Implementing the complete beam hardening effect of the bowtie filter versus scaling beam intensities: effects on dosimetric applications in computed tomography.

Xochitl Lopez-Rendon1, Guozhi Zhang2, Hilde Bosmans3, Raymond Oyen3, Federica Zanca1.   

Abstract

The bowtie filter is an essential element of computed tomography scanners. Implementation of this filter in a Monte Carlo dosimetry platform can be based on Turner's method, which describes how to measure the filter thickness and relate the x-ray beam as a function of bowtie angle to the central beam. In that application, the beam hardening is accounted for by means of weighting factors that are associated to the photons according to their position (fan angle) and energy. We assessed an alternative approximation in which the photon spectrum is given a fan angle-dependent scaling factor. The aim of our investigation was to evaluate the effects on dose accuracy estimation when using the gold standard bowtie filter method versus a beam scaling approximation method. In particular, we wanted to assess the percentage dose differences between the two methods for several water thicknesses representative for different patients of different body mass index. The largest percentage differences were found for the thickest part of the bowtie filter and increased with patient size.

Entities:  

Keywords:  Monte Carlo; bowtie filter; computed tomography; dose; patient size

Year:  2014        PMID: 26158059      PMCID: PMC4478776          DOI: 10.1117/1.JMI.1.3.033507

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  10 in total

1.  History by history statistical estimators in the BEAM code system.

Authors:  B R B Walters; I Kawrakow; D W O Rogers
Journal:  Med Phys       Date:  2002-12       Impact factor: 4.071

2.  Spektr: a computational tool for x-ray spectral analysis and imaging system optimization.

Authors:  J H Siewerdsen; A M Waese; D J Moseley; S Richard; D A Jaffray
Journal:  Med Phys       Date:  2004-11       Impact factor: 4.071

Review 3.  Principles of CT and CT technology.

Authors:  Lee W Goldman
Journal:  J Nucl Med Technol       Date:  2007-09

4.  A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models.

Authors:  G Jarry; J J DeMarco; U Beifuss; C H Cagnon; M F McNitt-Gray
Journal:  Phys Med Biol       Date:  2003-08-21       Impact factor: 3.609

5.  Bowtie filtration for dedicated cone beam CT of the head and neck: a simulation study.

Authors:  G Zhang; N Marshall; R Jacobs; Q Liu; H Bosmans
Journal:  Br J Radiol       Date:  2013-05-31       Impact factor: 3.039

6.  Patient dosimetry for x rays used in medical imaging.

Authors: 
Journal:  J ICRU       Date:  2005-12

7.  A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT Monte Carlo dosimetry simulations.

Authors:  Adam C Turner; Di Zhang; Hyun J Kim; John J DeMarco; Chris H Cagnon; Erin Angel; Dianna D Cody; Donna M Stevens; Andrew N Primak; Cynthia H McCollough; Michael F McNitt-Gray
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

8.  Monte Carlo modeling for dose assessment in cone beam CT for oral and maxillofacial applications.

Authors:  G Zhang; N Marshall; R Bogaerts; R Jacobs; H Bosmans
Journal:  Med Phys       Date:  2013-07       Impact factor: 4.071

9.  Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product.

Authors:  Paul D Deak; Yulia Smal; Willi A Kalender
Journal:  Radiology       Date:  2010-10       Impact factor: 11.105

10.  Effective dose: how should it be applied to medical exposures?

Authors:  C J Martin
Journal:  Br J Radiol       Date:  2007-07-23       Impact factor: 3.039

  10 in total
  4 in total

1.  Comparing different methods for estimating radiation dose to the conceptus.

Authors:  X Lopez-Rendon; M S Walgraeve; S Woussen; A Dedulle; G Zhang; H Bosmans; F Zanca
Journal:  Eur Radiol       Date:  2016-05-10       Impact factor: 5.315

2.  Clinical indications and radiation doses to the conceptus associated with CT imaging in pregnancy: a retrospective study.

Authors:  S Woussen; X Lopez-Rendon; D Vanbeckevoort; H Bosmans; R Oyen; F Zanca
Journal:  Eur Radiol       Date:  2015-07-23       Impact factor: 5.315

3.  Do we need 3D tube current modulation information for accurate organ dosimetry in chest CT? Protocols dose comparisons.

Authors:  Xochitl Lopez-Rendon; Guozhi Zhang; Walter Coudyzer; Wim Develter; Hilde Bosmans; Federica Zanca
Journal:  Eur Radiol       Date:  2017-05-19       Impact factor: 5.315

4.  Determination and verification of the x-ray spectrum of a CT scanner.

Authors:  Ahmad Ibrahim Hassan; Martin Skalej; Helmut Schlattl; Christoph Hoeschen
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-07
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.