Literature DB >> 19522050

Learning curves of the Glidescope, the McGrath and the Airtraq laryngoscopes: a manikin study.

Georges L Savoldelli1, Eduardo Schiffer, Christoph Abegg, Vincent Baeriswyl, François Clergue, Jean-Luc Waeber.   

Abstract

BACKGROUND AND
OBJECTIVE: Several video and optical laryngoscopes have been developed but few have been compared in terms of their learning curves and efficacy. Using a manikin with normal airways we compared the Glidescope, the McGrath, the Airtraq and the Macintosh laryngoscopes.
METHODS: Sixty anaesthetists (20 staff, 20 residents and 20 nurses) participated in the study. All subjects were novice with the new devices. They intubated a Laerdal SimMan manikin (with normal airway) five times in a row with all laryngoscopes. The sequence of use of the devices was randomized. Before using a device, a presentation and a demonstration were provided. Outcome measures were: duration of intubation attempt, modified Cormack grades, dental trauma and difficulty of use.
RESULTS: The Airtraq had the most favourable learning curve and mirrored the Macintosh after two intubation attempts. The Glidescope and McGrath had steep learning curves but, after five attempts, differences persisted when compared with the Macintosh and Airtraq. Time taken to visualize the glottis was similar but time taken to position the endotracheal tube was shorter for the Airtraq when compared with the Glidescope and McGrath. Indirect laryngoscopes seemed to have advantages over the Macintosh blade in terms of laryngeal exposure and potential dental trauma.
CONCLUSIONS: In a 'normal airway' model, intubation skills with the new devices appeared to be rapidly mastered. The three indirect laryngoscopes provided a better glottic exposure than the Macintosh. The Airtraq displayed the most favourable learning curve, probably reflecting differences in the techniques of endotracheal tube placement: guiding channel versus steering technique.

Entities:  

Mesh:

Year:  2009        PMID: 19522050     DOI: 10.1097/eja.0b013e3283269ff4

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  33 in total

1.  [Rapid sequence induction and intubation in patients with risk of aspiration : Recommendations for action for practical management of anesthesia].

Authors:  C Eichelsbacher; H Ilper; R Noppens; J Hinkelbein; T Loop
Journal:  Anaesthesist       Date:  2018-06-29       Impact factor: 1.041

Review 2.  [Indirect laryngoscopy : Alternatives to securing the airway].

Authors:  R R Noppens; C Werner; T Piepho
Journal:  Anaesthesist       Date:  2010-02       Impact factor: 1.041

3.  [Comparison of GlideScope® Cobalt and McGrath® Series 5 video laryngoscopes with direct laryngoscopy in a simulated regurgitation/aspiration scenario].

Authors:  M Kriege; T Piepho; H Buggenhagen; R R Noppens
Journal:  Med Klin Intensivmed Notfmed       Date:  2015-01-11       Impact factor: 0.840

4.  Comparison of Airtraq optical laryngoscope and Storz video laryngoscope in a cadaver model.

Authors:  Michael C Wadman; Travis W Dierks; Chad E Branecki; Claudia L Barthold; Lance H Hoffman; Lina Lander; Carol S Lomneth; Richard A Walker
Journal:  World J Emerg Med       Date:  2011

5.  Comparison of the McGrath® Series 5 and GlideScope® Ranger with the Macintosh laryngoscope by paramedics.

Authors:  Tim Piepho; Kathrin Weinert; Florian M Heid; Christian Werner; Rüdiger R Noppens
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2011-01-17       Impact factor: 2.953

6.  Airtraq, LMA CTrach and Macintosh Laryngoscopes in Tracheal Intubation Training: A Randomized Comparative Manikin Study.

Authors:  Ayten Saraçoğlu; Didem Dal; Ömer Baygın; Fevzi Yılmaz Göğüş
Journal:  Turk J Anaesthesiol Reanim       Date:  2016-04-01

7.  Intubation of a Paediatric Manikin in Tongue Oedema and Face-to-Face Simulations by Novice Personnel: a Comparison of Glidescope, Airtraq and Direct Laryngoscopy.

Authors:  Zehra İpek Arslan; Canan Turna; Nevin Esra Gümüş; Kamil Toker; Mine Solak
Journal:  Turk J Anaesthesiol Reanim       Date:  2016-04-01

8.  Fluoroscopic Comparison of Cervical Spine Motion Using LMA CTrach, C-MAC Videolaryngoscope and Macintosh Laryngoscope.

Authors:  Tülay Şahin; Zehra İpek Arslan; Gür Akansel; Onur Balaban; Derya Berk; Mine Solak; Kamil Toker
Journal:  Turk J Anaesthesiol Reanim       Date:  2018-02-01

9.  Use of Airtraq, C-Mac, and Glidescope laryngoscope is better than Macintosh in novice medical students' hands: A manikin study.

Authors:  Abdullah M Kaki; Waleed A Almarakbi; Hazem M Fawzi; Abdulaziz M Boker
Journal:  Saudi J Anaesth       Date:  2011-10

10.  The use of the Airtraq® optical laryngoscope for routine tracheal intubation in high-risk cardio-surgical patients.

Authors:  Gereon Schälte; Ulrike Scheid; Steffen Rex; Mark Coburn; Britta Fiedler; Rolf Rossaint; Norbert Zoremba
Journal:  BMC Res Notes       Date:  2011-10-19
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.