Literature DB >> 18613095

Fretting crevice corrosion of stainless steel stem-CoCr femoral head connections: comparisons of materials, initial moisture, and offset length.

Jeremy L Gilbert1, Manav Mehta, Bryan Pinder.   

Abstract

Modular tapers continue to be used in a wide variety of orthopedic implants. In this study, stainless steel (ASTM F-1568) femoral hip stems combined with Co-Cr-Mo alloy heads (SS/CoCr) were tested in an in vitro fretting corrosion test set-up to assess the propensity for mechanically assisted corrosion. Three different aspects of the modular design were evaluated in this study: (1) material combination compared to CoCr/CoCr, (2) wet versus dry assembly for SS/CoCr couples, and (3) 0- and 6-mm head offset for SS/CoCr couples. Fretting corrosion tests over a range of cyclic loads up to 3300 N were performed, and continuous cyclic loading at 3300 N for 1 M cycles were performed on each group (n = 5). Fretting micromotion was measured as a function of cyclic load on select couples to detect the nature and extent of motion present. The results showed that SS/CoCr couples were more susceptible to fretting corrosion than CoCr/CoCr couples, that dry assembly does not prevent fretting corrosion from taking place but raises the onset load, and that 6-mm offset heads had higher visual evidence of fretting damage but showed mixed statistical results in terms of onset loads and OCP shifts and currents compared to the 0-mm offset samples. Current and voltage excursions over 1 million cycles tended to diminish towards their unloaded control levels but did not fully recover until cyclic loading ceased. Micromotion measurements indicated fretting motions in the range of 10-25 microm where 0-mm heads tended to piston on the trunion, while 6 mm heads tended to rock. (c) 2008 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 18613095     DOI: 10.1002/jbm.b.31164

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  18 in total

1.  Modern trunnions are more flexible: a mechanical analysis of THA taper designs.

Authors:  David A Porter; Robert M Urban; Joshua J Jacobs; Jeremy L Gilbert; José A Rodriguez; H John Cooper
Journal:  Clin Orthop Relat Res       Date:  2014-09-30       Impact factor: 4.176

2.  The effect of manufacturing tolerances on the mechanical environment of taper junctions in modular TKR.

Authors:  Kyle Snethen; Jorge Hernandez; Melinda Harman
Journal:  J Mech Behav Biomed Mater       Date:  2019-04-18

3.  Severe metal-induced osteolysis many years after unipolar hip endoprosthesis.

Authors:  Matthew A Mann; Dylan Tanzer; Michael Tanzer
Journal:  Clin Orthop Relat Res       Date:  2013-01-05       Impact factor: 4.176

Review 4.  Electrical implications of corrosion for osseointegration of titanium implants.

Authors:  R A Gittens; R Olivares-Navarrete; R Tannenbaum; B D Boyan; Z Schwartz
Journal:  J Dent Res       Date:  2011-05-09       Impact factor: 6.116

5.  Otto Aufranc Award: Large Heads Do Not Increase Damage at the Head-neck Taper of Metal-on-polyethylene Total Hip Arthroplasties.

Authors:  Georgios K Triantafyllopoulos; Marcella E Elpers; Jayme C Burket; Christina I Esposito; Douglas E Padgett; Timothy M Wright
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

6.  Case reports: fractures of threaded cups: rare complications of a well-established implant.

Authors:  Arnd T Hoburg; Manav Mehta; Stephan Tohtz; Carsten Perka
Journal:  Clin Orthop Relat Res       Date:  2008-09-27       Impact factor: 4.176

7.  Properties and Corrosion Performance of Self-reinforced Composite PEEK for Proposed Use as a Modular Taper Gasket.

Authors:  Eric S Ouellette; Jeremy L Gilbert
Journal:  Clin Orthop Relat Res       Date:  2016-11       Impact factor: 4.176

8.  Does Surface Topography Play a Role in Taper Damage in Head-neck Modular Junctions?

Authors:  Robin Pourzal; Deborah J Hall; Nguyen Q Ha; Robert M Urban; Brett R Levine; Joshua J Jacobs; Hannah J Lundberg
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

9.  Acute trunnion failure of a TMZF alloy stem with large diameter femoral heads.

Authors:  E J Fahey; E Pomeroy; F E Rowan
Journal:  J Orthop       Date:  2020-01-10

10.  The role of fretting-frequency on the damage modes of THR modular junction: In-vitro study.

Authors:  Bartlomiej Radzik; Divya Bijukumar; Kai-Yuan Cheng; Ravindra V Badhe; Mark Barba; Mathew T Mathew
Journal:  Mater Sci Eng C Mater Biol Appl       Date:  2021-04-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.