Samik Banerjee, Jeffrey J Cherian, James V Bono, Steven M Kurtz, Rudolph Geesink, R Michael Meneghini, Ronald E Delanois, Michael A Mont.   

Abstract

Unfavorable outcomes from trunnion fretting and corrosion damage have been reported in the literature, gross failures of tapers in primary total hip arthroplasties have been less frequently reported. We report on 5 patients, who presented with gross trunnion failures of modular metal-on-polyethylene or ceramic-on-polyethylene bearings from 5 implant manufacturers, all necessitating revision surgery. None of these patients had an antecedent history of trauma, and the majority presented with pain or instability. No common factor was identified that may be predictive of these type of failures. Since there were 5 different stem designs, this suggests that it may be a rare generic phenomenon occurring with multiple designs. Currently, further investigations are necessary, including retrieval analysis, to identify risk factors that may predispose to such failures.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  failure; fracture; total hip arthroplasty; trunnion; trunnionosis; wear

Mesh:

Substances:

Year:  2014        PMID: 25499678     DOI: 10.1016/j.arth.2014.11.023

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  20 in total

Review 1.  What Factors Drive Taper Corrosion?

Authors:  Robin Pourzal; Hannah J Lundberg; Deborah J Hall; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2018-03-30       Impact factor: 4.757

Review 2.  Catastrophic Femoral Neck Failure after THA with the Accolade(®) I Stem in Three Patients.

Authors:  Jonathon Spanyer; Jennifer Hines; Christopher Maxwell Beaumont; Jonathan Yerasimides
Journal:  Clin Orthop Relat Res       Date:  2015-07-08       Impact factor: 4.176

3.  Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.

Authors:  Kenneth L Urish; Brian R Hamlin; Anton Y Plakseychuk; Timothy J Levison; Genymphas B Higgs; Steven M Kurtz; Anthony M DiGioia
Journal:  J Arthroplasty       Date:  2017-04-08       Impact factor: 4.757

4.  Do Well-functioning THAs Retrieved at Autopsy Exhibit Evidence of Fretting and Corrosion?

Authors:  Jeffrey Lange; Amanda Wach; Chelsea N Koch; Robert H Hopper; Henry Ho; Charles A Engh; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-10       Impact factor: 4.176

5.  Adventure sports and sexual freedom hip replacement: the tripolar hip.

Authors:  James W Pritchett
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-06-28

6.  Head-stem trunnion dissociation due to corrosion in total hip arthroplasty.

Authors:  Michael R Whitehouse
Journal:  Ann Transl Med       Date:  2017-09

Review 7.  Gross trunnion failure in metal on polyethylene total hip arthroplasty-a systematic review of literature.

Authors:  Tungish Bansal; Sameer Aggarwal; Mandeep Singh Dhillon; Sandeep Patel
Journal:  Int Orthop       Date:  2020-01-04       Impact factor: 3.075

8.  Acute trunnion failure of a TMZF alloy stem with large diameter femoral heads.

Authors:  E J Fahey; E Pomeroy; F E Rowan
Journal:  J Orthop       Date:  2020-01-10

9.  What Is the Risk of THA Revision for ARMD in Patients with Non-metal-on-metal Bearings? A Study from the Australian National Joint Replacement Registry.

Authors:  R N de Steiger; Alesha Hatton; Yi Peng; Stephen Graves
Journal:  Clin Orthop Relat Res       Date:  2020-06       Impact factor: 4.755

Review 10.  Trunnionosis in total hip arthroplasty: a review.

Authors:  Jaydev B Mistry; Morad Chughtai; Randa K Elmallah; Aloise Diedrich; Sidney Le; Melbin Thomas; Michael A Mont
Journal:  J Orthop Traumatol       Date:  2016-02-11
View more