| Literature DB >> 35236443 |
P G van Doesburg1, E J van Langelaan2, I Apachitei2, M R Bénard3, S H M Verdegaal3.
Abstract
PURPOSE: Head-neck modularity was introduced into total hip arthroplasty to provide more intraoperative surgical options. However, modularity led to new problems, such as trunnionosis and fractures of the femoral prosthesis neck. The purpose of this study was to identify risk factors for hip neck fractures and to provide recommendations to prevent damage and fractures of the neck.Entities:
Keywords: Corrosion; Fatigue fracture; Total hip arthroplasty; Trunnionosis
Year: 2020 PMID: 35236443 PMCID: PMC8796592 DOI: 10.1186/s42836-020-00047-3
Source DB: PubMed Journal: Arthroplasty ISSN: 2524-7948
Fig. 1Two types of hip neck fractures: head-neck fractures are located at the proximal part of the neck and neck-shoulder fractures at the distal part of the neck
Critical appraisal instrument
| Study | ||
|---|---|---|
| Domains | Questions | Answer |
| Selection | 1. Does the patient(s) represent(s) the whole experience of the investigator (center) or is the selection method unclear to the extent that other patients with similar presentation may not have been reported? | |
| Ascertainment | 2. Was the outcome adequately ascertained? | |
| Causality | 3. Were other alternative causes that may explain the observation ruled out? | |
| 4. Was follow-up long enough for outcomes to occur? | ||
| 5. Was there further analysis done to investigate the cause of fracture? | ||
| Reporting | 6. Is the case(s) described with sufficient details to allow other investigators to replicate the research or to allow practitioners make inferences related to their own practice? | |
| – or +/− or + | ||
Fig. 2PRISMA flowchart of the review process of our study
Characteristics of the 33 studies included
| Study | Year | Number of fractures | Fracture location | Mean age (y) - weight (kg) - BMI (kg/m2) | Gender | Type of stem | Methodological score | |
|---|---|---|---|---|---|---|---|---|
| 1 | Peterson | 2019 | 1 | Head-neck: 1 | 45 y - 94 kg - 27 kg/m2 | 1–0 | C-taper stem | + |
| 2 | Takai | 2019 | 1 | Neck-shoulder: 1 | 57 y – 70 kg – 27 kg/m2 | 0–1 | AHFIX Q | + |
| 3 | Ryniewicz | 2018 | 1 | Neck-shoulder: 1 | 73 y -? -? | 1–0 | Aura II | + |
| 4 | Bolland | 2016 | 26 | Head-neck: 20 Neck-shoulder: 6 | 69 y - 107 kg -? | 20–6 | Exeter | +/− |
| 5 | Facek | 2016 | 1 | Neck-shoulder: 1 | 70 y - 71 kg -? | 1–0 | Exeter | + |
| 6 | Morlock | 2016 | 3 | Head-neck: 3 | 63 y - 109 kg - 33 kg/m2 | 3–0 | BiMetric | + |
| 7 | Reito | 2015 | 3 | Head-neck: 3 | 79 y - 99 kg - 32 kg/m2 | 2–1 | Exeter V40 | +/− |
