High failure rate of a new pressfit cup in mid-term follow-up.

Steffen Brodt1, Georg Matziolis, Christoph Windisch, Andreas Gosse, Matthias Spalteholz, Ralf Herbert Gahr.   

Abstract

BACKGROUND: Unacceptably high rates of loosening of a novel cementless cup system (Mathys, SeleXys TH+) have been reported. In this study, our mid-term-results with the SeleXys TH+ cup are compared with an established pressfit cup (Ceraver, Cerafit-R).
METHODS: Between 2008 and 2011, a total of 407 implantations, 19 of them bilaterally, were conducted on 388 patients with the implants in question. A total of 280 cups were available for follow-up. Of these, 160 SeleXys TH+ cups and 120 Cerafit-R cups had been used.
RESULTS: The mean follow-up period was 48.6 months (min. 28, max. 75). A total of 18 cases of loosening were found in the SeleXys TH+ group. Of these, one was of a traumatic nature and one was of septic origin. This left 16 cases of aseptic, atraumatic loosening (10% loosening rate). In the Cerafit-R group, only two cases (1.7%) of loosening were seen over the same period. Both were of traumatic origin. No cases of aseptic, atraumatic loosening were seen in the follow-up period. Further, significantly more cases of loosening were observed in women than in men with the SeleXys cup.
CONCLUSIONS: The SeleXys TH+ cup shows an unacceptably high failure rate after six years. The time of revision is not limited to the first months after implantation, which means that close clinical and radiological controls will continue to be necessary in future in patients treated with SeleXys TH+ cups. Level of Evidenve: III (retrospective cohort study).

Entities:  

Mesh:

Year:  2015        PMID: 26152245     DOI: 10.1007/s00264-015-2872-x

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  27 in total

1.  Bone ingrowth on a smooth-surfaced hydroxyapatite-coated acetabular cup.

Authors:  Y Y Chung; H D Kim; K S Kim
Journal:  Int Orthop       Date:  2002-08-01       Impact factor: 3.075

2.  Stability of press-fit acetabular cups.

Authors:  E Adler; S A Stuchin; F J Kummer
Journal:  J Arthroplasty       Date:  1992-09       Impact factor: 4.757

3.  Deformation characteristics and eigenfrequencies of press-fit acetabular cups.

Authors:  Arne Hothan; Gerd Huber; Cornelius Weiss; Norbert Hoffmann; Michael Morlock
Journal:  Clin Biomech (Bristol, Avon)       Date:  2011-01       Impact factor: 2.063

4.  The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis.

Authors:  Paul Goebel; Daniel Kluess; Jan Wieding; Robert Souffrant; Horst Heyer; Manuela Sander; Rainer Bader
Journal:  J Orthop Sci       Date:  2013-02-03       Impact factor: 1.601

5.  Observations on the effect of movement on bone ingrowth into porous-surfaced implants.

Authors:  R M Pilliar; J M Lee; C Maniatopoulos
Journal:  Clin Orthop Relat Res       Date:  1986-07       Impact factor: 4.176

6.  The transgluteal approach to the hip joint.

Authors:  R Bauer; F Kerschbaumer; S Poisel; W Oberthaler
Journal:  Arch Orthop Trauma Surg       Date:  1979-10

Review 7.  Thirty years of experience with alumina-on-alumina bearings in total hip arthroplasty.

Authors:  Didier Hannouche; Amine Zaoui; Frédéric Zadegan; Laurent Sedel; Rémy Nizard
Journal:  Int Orthop       Date:  2010-12-30       Impact factor: 3.075

8.  Early loosening of a press-fit cup with ceramic-on-ceramic articulation: our early results.

Authors:  Daniël Haverkamp; Stijn Westerbos; Martin M Campo; Ritsert H Boonstra; G H Rob Albers; Harm M van der Vis
Journal:  Arch Orthop Trauma Surg       Date:  2013-10-02       Impact factor: 3.067

9.  Effect of stem modulus in a total hip arthroplasty model.

Authors:  G L Maistrelli; V Fornasier; A Binnington; K McKenzie; V Sessa; I Harrington
Journal:  J Bone Joint Surg Br       Date:  1991-01

10.  Long-term results of cementless hip arthroplasty with ceramic-on-ceramic articulation.

Authors:  Marek Synder; Marek Drobniewski; Marcin Sibiński
Journal:  Int Orthop       Date:  2012-08-19       Impact factor: 3.075

View more
  3 in total

1.  Comment to "Ceramic-on-ceramic articulation in press-fit total hip arthroplasty as a potential reason for early failure, what about the survivors: a ten year follow-up".

Authors:  Zhi-Qiang Fan; Xin Xu
Journal:  Int Orthop       Date:  2021-03-16       Impact factor: 3.075

2.  Midterm analysis of the seleXys cup with ceramic inlay.

Authors:  Tamas Halasi; David Kieser; Christopher M A Frampton; Gary J Hooper
Journal:  Arthroplast Today       Date:  2016-11-09

3.  Ceramic-on-ceramic articulation in press-fit total hip arthroplasty as a potential reason for early failure, what about the survivors: a ten year follow-up.

Authors:  J van Loon; A M J S Vervest; H M van der Vis; I N Sierevelt; D C Baas; K T M Opdam; G M M J Kerkhoffs; D Haverkamp
Journal:  Int Orthop       Date:  2021-01-18       Impact factor: 3.075

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.