BACKGROUND: Direct laryngoscopy used for tracheal intubation requires aligning the pharyngeal, laryngeal and oral axes to achieve a line of sight. Video laryngoscopy provides a better view of the glottis without the need for aligning the three axes. AIMS: To evaluate the effectiveness of King vision laryngoscope over Macintosh laryngoscope in visualizing the glottis and intubating the trachea, when used on a same patient as in a cross over manner. SETTINGS ANDEntities:
Keywords: Cormack-Lehane grade; Macintosh laryngoscope; king vision laryngoscope; laryngoscopy
Year: 2019 PMID: 31031477 PMCID: PMC6444949 DOI: 10.4103/aer.AER_165_18
Source DB: PubMed Journal: Anesth Essays Res ISSN: 2229-7685
Figure 1Standard King Vision laryngoscope with monitor and TrueFlex stylet
Figure 2King Vision video laryngoscope and endotracheal tube loaded with TrueFlex stylet
Patients’ characteristics and airway data in Macintosh and King Vision groups. Data are expressed as mean (SD)
| Variable | Mcintosh | King Vision |
|---|---|---|
| Age (yrs) | 38 (8) | 42 (14) |
| Weight (kg) | 63 (12) | 67 (11) |
| Physical Status (ASA I/II/III) | 15/15/0 | 13/17/0 |
| Mallampati Class (I/II/III) | 15/12/3 | 16/11/3 |
| Thyromental distance (cm) | 10 (1) | 9.52 (0.93) |
| Cormack and Lehane Grade (1/2/3) | 19/6/5 | 18/8/4 |
Hemodynamic variables (mean arterial pressure in mmHg, heart rate in beats/min)
| Variable | Macintosh group | King Vision group | |
|---|---|---|---|
| MAP (immediate post intubation) | 92.87±20.75 | 101.67±14.36 | 0.611 |
| MAP (3 min post intubation) | 84.23±12.38 | 86.87±9.41 | 0.3563 |
| MAP (5 min post intubation) | 79.2±6.7 | 83.3±9.78 | 0.632 |
| HR (immediate post intubation) | 97.1±13.35 | 97±16.16 | 0.9792 |
| HR (3 min post intubation) | 89±11.27 | 86.6±10.88 | 0.4048 |
| HR (5 min post intubation) | 83.17±10.08 | 85.9±10.65 | 0.3121T |
MAP: Mean Arterial Pressure, HR: Heart Rate
Comparison of time to intubate between Macintosh and King Vision groups. Values are expressed as mean (95% Confidence Interval)
| Macintosh ( | King Vision ( | |
|---|---|---|
| TTI (sec) | 29.97 (28.05-31.89) | 42.77 (40.05-45.48) |
| TTI (sec) for C-L grade 1 | 30.84 (28.56-33.12) | 42.11 (38.98-45.24) |
| TTI (sec) for C-L grade 2 | 31 (23.79-38.21) | 45.25 (37.84-52.66) |
| TTI (sec) for C-L grade 3 | 25.4 (24.29-26.51) | 40.75 (26.55-54.95) |
TTI: Time To Intubate, Sec: seconds
Comparison of laryngoscopy grades with the King Vision scope in the King Vision group (n=30)
| Macintosh C-L grade | King Vision C-L grade | |||
|---|---|---|---|---|
| Grade 1 | Grade 2 | Grade 3 | Total | |
| Grade 1 | 18 | 0 | 0 | 18 |
| Grade 2 | 6* | 2 | 0 | 8 |
| Grade 3 | 2* | 1* | 1 | 4 |
| Total | 26 | 3 | 1 | 30 |
P<0.05 (McNemar χ2-test). *Patients with improvement in C-L grade