Patrick Schoettker1, Jocelyn Corniche1.
Abstract
We conducted a study assessing the quality and speed of intubation between the Airtraq with its new iPhone AirView app and the King Vision in a manikin. The primary endpoint was reduction of time needed for intubation. Secondary endpoints included times necessary for intubation. 30 anaesthetists randomly performed 3 intubations with each device on a difficult airway manikin. Participants had a professional experience of 12 years: 60.0% possessed the Airtraq in their hospital, 46.7% the King Vision, and 20.0% both. Median time difference [IQR] to identify glottis (1.1 [-1.3; 3.9] P = 0.019), for tube insertion (2.1 [-2.6; 9.4] P = 0.002) and lung ventilation (2.8 [-2.4; 11.5] P = 0.001), was shorter with the Airtraq-AirView. Median time for glottis visualization was significantly shorter with the Airtraq-AirView (5.3 [4.0; 8.4] versus 6.4 [4.6; 9.1]). Cormack Lehane before intubation was better with the King Vision (P = 0.03); no difference was noted during intubation, for subjective device insertion or quality of epiglottis visualisation. Assessment of tracheal tube insertion was better with the Airtraq-AirView. The Airtraq-AirView allows faster identification of the landmarks and intubation in a difficult airway manikin, while clinical relevance remains to be studied. Anaesthetists assessed the intubation better with the Airtraq-AirView.Entities:
Mesh:
Year: 2015 PMID: 26161393 PMCID: PMC4486309 DOI: 10.1155/2015/284142
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1
Figure 2|
Airtraq-AirView |
King Vision | Delta K-A |
| ||||
|---|---|---|---|---|---|---|---|
| Median | IQR | Median | IQR | Median | IQR | ||
| Time to identify glottis | 5.3 | [3.9; 8.4] | 6.4 | [4.6; 9.1] | 1.1 | [−1.3; 3.9] | 0.019 |
| Time to inflate the cuff | 14.2 | [9.1; 18.2] | 15.2 | [10.9; 26.1] | 2.1 | [−2.6; 9.4] | 0.002 |
| Time to ventilate the lungs | 16.6 | [11.9; 21.1] | 17.9 | [13.6; 28.5] | 2.8 | [−2.4; 11.5] | 0.0002 |
Times are expressed in seconds.
| Cormack Lehane | Airtraq-AirView | King Vision | ||
|---|---|---|---|---|
|
| % |
| % | |
| 1 | 70 | 77.8 | 81 | 90.0 |
| 2a | 20 | 22.2 | 8 | 8.9 |
| 2b | 0 | 1 | 1.1 | |
| Ease of device insertion | Ease of epiglottis visualisation | Ease of intubation | ||||
|---|---|---|---|---|---|---|
| AT-AirView | King Vision | AT-AirView | King Vision | AT-AirView | King Vision | |
| 1 | 51 | 48 | 66 | 68 | 63 | 52 |
| 2 | 33 | 34 | 24 | 20 | 22 | 15 |
| 3 | 6 | 8 | 0 | 1 | 5 | 14 |
| 4 | 0 | 0 | 0 | 1 | 0 | 9 |