| Literature DB >> 22670846 |
D Ranieri1, S M Filho, S Batista, P do Nascimento.
Abstract
This study compared intubation conditions produced by the Macintosh and Airtraq™ laryngoscopes when used in obese patients in the ramped position. One hundred and thirty-two patients having bariatric surgery were placed in the ramped position and randomly assigned to have their tracheas intubated using either the Macintosh (n=64) or an Airtraq (n=68) laryngoscope. Mean (SD) intubation times were 37 (23) s and 14 (3) s for Macintosh and Airtraq, respectively (p<0.0001). Compared with the Macintosh laryngoscope, the Airtraq laryngoscope provided an improved vocal cord view as assessed by the Cormack and Lehane score (number of patients with Cormack-Lehane scores of 1/2/3/4/37/20/4/3 for the Macintosh, and 65/3/0/0 for the Airtraq laryngoscopes, p<0.0001). One patient in the Macintosh group had a failed intubation and intubation was achieved with the Airtraq. For obese patients in ramped position, Airtraq affords faster tracheal intubation than the Macintosh laryngoscope. AnaesthesiaEntities:
Mesh:
Substances:
Year: 2012 PMID: 22670846 DOI: 10.1111/j.1365-2044.2012.07200.x
Source DB: PubMed Journal: Anaesthesia ISSN: 0003-2409 Impact factor: 6.955