Literature DB >> 6985558

Fracture of the femoral component after total hip replacement. An analysis of fifty-eight cases.

E Y Chao, M B Coventry.   

Abstract

Clinical, radiographic, biomechanical, and metallurgical data from fifty-eight patients with a fractured stem of the femoral component of a total hip replacement (thirty-seven Charnley, sixteen Müller, three Trapezoidal-28, and two Bechol prostheses) were analyzed, as well as the cases of twenty-seven control patients with matching clinical and radiographic data who had a Charnley prosthesis but no fracture of the stem. Radiographic data in the form of defined measurements were classified into discrete variables for statistical analysis. These radiographic variables, along with clinical variables, were rated by linear discriminant analysis and a fracture risk index (the sum of the rating scores or regressive coefficient) was derived for each patient. Using this index, two zones (of index values) were defined: one for the patients at risk for fracture and the other for those not at risk for fracture according to the discriminative scale. All patients in both the fracture group and the non-fracture group were separated successfully into one of these two zones, except for one patient with a metallurgical defect of the prosthesis and a stem in marked valgus orientation. Based on the values of the risk index. the zone for those at risk for fracture and the zone for those not at risk were each divided into three regions - questionable, marginal, and positive - to give more weight to the contributions of clinical, biomechanical, and material factors in the prediction of fractures of prosthetic stems. The mechanisms causing stem fracture that seemed to involve multiple variables of different origins were identified and an attempt was made to rank these groups of variables as causes of stem failures. A risk index developed in this way could be useful in predicting the possibility of fracture or loosening of a prosthetic stem after total hip replacement.

Entities:  

Mesh:

Substances:

Year:  1981        PMID: 6985558

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  14 in total

Review 1.  Catastrophic Femoral Neck Failure after THA with the Accolade(®) I Stem in Three Patients.

Authors:  Jonathon Spanyer; Jennifer Hines; Christopher Maxwell Beaumont; Jonathan Yerasimides
Journal:  Clin Orthop Relat Res       Date:  2015-07-08       Impact factor: 4.176

2.  Laser etching causing fatigue fracture at the neck-shoulder junction of an uncemented femoral stem: A case report.

Authors:  Bob Jang; Andrew Kanawati; Declan Brazil; Warwick Bruce
Journal:  J Orthop       Date:  2013-05-24

3.  No influence of body mass index on early outcome following total hip arthroplasty.

Authors:  T Ibrahim; S Hobson; A Beiri; C N Esler
Journal:  Int Orthop       Date:  2005-09-24       Impact factor: 3.075

4.  Failure of the stem in total hip replacement. A study of aetiology and mechanism of failure in 13 cases.

Authors:  U E Pazzaglia; F Ghisellini; D Barbieri; L Ceciliani
Journal:  Arch Orthop Trauma Surg       Date:  1988

5.  A clinical study of the results of cementless total hip replacement.

Authors:  Y Itami; N Akamatsu; Y Tomita; M Nagai; I Nakajima
Journal:  Arch Orthop Trauma Surg       Date:  1983

6.  Failures of total hip arthroplasty and probable incidence of revision surgery in the future. Calculations according to a mathematical model based on a ten years' experience in total hip arthroplasty.

Authors:  E Morscher; A Schmassmann
Journal:  Arch Orthop Trauma Surg       Date:  1983

7.  Fracture of Fully-coated Femoral Stem after Primary Total Hip Arthroplasty for Nonunion of Intertrochanteric Fracture: A Case Report.

Authors:  Young Soo Chun; Hyung Suk Juh; Yoon Je Cho; Kee Hyung Rhyu
Journal:  Hip Pelvis       Date:  2015-09-30

8.  Hip and knee arthroplasty implants contraindicated in obesity.

Authors:  J D Craik; M D Bircher; M Rickman
Journal:  Ann R Coll Surg Engl       Date:  2016-03-29       Impact factor: 1.891

9.  A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development.

Authors:  Tomohiro Mimura
Journal:  Case Rep Orthop       Date:  2015-05-21

10.  Neck fracture of a cementless forged titanium alloy femoral stem following total hip arthroplasty: a case report and review of the literature.

Authors:  Theodoros B Grivas; Olga D Savvidou; Spyridon A Psarakis; Pierre-Francois Bernard; George Triantafyllopoulos; Ioannis Kovanis; Panagiotis Alexandropoulos
Journal:  J Med Case Rep       Date:  2007-12-06
View more