Literature DB >> 26208607

Otto Aufranc Award: Large Heads Do Not Increase Damage at the Head-neck Taper of Metal-on-polyethylene Total Hip Arthroplasties.

Georgios K Triantafyllopoulos1, Marcella E Elpers1, Jayme C Burket1, Christina I Esposito1, Douglas E Padgett1, Timothy M Wright2.   

Abstract

BACKGROUND: Fretting and corrosion at head-neck junctions of total hip arthroplasties (THAs) have been associated with adverse local tissue reactions in patients with both metal-on-polyethylene (MoP) and metal-on-metal (MoM) prostheses. Femoral head size contributes to the severity of fretting and corrosion in large-diameter MoM THAs, but its impact on such damage in MoP THAs remains unknown. QUESTIONS/PURPOSES: (1) Is femoral head size associated with increased fretting or corrosion at the head-neck junction in MoP total hips? (2) Is duration of implantation associated with increased fretting or corrosion?
METHODS: The severity of fretting/corrosion on surfaces of head tapers and stem trunnions was visually examined in 154 MoP THAs retrieved as part of 3282 revision surgeries performed at our institution between January 1, 2007, and December 31, 2013. Fretting and corrosion damage were subjectively graded by two independent observers on a 1 to 4 scale, and their relations to head size, alloy combinations, taper/trunnion design, length of implantation (LOI), and location were investigated. Differences in scores never exceeded one grade, and this occurred in only 17% of examined implants. With the available implants, the study provided 88% power to detect differences of 0.5 in fretting or corrosion scores in these analyses.
RESULTS: Fretting and corrosion of the tapers and the trunnions were not affected by head size (p = 0.247, p = 0.471, p = 0.837, and p = 0.868, respectively), although taper/trunnion design affected taper fretting (p = 0.005) and corrosion (p = 0.0031) and trunnion fretting (p = 0.0028). Head taper fretting (observed in 73% of heads) increased with LOI, but head taper corrosion (noted in 93% of heads) was not affected. Trunnion fretting (observed in 86% of stems) was more severe in mixed-alloy combinations and with increased LOI and was more severe proximally. Trunnion corrosion (noted in 72% of stems) was also location-dependent with greater corrosion distally.
CONCLUSIONS: Fretting and corrosion are regular occurrences in MoP THAs, but neither damage type was related to femoral head size. Conversely, taper design, LOI, and alloy combination affected the severity of both fretting and corrosion. CLINICAL RELEVANCE: Although it has been suggested that trunnion corrosion seen in MoP bearings is a function of larger diameter heads, our data suggest that larger femoral heads may be used for increased damage at the modular junction of MoP THAs.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26208607      PMCID: PMC4709304          DOI: 10.1007/s11999-015-4468-6

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  33 in total

1.  Life expectancy of modular Ti6Al4V hip implants: influence of stress and environment.

Authors:  A Chandra; J J Ryu; P Karra; P Shrotriya; V Tvergaard; M Gaisser; T Weik
Journal:  J Mech Behav Biomed Mater       Date:  2011-07-03

2.  Fretting crevice corrosion of stainless steel stem-CoCr femoral head connections: comparisons of materials, initial moisture, and offset length.

Authors:  Jeremy L Gilbert; Manav Mehta; Bryan Pinder
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2009-01       Impact factor: 3.368

3.  Fretting corrosion of CoCrMo and Ti6Al4V interfaces.

Authors:  Viswanathan Swaminathan; Jeremy L Gilbert
Journal:  Biomaterials       Date:  2012-05-09       Impact factor: 12.479

4.  Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses.

Authors:  Annelie Rehmer; Nicholas E Bishop; Michael M Morlock
Journal:  Clin Biomech (Bristol, Avon)       Date:  2011-09-08       Impact factor: 2.063

Review 5.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

6.  Potential and frequency effects on fretting corrosion of Ti6Al4V and CoCrMo surfaces.

Authors:  Viswanathan Swaminathan; Jeremy L Gilbert
Journal:  J Biomed Mater Res A       Date:  2013-02-13       Impact factor: 4.396

7.  Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings.

