The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis.

Paul Goebel1, Daniel Kluess, Jan Wieding, Robert Souffrant, Horst Heyer, Manuela Sander, Rainer Bader.   

Abstract

BACKGROUND: To increase the range of motion of total hip endoprostheses, prosthetic heads need to be enlarged, which implies that the cup and/or liner thickness must decrease. This may have negative effects on the wear rate, because the acetabular cups and liners could deform during press-fit implantation and hip joint loading. We compared the metal cup and polyethylene liner deformations that occurred when different wall thicknesses were used in order to evaluate the resulting changes in the clearance of the articulating region.
METHODS: A parametric finite element model utilized three cup and liner wall thicknesses to analyze cup and liner deformations after press-fit implantation into the pelvic bone. The resultant hip joint force during heel strike was applied while the femur was fixed, accounting for physiological muscle forces. The deformation behavior of the liner under joint loading was therefore assessed as a function of the head diameter and the resulting clearance.
RESULTS: Press-fit implantation showed diametral cup deformations of 0.096, 0.034, and 0.014 mm for cup wall thicknesses of 3, 5, and 7 mm, respectively. The largest deformations (average 0.084 ± 0.003 mm) of liners with thicknesses of 4, 6, and 8 mm occurred with the smallest cup wall thickness (3 mm). The smallest liner deformation (0.011 mm) was obtained with largest cup and liner wall thicknesses. Under joint loading, liner deformations in thin-walled acetabular cups (3 mm) reduced the initial clearance by about 50 %.
CONCLUSION: Acetabular press-fit cups with wall thicknesses of ≤5 mm should only be used in combination with polyethylene liners >6 mm thick in order to minimize the reduction in clearance.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23377753     DOI: 10.1007/s00776-012-0340-7

Source DB:  PubMed          Journal:  J Orthop Sci        ISSN: 0949-2658            Impact factor:   1.601


  4 in total

1.  Correlation between in vivo stresses and oxidation of UHMWPE in total hip arthroplasty.

Authors:  M Regis; P Bracco; L Giorgini; S Fusi; P Dalla Pria; L Costa; C Schmid
Journal:  J Mater Sci Mater Med       Date:  2014-06-29       Impact factor: 3.896

2.  Quantification of clearance and creep in acetabular wear measurements.

Authors:  Mo Saffarini; Thomas Gregory; Eric Vandenbussche
Journal:  Ann Transl Med       Date:  2016-04

3.  High failure rate of a new pressfit cup in mid-term follow-up.

Authors:  Steffen Brodt; Georg Matziolis; Christoph Windisch; Andreas Gosse; Matthias Spalteholz; Ralf Herbert Gahr
Journal:  Int Orthop       Date:  2015-07-08       Impact factor: 3.075

4.  Total hip arthroplasty for dysplastic coxarthrosis using a cementless Wagner Cone stem.

Authors:  Giuseppe Solarino; Giovanni Vicenti; Andrea Piazzolla; Francesco Maruccia; Angela Notarnicola; Biagio Moretti
Journal:  J Orthop Traumatol       Date:  2021-04-17
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.