Faulty design increases the risk of neck fracture in a hip prosthesis.

Nosrat Vatani1, Daniel Comando, José Acuña, Daniel Prieto, Horacio Caviglia.   

Abstract

From October 1993 to May 1994, we performed 35 total hip arthroplasties (20 women) using a Charnley type modular 28 head manufactured by Medical Tee Brazil. 9 fractures through the neck of the stem prostheses occurred, all but 1 in men, after mean 4 (0.3-6) years. All fractures were due to an inadequate confluent radius causing abnormal transmission forces through the neck. The higher the load (men, heavy body weight), the greater the risk of fracture. The metallographic, chemical and microhardness analyses showed no abnormalities. We ascribe these fractures to faulty design of the stem prosthesis.

Entities:  

Mesh:

Year:  2002        PMID: 12440493     DOI: 10.1080/000164702321022776

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  4 in total

Review 1.  Catastrophic Femoral Neck Failure after THA with the Accolade(®) I Stem in Three Patients.

Authors:  Jonathon Spanyer; Jennifer Hines; Christopher Maxwell Beaumont; Jonathan Yerasimides
Journal:  Clin Orthop Relat Res       Date:  2015-07-08       Impact factor: 4.176

Review 2.  The incidence of implant fractures after total hip arthroplasty.

Authors:  Patrick Sadoghi; Wolfram Pawelka; Michael C Liebensteiner; Alexandra Williams; Andreas Leithner; Gerold Labek
Journal:  Int Orthop       Date:  2014-01       Impact factor: 3.075

3.  Neck fracture of a cementless forged titanium alloy femoral stem following total hip arthroplasty: a case report and review of the literature.

Authors:  Theodoros B Grivas; Olga D Savvidou; Spyridon A Psarakis; Pierre-Francois Bernard; George Triantafyllopoulos; Ioannis Kovanis; Panagiotis Alexandropoulos
Journal:  J Med Case Rep       Date:  2007-12-06

4.  Femoral stem fractures after primary and revision hip replacements: A single-center experience.

Authors:  Alper Köksal; Ali Öner; Osman Çimen; Osman Emre Aycan; Hakan Akgün; Furkan Yapıcı; Yalkın Çamurcu
Journal:  Jt Dis Relat Surg       Date:  2020