Authors:  Nicholas Bishop; Florian Witt; Robin Pourzal; Alfons Fischer; Marcel Rütschi; Markus Michel; Michael Morlock
Journal:  J Orthop Res       Date:  2013-02-25       Impact factor: 3.494

8.  An analysis of the head-neck taper interface in retrieved hip prostheses.

Authors:  J R Lieberman; C M Rimnac; K L Garvin; R W Klein; E A Salvati
Journal:  Clin Orthop Relat Res       Date:  1994-03       Impact factor: 4.176

9.  Do the potential benefits of metal-on-metal hip resurfacing justify the increased cost and risk of complications?

Authors:  Kevin J Bozic; Christine M Pui; Matthew J Ludeman; Thomas P Vail; Marc D Silverstein
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

10.  Enhanced wear and corrosion in modular tapers in total hip replacement is associated with the contact area and surface topography.

Authors:  Anna Panagiotidou; Jay Meswania; Jia Hua; Sarah Muirhead-Allwood; Alister Hart; Gordon Blunn
Journal:  J Orthop Res       Date:  2013-08-21       Impact factor: 3.494

View more
  19 in total

1.  Do Well-functioning THAs Retrieved at Autopsy Exhibit Evidence of Fretting and Corrosion?

Authors:  Jeffrey Lange; Amanda Wach; Chelsea N Koch; Robert H Hopper; Henry Ho; Charles A Engh; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-10       Impact factor: 4.176

2.  How Useful Is Magnetic Resonance Imaging in Evaluating Adverse Local Tissue Reaction?

Authors:  Douglas E Padgett; Edwin P Su; Timothy M Wright; Alissa J Burge; Hollis G Potter
Journal:  J Arthroplasty       Date:  2020-01-15       Impact factor: 4.757

3.  Spontaneous Fractures of a Modern Modular Uncemented Femoral Stem.

Authors:  Chelsea N Koch; Laura Serrano Mateo; Stephen Kayiaros; Kevin A Cassidy; Seth A Jerabek; Alejandro Gonzalez Della Valle
Journal:  HSS J       Date:  2016-06-21

4.  Salvage of a monoblock metal-on-metal cup using a dual mobility liner: a two-year MRI follow-up study.

Authors:  Maximilian F Kasparek; Lisa Renner; Martin Faschingbauer; Wenzel Waldstein; Kilian Rueckl; Friedrich Boettner
Journal:  Int Orthop       Date:  2017-09-19       Impact factor: 3.075

Review 5.  Metal-on-metal total hip arthroplasty: risk factors for pseudotumours and clinical systematic evaluation.

Authors:  Ming Han Lincoln Liow; Young-Min Kwon
Journal:  Int Orthop       Date:  2016-10-20       Impact factor: 3.075

6.  Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: a Retrieval Analysis.

Authors:  Chelsea N Koch; Mark Figgie; Mark P Figgie; Marcella E Elpers; Timothy M Wright; Douglas E Padgett
Journal:  HSS J       Date:  2017-06-22

7.  Low failure rate at short term for 40 mm heads and second generation triple annealed HCLPE liners in hybrid hip replacements.

Authors:  Rajkumar Thangaraj; Jan Kuiper; Ralph D Perkins
Journal:  J Clin Orthop Trauma       Date:  2018-01-03

Review 8.  Review: Trunnionosis leading to modular femoral head dissociation.

Authors:  Agneish Dutta; James Nutt; Guy Slater; Syed Ahmed
Journal:  J Orthop       Date:  2021-01-30

Review 9.  Trunnionosis in total hip arthroplasty: a review.

Authors:  Jaydev B Mistry; Morad Chughtai; Randa K Elmallah; Aloise Diedrich; Sidney Le; Melbin Thomas; Michael A Mont
Journal:  J Orthop Traumatol       Date:  2016-02-11

10.  In Vivo Damage of the Head-Neck Junction in Hard-on-Hard Total Hip Replacements: Effect of Femoral Head Size, Metal Combination, and 12/14 Taper Design.

Authors:  Massimiliano Baleani; Paolo Erani; Barbara Bordini; Federica Zuccheri; Mateusz Kordian Mąkosa; Dalila De Pasquale; Alina Beraudi; Susanna Stea
Journal:  Materials (Basel)       Date:  2017-07-01       Impact factor: 3.623

View more