Ursula Mothiram1, Patrick C Brennan1, John Robinson1, Sarah J Lewis1, Bernadette Moran2.
Abstract
INTRODUCTION: Following X-ray exposure, radiographers receive immediate feedback on detector exposure in the form of the exposure index (EI).Entities:
Keywords: Computed radiography (CR); digital radiography (DR); exposure index (EI); plain radiographs; quality improvement
Year: 2013 PMID: 26229619 PMCID: PMC4175822 DOI: 10.1002/jmrs.25
Source DB: PubMed Journal: J Med Radiat Sci ISSN: 2051-3895
Descriptive statistics of exposure index value distributions for the manufacturer/examinations included in this study (for manufacturer-recommended exposure index value ranges see Table 1).
| Exposure | ||||||
|---|---|---|---|---|---|---|
| Manufacturer/examination | Number ( | Median | First quartile ( | Third quartile ( | % Over | % Under |
| Siemens DR | ||||||
| Chest posterior–anterior | 2453 | 228 | 195 | 267 | 1 | 6 |
| Chest lateral | 107 | 319 | 278 | 360 | 13 | 1 |
| Chest supine | 15 | 226 | 194 | 289 | 0 | 0 |
| Chest lateral decubitus | 2 | 193 | 142 | 243 | 0 | 50 |
| Chest anterior–posterior | 4 | 249 | 146 | 270 | 0 | 25 |
| Abdomen | 377 | 233 | 205 | 256 | 0 | 1 |
| Pelvis | 239 | 467 | 369 | 581 | 67 | 0 |
| Philips DR | ||||||
| Chest posterior–anterior | 213 | 400 | 320 | 400 | 0 | 1 |
| Chest lateral | 15 | 400 | 320 | 400 | 0 | 0 |
| Chest anterior–posterior | 9 | 400 | 250 | 630 | 11 | 11 |
| Abdomen | 26 | 630 | 630 | 630 | 0 | 0 |
| Pelvis | 3 | 630 | 500 | 630 | 0 | 0 |
| Carestream Health CR | ||||||
| Mobile chests | 1261 | 2040 | 1930 | 2170 | 79 | 3 |
| Departmental chests | 273 | 2020 | 1880 | 2130 | 72 | 4 |
| Abdomen | 1 | 1400 | 1400 | 1400 | 0 | 100 |
| Pelvis | 2 | 1650 | 1510 | 1790 | 0 | 50 |
| Total | 5000 | 28 | 4 | |||
DR, digital radiography; CR, computed radiography.
Exposure index values for in hours and out of hours examinations (for manufacturer-recommended exposure index value ranges see Table 1).
| Exposure | ||||||||
|---|---|---|---|---|---|---|---|---|
| Manufacturer/examination | In hours (I)/out of hours (O) | Number ( | Median | First quartile ( | Third quartile ( | % Over | % Under | |
| Siemens DR | ||||||||
| | ||||||||
| | ||||||||
| Pelvis | I | 194 | 469 | 376 | 568 | 69 | 0 | ns |
| O | 45 | 428 | 343 | 615 | 62 | 0 | ||
| Philips DR | ||||||||
| | ||||||||
| Abdomen | I | 7 | 630 | 630 | 630 | 0 | 0 | ns |
| O | 19 | 630 | 630 | 630 | 0 | 0 | ||
| Pelvis | I | 2 | 630 | 630 | 630 | 0 | 0 | nf |
| O | 1 | 500 | 500 | 500 | 0 | 0 | ||
| Carestream Health CR | ||||||||
| | ||||||||
| | ||||||||
| Abdomen | I | 1 | 1400 | 1400 | 1400 | 0 | 100 | nf |
| O | 0 | 0 | 0 | |||||
| Pelvis | I | 1 | 1510 | 1510 | 1510 | 0 | 100 | nf |
| O | 1 | 1790 | 1790 | 1790 | 0 | 0 | ||
| Total | 5000 | |||||||
Rows with values in bold indicate significant differences in median exposure index values (P < 0.05), where medians were equal the mean was used to determine significance. nf, numbers too few; ns, not significant; DR, digital radiography; CR, computed radiography.
