Literature DB >> 26152781

Catastrophic Femoral Neck Failure after THA with the Accolade(®) I Stem in Three Patients.

Jonathon Spanyer1, Jennifer Hines2, Christopher Maxwell Beaumont3, Jonathan Yerasimides4,5.   

Abstract

CASE DESCRIPTION: We report a series of three femoral stem failures, each occurring at the head-neck junction, with all patients experiencing limited and painful ambulation, leading to subsequent revision arthroplasty. All patients were male with high-offset femoral stems and increased head lengths, and each had undergone primary THA at a minimum of 7 years before presentation (average, 94 months). There were no associated deep infections or cases of aseptic loosening in the cohort. LITERATURE REVIEW: There is a paucity of similar reports in the literature regarding femoral stem failure at the head-neck junction. When failures of titanium stems have been reported, failure has been attributed to material design and geometry, laser etching, overload, implant alignment, and patient characteristics. PURPOSE AND CLINICAL RELEVANCE: Catastrophic failures of femoral stems at the head-neck junction are a rare cause for revision after THA. Component material and design, surgical technique, and patient factors may contribute.

Entities:  

Mesh:

Year:  2015        PMID: 26152781      PMCID: PMC4814423          DOI: 10.1007/s11999-015-4438-z

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  18 in total

1.  Prosthetic component failures in hip arthroplasty surgery.

Authors:  D A Heck; C M Partridge; J D Reuben; W L Lanzer; C G Lewis; E M Keating
Journal:  J Arthroplasty       Date:  1995-10       Impact factor: 4.757

2.  The tradeoffs associated with modular hip prostheses.

Authors:  J P Collier; M B Mayor; I R Williams; V A Surprenant; H P Surprenant; B H Currier
Journal:  Clin Orthop Relat Res       Date:  1995-02       Impact factor: 4.176

3.  Fracture failure of a forged vitallium prosthesis. A case report.

Authors:  E H Miller; R Shastri; C I Shih
Journal:  J Bone Joint Surg Am       Date:  1982-12       Impact factor: 5.284

4.  Early fatigue failures of cemented, forged, cobalt-chromium femoral stems at the neck-shoulder junction.

Authors:  E W Lee; H T Kim
Journal:  J Arthroplasty       Date:  2001-02       Impact factor: 4.757

5.  Fatigue fracture of a forged cobalt-chromium-molybdenum femoral component inserted with cement. A report of ten cases.

Authors:  S T Woolson; J P Milbauer; J D Bobyn; S Yue; W J Maloney
Journal:  J Bone Joint Surg Am       Date:  1997-12       Impact factor: 5.284

6.  Trunion fracture of a fully porous-coated femoral stem. Case report.

Authors:  Torey P Botti; Justin Gent; John M Martell; David W Manning
Journal:  J Arthroplasty       Date:  2005-10       Impact factor: 4.757

7.  Gross trunnion failure after primary total hip arthroplasty.

Authors:  Samik Banerjee; Jeffrey J Cherian; James V Bono; Steven M Kurtz; Rudolph Geesink; R Michael Meneghini; Ronald E Delanois; Michael A Mont
Journal:  J Arthroplasty       Date:  2014-11-26       Impact factor: 4.757

8.  Fracture of the femoral component after total hip replacement. An analysis of fifty-eight cases.

Authors:  E Y Chao; M B Coventry
Journal:  J Bone Joint Surg Am       Date:  1981-09       Impact factor: 5.284

9.  Faulty design increases the risk of neck fracture in a hip prosthesis.

Authors:  Nosrat Vatani; Daniel Comando; José Acuña; Daniel Prieto; Horacio Caviglia
Journal:  Acta Orthop Scand       Date:  2002-10

10.  Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants.

Authors:  J L Gilbert; C A Buckley; J J Jacobs; K C Bertin; M R Zernich
Journal:  J Bone Joint Surg Am       Date:  1994-01       Impact factor: 5.284

View more
  5 in total

1.  Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.

Authors:  Kenneth L Urish; Brian R Hamlin; Anton Y Plakseychuk; Timothy J Levison; Genymphas B Higgs; Steven M Kurtz; Anthony M DiGioia
Journal:  J Arthroplasty       Date:  2017-04-08       Impact factor: 4.757

Review 2.  Gross trunnion failure in metal on polyethylene total hip arthroplasty-a systematic review of literature.

Authors:  Tungish Bansal; Sameer Aggarwal; Mandeep Singh Dhillon; Sandeep Patel
Journal:  Int Orthop       Date:  2020-01-04       Impact factor: 3.075

3.  Comparison of Asymptomatic and Symptomatic Adverse Local Tissue Reaction in Patients With Head-Neck Taper Corrosion.

Authors:  Margaret A Weber; Matthew J Snyder; Kalain K Workman; Margaret M Sims; Clair N Smith; Deepak Kumar; Akshay V Daji; Camilo G Borrero; Andrew C Cordle; Anthony M DiGioia; Brian R Hamlin; Anton Y Plakseychuk; Kenneth L Urish
Journal:  J Arthroplasty       Date:  2020-09-17       Impact factor: 4.435

4.  Femoral head disengagement from Accolade femoral stem in MOM Arthroplasty: a case study and literature review.

Authors:  Mohamed F H Elsheikh; Rehan Gul
Journal:  SICOT J       Date:  2019-08-20

Review 5.  Femoral prosthesis neck fracture following total hip arthroplasty - a systematic review.

Authors:  P G van Doesburg; E J van Langelaan; I Apachitei; M R Bénard; S H M Verdegaal
Journal:  Arthroplasty       Date:  2020-10-16
  5 in total