Bowtie filtration for dedicated cone beam CT of the head and neck: a simulation study.

G Zhang1, N Marshall, R Jacobs, Q Liu, H Bosmans.   

Abstract

OBJECTIVE: To investigate the influence of bowtie filtration on dedicated cone beam CT (CBCT) of the head and neck.
METHODS: A validated hybrid simulation technique was used to model a commercial CBCT system with offset scanning geometry, 90 kV tube potential and 145×75 mm imaging field of view. Three bowtie filters were formulated to produce uniform flux intensity in the projection image of cylindrical objects of diameter 14, 16 and 18 cm. The influence of these simulated filters was compared with the original flat filtration in terms of the output radiation field, the dose delivered to the object, the scatter distribution in projections and the quality of the reconstructed image.
RESULTS: Compared against flat filtration, dose reduction for the bowtie case, examined as a function of radial distance within a 16-cm-diameter water cylinder, varied from 8.7% at the centre to 53.8% at the periphery. Scatter reduction, quantified using scatter-to-primary ratio in projection images, was up to 37.6% for a 14-cm-diameter cylindrical contrast phantom. Using the supplied routine image reconstruction, bowtie filtration resulted in comparable noise appearance, contrast resolution and artefact pattern for computational anatomical phantoms, with <5% difference in contrast-to-noise ratio.
CONCLUSION: Bowtie filtration can effectively reduce the dose and scatter in CBCT of the head and neck. For better image quality, corresponding modification to the image pre-processing and reconstruction is needed. ADVANCES IN KNOWLEDGE: The hybrid simulation approach can usefully explore the impact of proposed system component and design changes.

Entities:  

Mesh:

Year:  2013        PMID: 23728948      PMCID: PMC3745055          DOI: 10.1259/bjr.20130002

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  10 in total

1.  X-ray micro-CT with a displaced detector array.

Authors:  Ge Wang
Journal:  Med Phys       Date:  2002-07       Impact factor: 4.071

2.  Characteristics of kilovoltage x-ray beams used for cone-beam computed tomography in radiation therapy.

Authors:  George X Ding; Dennis M Duggan; Charles W Coffey
Journal:  Phys Med Biol       Date:  2007-02-27       Impact factor: 3.609

Review 3.  Conebeam CT of the head and neck, part 1: physical principles.

Authors:  A C Miracle; S K Mukherji
Journal:  AJNR Am J Neuroradiol       Date:  2009-05-13       Impact factor: 3.825

4.  The influence of bowtie filtration on cone-beam CT image quality.

Authors:  N Mail; D J Moseley; J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

Review 5.  Artefacts in CBCT: a review.

Authors:  R Schulze; U Heil; D Gross; D D Bruellmann; E Dranischnikow; U Schwanecke; E Schoemer
Journal:  Dentomaxillofac Radiol       Date:  2011-07       Impact factor: 2.419

6.  Method for evaluating bow tie filter angle-dependent attenuation in CT: theory and simulation results.

Authors:  John M Boone
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

7.  Development and validation of a hybrid simulation technique for cone beam CT: application to an oral imaging system.

Authors:  G Zhang; R Pauwels; N Marshall; E Shaheen; J Nuyts; R Jacobs; H Bosmans
Journal:  Phys Med Biol       Date:  2011-08-16       Impact factor: 3.609

8.  The effects of compensator and imaging geometry on the distribution of x-ray scatter in CBCT.

Authors:  G J Bootsma; F Verhaegen; D A Jaffray
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

9.  Computerized three-dimensional segmented human anatomy.

Authors:  I G Zubal; C R Harrell; E O Smith; Z Rattner; G Gindi; P B Hoffer
Journal:  Med Phys       Date:  1994-02       Impact factor: 4.071

10.  Compensators for dose and scatter management in cone-beam computed tomography.

Authors:  S A Graham; D J Moseley; J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

  10 in total
  4 in total

1.  Implementing the complete beam hardening effect of the bowtie filter versus scaling beam intensities: effects on dosimetric applications in computed tomography.

Authors:  Xochitl Lopez-Rendon; Guozhi Zhang; Hilde Bosmans; Raymond Oyen; Federica Zanca
Journal:  J Med Imaging (Bellingham)       Date:  2014-12-30

2.  A dedicated breast-PET/CT scanner: Evaluation of basic performance characteristics.

Authors:  Raymond R Raylman; Will Van Kampen; Alexander V Stolin; Wenbo Gong; Gangadhar Jaliparthi; Peter F Martone; Mark F Smith; David Sarment; Neal H Clinthorne; Mark Perna
Journal:  Med Phys       Date:  2018-02-23       Impact factor: 4.071

3.  Scatter-to-primary ratio in dentomaxillofacial cone-beam CT: effect of field of view and beam energy.

Authors:  Ruben Pauwels; Pisha Pittayapat; Phonkit Sinpitaksakul; Soontra Panmekiate
Journal:  Dentomaxillofac Radiol       Date:  2021-04-29       Impact factor: 2.419

4.  0° vs. 180° CT localiser: The effect of vertical off-centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography.

Authors:  Yazan Al-Hayek; Xiaoming Zheng; Rob Davidson; Christopher Hayre; Dana Al-Mousa; Campbell Finlay; Kelly Spuur
Journal:  J Med Radiat Sci       Date:  2021-08-17
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.