The manufacturer-recommended exposure index value ranges.
| Manufacturer/model | MREI range | Exposure indicator |
|---|---|---|
| Siemens Medical VX and MX Axiom Aristos DR (Erlangen, Germany) | 150–400 (Schramm H, 2012, personal communication) | EXI |
| Philips DR DigitalDiagnost system version 1.4 (Hamburg, Germany) | 250–630 (Neitzel U, 2012, personal communication) | EI |
| Carestream Health Direct View CR500 CR (Rochester, New York) | 1700–1900 | EI |
DR, digital radiography; CR, computed radiography; MREI, manufacturer-recommended exposure index; EXI, exposure index; EI, exposure index. [Correction added on 27 November 2013, after first online publication: In the first row of Table 1, ‘EXI’ has been corrected from ‘EI’, and the corresponding legend ‘EXI, exposure index;’ has been inserted.]
Exposure index values for each patient gender and examination type included in this study (for manufacturer-recommended exposure index value ranges see Table 1).
| Exposure | ||||||||
|---|---|---|---|---|---|---|---|---|
| Manufacturer/examination | Gender | Number ( | Median | First quartile ( | Third quartile ( | % Over | % Under | |
| Siemens DR | ||||||||
| | ||||||||
| | ||||||||
| | ||||||||
| Chest lateral decubitus | F | 2 | 193 | 142 | 243 | 0 | 50 | nf |
| M | 0 | 0 | 0 | |||||
| Chest anterior–posterior | F | 2 | 269 | 267 | 271 | 0 | 0 | nf |
| M | 2 | 174 | 118 | 230 | 0 | 50 | ||
| | ||||||||
| Pelvis | F | 157 | 479 | 384 | 596 | 70 | 0 | ns |
| M | 82 | 447 | 357 | 526 | 62 | 0 | ||
| Philips DR | ||||||||
| | ||||||||
| Chest lateral | F | 3 | 400 | 400 | 400 | 0 | 0 | ns |
| M | 12 | 400 | 320 | 475 | 0 | 0 | ||
| Chest anterior–posterior | F | 5 | 320 | 225 | 515 | 20 | 0 | ns |
| M | 4 | 630 | 345 | 758 | 0 | 25 | ||
| Abdomen | F | 17 | 630 | 630 | 630 | 0 | 0 | ns |
| M | 9 | 630 | 630 | 630 | 0 | 0 | ||
| Pelvis | F | 2 | 565 | 500 | 630 | 0 | 0 | nf |
| M | 1 | 630 | 630 | 630 | 0 | 0 | ||
| Carestream Health CR | ||||||||
| Mobile chests | F | 483 | 2060 | 1930 | 2190 | 81 | 3 | ns |
| M | 778 | 2030 | 1920 | 2160 | 78 | 3 | ||
| | ||||||||
| Abdomen | F | 0 | 0 | 0 | nf | |||
| M | 1 | 1400 | 1400 | 1400 | 0 | 100 | ||
| Pelvis | F | 2 | 1650 | 1510 | 1790 | 0 | 50 | nf |
| M | 0 | 0 | 0 | |||||
| Total | 5000 | |||||||
Rows with values in bold indicate significant differences in median exposure index values (P < 0.05), where medians were equal the mean was used to determine significance. M, male; F, female; nf, numbers too few; ns, not significant; DR, digital radiography; CR, computed radiography.
Exposure index values for Philips digital radiography chest X-rays and grid usage (for manufacturer-recommended exposure index value ranges see Table 1).
| Exposure | |||||||
|---|---|---|---|---|---|---|---|
| Philips chest DR | Number ( | Median | First quartile ( | Third quartile ( | % Over | % Under | |
| Total | 237 | ||||||
Rows with values in bold indicate significant differences in median exposure index values (P < 0.05). DR, digital radiography.