| 7 | Spanyer | 2015 | 2 | Head-neck:1 Neck-shoulder: 1 | 52 y -? - 28 kg/m2 | 2–0 | Accolade1 | +/− |
| 8 | Yoshimoto | 2015 | 2 | Neck-shoulder: 2 | 58 y - 74 kg - 25 kg/m2 | 2–0 | Kyocera | + |
| 10 | Banerjee | 2014 | 1 | Head-neck: 1 | 55 y - 81 kg - 47 kg/m2 | 1–0 | RMHS | +/− |
| 11 | Baratz | 2014 | 1 | Head-neck: 1 | 73 y - 89 kg - 32 kg/m2 | 1–0 | Meridian | +/− |
| 12 | Hamlin | 2014 | 1 | Head-neck: 1 | 76 y – 141 kg - 45 kg/m2 | 1–0 | Exeter | +/− |
| 13 | Jang | 2013 | 1 | Neck-shoulder: 1 | 74 y - 107 kg - 35 kg/m2 | 1–0 | Corail | + |
| 14 | Lizano-Diez | 2013 | 1 | Head-neck: 1 | 67 y - 110 kg - 32 kg/m2 | 1–0 | Bicontact | +/− |
| 15 | Nganbe | 2013 | 1 | Neck-shoulder: 1 | 75 y -? -? | 1–0 | Lord | + |
| 16 | Haasper | 2012 | 2 | Neck-shoulder: 2 | 50 y -? -? | 1–1 | Collum Femoris Preserving (CFP) | +/− |
| 17 | Morley | 2012 | 1 | Head-neck: 1 | 53 y - 110 kg - 32 kg/m2 | 1–0 | C-stem | +/− |
| 18 | Bos | 2011 | 1 | Head-neck: 1 | 58 y - 92 kg - 30 kg/m2 | 1–0 | Elite Plus | +/− |
| 19 | Garg | 2011 | 1 | Head-neck: 1 | 31 y - 78 kg - 26 kg/m2 | 1–0 | Anatomic Medullary Locking | – |
| 20 | O’neill | 2011 | 1 | Neck-shoulder: 1 | 76 y -? - 28 kg/m2 | 0–1 | Exeter revision | – |
| 21 | Unnanuntana | 2010 | 2 | Head-neck: 2 | 60 y - 88 kg - 29 kg/m2 | 2–0 | Anatomic Medullary Locking | + |
| 22 | Lam | 2008 | 4 | Head-neck: 4 | 80 y - 103 kg - 34 kg/m2 | ? -? | Omnifit | +/− |
| 23 | Briant-Evans | 2007 | 2 | Neck-shoulder: 2 | 55 y -? - 39 kg/m2 | 0–2 | Corin Eurocone Taper-Fit CDH | +/− |
| 24 | Grivas | 2007 | 1 | Neck-shoulder: 1 | 68 y - 70 kg - 26 kg/m2 | 1–0 | SEM3 | + |
| 25 | Harvie | 2007 | 1 | Neck-shoulder: 1 | 73 y - 90 kg - 30 kg/m2 | 1–0 | JRI-Furlong | + |
| 26 | Botti | 2005 | 1 | Head-neck: 1 | 85 y - 90 kg - 28 kg/m2 | 1–0 | Anatomic Medullary Locking | +/− |
| 27 | Morgan-Hough | 2004 | 1 | Neck-shoulder: 1 | 64 y - 74 - 24 kg/m2 | 1–0 | JRI Limited | + |
| 28 | Vatani | 2002 | 9 | Head-neck: 9 | 62 y - 79 kg - 26 kg/m2 | 8–1 | Modular Charnley by Medical Tec. | + |
| 29 | Lee | 2001 | 2 | Neck-shoulder: 2 | 65 y - 104 kg - 33 kg/m2 | 2–0 | Exactech Opteon | +/− |
| 30 | Allcock | 1997 | 1 | Neck-shoulder: 1 | 52 y - 90 kg - 26 kg/m2 | 1–0 | BHC prosthesis | – |
| 31 | Artime | 1997 | 1 | Head-neck: 1 | 37 y -? -? | 1–0 | Lord | – |
| 32 | Gilbert | 1994 | 2 | Head-neck: 2 | 65 y - 112 kg - 36 kg/m2 | 2–0 | PCA | + |
| 33 | Barrack | 1993 | 1 | Head-neck:1 | 66 y - 91 kg - 27 kg/m2 | 1–0 | Osteonics | – |
| 80 | Head-neck: 55 Neck-shoulder: 25 | 65 y - 98 kg - 30 kg/m2 | 63–13 (4?) |