Nancy Dunne1, Maria T Viggiani2, Scott Pardo2, Cynthia Robinson2, Joan Lee Parkes2.
Abstract
INTRODUCTION: The aim of this study was to compare the accuracy of 5 blood glucose monitoring systems (BGMSs; CONTOUR(®)PLUS [CP], Accu-Chek(®) Active [ACA], Accu-Chek(®) Performa [ACP], FreeStyle Freedom™ [FF], OneTouch(®) SelectSimple™ [OTSS]).Entities:
Keywords: Accuracy; MARD; Self-monitoring of blood glucose
Year: 2015 PMID: 26169192 PMCID: PMC4575307 DOI: 10.1007/s13300-015-0121-3
Source DB: PubMed Journal: Diabetes Ther ISSN: 1869-6961 Impact factor: 2.945
Subject demographic and baseline characteristics
| Characteristic |
|
|---|---|
| Gender, | |
| Female | 60 (57) |
| Male | 46 (43) |
| Type of diabetes, | |
| Type 1 | 8 (8) |
| Type 2 | 90 (85) |
| Type unknown | 2 (2) |
| Do not have diabetes | 6 (6) |
| Age, y | |
| Mean (range) | 57.9 (18–84) |
| Race, | |
| White/Caucasian | 87 (82) |
| Black/African American | 10 (9) |
| American Indian/Alaskan Native | 2 (2) |
| Other | 7 (7) |
| Duration of diabetes, | |
| 1–3 mo | 2 (2) |
| 4–6 mo | 2 (2) |
| 7–12 mo | 1 (1) |
| 13 mo–2 y | 5 (5) |
| 3–5 y | 11 (11) |
| 6–10 y | 25 (25) |
| >10 y | 54 (54) |
| Frequency of daily SMBG, | |
| >4 | 12 (12) |
| 4 | 13 (13) |
| 3 | 19 (19) |
| 2 | 20 (20) |
| 1 | 23 (23) |
| <1 | 10 (10) |
| Did not test | 3 (3) |
y years, mo months, SMBG self-monitoring of blood glucose
aPercentages may not total 100% because of rounding
bOnly for subjects with diabetes, n = 100
MARD comparisons to the CP BGMS
| CP | ACA | ACP | FF | OTSS | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| MARD (%) |
| MARD (%) |
| MARD (%) |
| MARD (%) |
| MARD (%) |
| ||
| All samples | Overall (YSI 27–460 mg/dL [1.5–25.5 mmol/L]; | 3.34 | NA | 5.83 | <0.0001b | 4.85 | <0.0001b | 15.32 | <0.0001b | 10.70 | <0.0001b |
| Low (YSI ≤80 mg/dL [≤4.4 mmol/L]; | 3.42 | NA | 6.38 | <0.0001b | 4.18 | 0.0475 | 17.73 | <0.0001b | 13.32 | <0.0001b | |
| Middle (YSI 81–180 mg/dL [4.5–10.0 mmol/L]; | 3.35 | NA | 5.42 | <0.0001b | 4.48 | 0.0494 | 14.11 | <0.0001b | 8.99 | <0.0001b | |
| High (YSI >180 mg/dL [>10.0 mmol/L]; | 3.26 | NA | 5.70 | 0.0006b | 5.90 | <0.0001b | 14.14 | <0.0001b | 9.79 | <0.0001b | |
| Unmodified samples | Overall (YSI 45–460 mg/dL [2.5–25.5 mmol/L]; | 2.95 | NA | 3.95 | 0.0579 | 4.55 | 0.0010b | 14.31 | <0.0001b | 6.68 | <0.0001b |
MARD mean absolute relative difference, CP CONTOUR®PLUS, BGMS blood glucose monitoring system, ACA Accu-Chek® Active, ACP Accu-Chek® Performa, FF FreeStyle Freedom™, OTSS OneTouch® SelectSimple™, YSI Yellow Springs Instruments, NA not applicable
aOf the 314 samples tested, 105 were unmodified and 209 were modified. Results for 4 samples were not analyzed as per data handling rules of the protocol
bVersus CP BGMS (P values <0.0125 are considered significant)
Fig. 1Modified Bland-Altman plots of the difference of BGMS results from laboratory reference results for a CP, b ACA, c ACP, d FF, and e OTSS. BGMS blood glucose monitoring system, CP CONTOUR®PLUS, ACA Accu-Chek® Active, ACP Accu-Chek® Performa, FF FreeStyle Freedom™, OTSS OneTouch® SelectSimple™, YSI Yellow Springs Instruments. Dashed lines ±15 mg/dL (±0.8 mmol/L) or ±15% of the reference result for samples with YSI glucose concentrations <100 mg/dL (<5.6 mmol/L) and ≥100 mg/dL (≥5.6 mmol/L), respectively
Parkes-Consensus Error Grid analysis of BGMS results
| Meter system | Number of results within error grid zones | ||
|---|---|---|---|
| Zone A | Zone B | Zones C, D, and E | |
| CP | 313/314 (99.7%) | 1/314 (0.3%) | 0/314 (0%) |
| ACA | 308/314 (98.1%) | 6/314 (1.9%) | 0/314 (0%) |
| ACP | 312/314 (99.4%) | 2/314 (0.6%) | 0/314 (0%) |
| FF | 296/314 (94.3%) | 18/314 (5.7%) | 0/314 (0%) |
| OTSS | 296/314 (94.3%) | 18/314 (5.7%) | 0/314 (0%) |
BGMS blood glucose monitoring system, CP CONTOUR®PLUS, ACA Accu-Chek® Active, ACP Accu-Chek® Performa, FF FreeStyle Freedom™, OTSS OneTouch® SelectSimple™
Comparison of BGMS specifications
| Meter system | Test strip name | Test strip chemistry | Interference from oxygen [ |
|---|---|---|---|
| CP | CP test strips | FAD-GDH | Not sensitive |
| ACA | ACA test strips | PQQ-GDH | Not sensitive |
| ACP | ACP test strips | PQQ-GDH | Not sensitive |
| FF | FL test strips | FAD-GDH | Not sensitive |
| OTSS | OTS test strips | GO | Sensitive |
BGMS blood glucose monitoring system, CP CONTOUR®PLUS, FAD flavin adenine dinucleotide, GDH glucose dehydrogenase, ACA Accu-Chek® Active, PQQ pyrroloquinoline quinone, ACP Accu-Chek® Performa, FF FreeStyle Freedom™, FL FreeStyle Lite™, OTSS OneTouch® SelectSimple™, OTS OneTouch® Select, GO glucose oxidase