on exposures. In this circumstance, it is possible that the lower EI values reflect the fact that if the prosthesis was within the clinical region of interest from which average pixel values were calculated. Naturally, the reduction in X-rays reaching the detector due to the prosthesis would reflect in lower pixel values and thus lower EIs, and if so, this clearly has little to do with improper technique. The purpose of this study was to retrospectively analyse EI values and it does not set out to compare the different technologies or manufacturers included in the study nor does it attempt to make a reference to patient dose. As this was a retrospective study it was not possible to analyse patient's size and body part thickness, which is a limitation of this study. Data on these variables could provide useful information relating to the EI gender variations demonstrated here, as male patients and female patients have variation in tissue distribution.

Conclusion

In summary, this study demonstrates a series of EI values that do not adhere to recommended values. However, to identify which exposures are inappropriate requires that manufacturer levels of EIs be optimized. However, it can be stated with confidence that the variations associated with patient gender, the time or date of exposure, and the use or non-use of a secondary radiation grid are very difficult to justify and corrective strategies should be explored. This study highlights the value of auditing EIs to assist with image quality improvement and ultimately to aid in patient dose reduction.
  17 in total

1.  Managing patient dose in digital radiology. A report of the International Commission on Radiological Protection.

Authors: 
Journal:  Ann ICRP       Date:  2004

2.  Transition from screen-film to digital radiography: evolution of patient radiation doses at projection radiography.

Authors:  Eliseo Vaño; José Miguel Fernández; José Ignacio Ten; Carlos Prieto; Luciano González; Ricardo Rodríguez; Hugo de las Heras
Journal:  Radiology       Date:  2007-03-13       Impact factor: 11.105

3.  The standardized exposure index for digital radiography: an opportunity for optimization of radiation dose to the pediatric population.

Authors:  J Anthony Seibert; Richard L Morin
Journal:  Pediatr Radiol       Date:  2011-04-14

Review 4.  Dose and perceived image quality in chest radiography.

Authors:  Wouter J H Veldkamp; Lucia J M Kroft; Jacob Geleijns
Journal:  Eur J Radiol       Date:  2009-07-03       Impact factor: 3.528

Review 5.  DR and CR: Recent advances in technology.

Authors:  C M Schaefer-Prokop; D W De Boo; M Uffmann; M Prokop
Journal:  Eur J Radiol       Date:  2009-08-19       Impact factor: 3.528

6.  New exposure indicators for digital radiography simplified for radiologists and technologists.

Authors:  Steven Don; Bruce R Whiting; Lois Jo Rutz; Bruce K Apgar
Journal:  AJR Am J Roentgenol       Date:  2012-12       Impact factor: 3.959

7.  Quality assurance: using the exposure index and the deviation index to monitor radiation exposure for portable chest radiographs in neonates.

Authors:  Mervyn D Cohen; Matt L Cooper; Kelly Piersall; Bruce K Apgar
Journal:  Pediatr Radiol       Date:  2010-12-30

8.  Are exposure index values consistent in clinical practice? A multi-manufacturer investigation.

Authors:  M L Butler; L Rainford; J Last; P C Brennan
Journal:  Radiat Prot Dosimetry       Date:  2010-03-11       Impact factor: 0.972

9.  Digital radiography: are the manufacturers' settings too high? Optimisation of the Kodak digital radiography system with aid of the computed radiography dose index.

Authors:  Sinead E Peters; Patrick C Brennan
Journal:  Eur Radiol       Date:  2001-12-13       Impact factor: 5.315

10.  Patient dose management in digital radiography.

Authors:  E Vano; Jm Fernandez Soto
Journal:  Biomed Imaging Interv J       Date:  2007-04-01
View more
  2 in total

1.  Portable abdomen radiography: moving to thickness-based protocols.

Authors:  Adrian A Sánchez; Ingrid Reiser; Tina Baxter; Yue Zhang; Joshua H Finkle; Zheng Feng Lu; Kate A Feinstein
Journal:  Pediatr Radiol       Date:  2017-11-12

Review 2.  Digital radiography exposure indices: A review.

Authors:  Ursula Mothiram; Patrick C Brennan; Sarah J Lewis; Bernadette Moran; John Robinson
Journal:  J Med Radiat Sci       Date:  2014-05-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.