Ceramic-on-ceramic articulation in press-fit total hip arthroplasty as a potential reason for early failure, what about the survivors: a ten year follow-up.

J van Loon1,2,3, A M J S Vervest3, H M van der Vis1, I N Sierevelt1,4, D C Baas3, K T M Opdam2, G M M J Kerkhoffs2, D Haverkamp5.   

Abstract

PURPOSE: In press-fit total hip arthroplasty (THA), primary stability is needed to avoid micromotion and hereby aseptic loosening, the main reason for early revision. High aseptic loosening revision rates of the seleXys TH+ cup (Mathys Medical) with Ceramys ceramic-on-ceramic (CoC) bearing are seen in literature. Since CoC is presumed to overcome long-term wear-related revisions, the reason for early failure of this cup is important to clarify. The aim is to investigate its ten year outcomes and differentiate between potential causes and identify risk factors for aseptic loosening.
METHODS: Retrospective screening of a prospectively documented series of 315 THAs was performed. Primary outcome was cumulative incidence of cup revision due to aseptic loosening. Secondary outcomes were component revision and reoperation. Additionally, potential predictive factors for aseptic loosening were evaluated.
RESULTS: At the median follow-up of 9.7 years [IQR 4.4; 10.3], 48 TH+ (15.2%) were revised due to aseptic loosening. Competing risk analysis showed a ten year cumulative incidence of cup revision due to aseptic loosening of 15.6% (95% CI 12.0-20.2). Stabilization of early revision rates was observed, following a high rate of respectively 81.3% (n = 39) and 95.8% (n = 46) within the first two and three years. No significant predictive factors for aseptic loosening were found.
CONCLUSION: The ten year results of seleXys TH+ cup with Ceramys CoC bearing showed an unacceptable high aseptic loosening rate, which stabilized over time after a high early failure incidence. This could be attributed to a problem with osseointegration during the transition of primary to definitive stability.

Entities:  

Keywords:  Aseptic loosening; Ceramic-on-ceramic; Early failure; Primary stability; Total hip arthroplasty

Year:  2021        PMID: 33459828     DOI: 10.1007/s00264-020-04895-1

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  19 in total

1.  Comparison of preoperative radiographs and intraoperative findings of fixation of hemispheric porous-coated sockets.

Authors:  P Udomkiat; Z Wan; L D Dorr
Journal:  J Bone Joint Surg Am       Date:  2001-12       Impact factor: 5.284

2.  Porous titanium construct cup compared to porous coated titanium cup in total hip arthroplasty. A randomised controlled trial.

Authors:  Mats Salemyr; Olle Muren; Thomas Eisler; Henrik Bodén; Ghazi Chammout; André Stark; Olof Sköldenberg
Journal:  Int Orthop       Date:  2014-10-22       Impact factor: 3.075

3.  Changes in periacetabular bone mineral density five years after resurfacing hip arthroplasty versus conventional total hip arthroplasty.

Authors:  Davey Mjm Gerhardt; José Mh Smolders; Elisabeth A Roovers; Ton Ajm Rijnders; Job Lc van Susante
Journal:  Hip Int       Date:  2018-11-14       Impact factor: 2.135

4.  Good diagnostic performance of early migration as a predictor of late aseptic loosening of acetabular cups: results from ten years of follow-up with Roentgen stereophotogrammetric analysis (RSA).

Authors:  Marc J Nieuwenhuijse; Edward R Valstar; Bart L Kaptein; Rob G H H Nelissen
Journal:  J Bone Joint Surg Am       Date:  2012-05-16       Impact factor: 5.284

5.  High failure rate of a new pressfit cup in mid-term follow-up.

Authors:  Steffen Brodt; Georg Matziolis; Christoph Windisch; Andreas Gosse; Matthias Spalteholz; Ralf Herbert Gahr
Journal:  Int Orthop       Date:  2015-07-08       Impact factor: 3.075

6.  Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years.

Authors:  Stephan M Röhrl; Ming Guo Li; Kjell-Gunnar Nilsson; Bo Nivbrant
Journal:  Acta Orthop       Date:  2007-12       Impact factor: 3.717

7.  Poor outcome of a spherical pressfit cup with a modern ceramic liner: a prospective cohort study of 181 cups.

Authors:  Thomas Ilchmann; Lukas Zwicky; Silke Gersbach; Martin Clauss
Journal:  Hip Int       Date:  2014-02-14       Impact factor: 2.135

8.  Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial.

Authors:  Richard W McCalden; Steven J MacDonald; Cecil H Rorabeck; Robert B Bourne; David G Chess; Kory D Charron
Journal:  J Bone Joint Surg Am       Date:  2009-04       Impact factor: 5.284

9.  Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients.

Authors:  H Malchau; J Kärrholm; Y X Wang; P Herberts
Journal:  Acta Orthop Scand       Date:  1995-10

10.  Assessing the Acetabular Cup Implant Primary Stability by Impact Analyses: A Cadaveric Study.

Authors:  Adrien Michel; Romain Bosc; Jean-Paul Meningaud; Philippe Hernigou; Guillaume Haiat
Journal:  PLoS One       Date:  2016-11-28       Impact factor: 3.240

View more
  2 in total

1.  Comparison of early migration patterns between a ceramic and polyethylene liner in uncemented Trabecular Titanium cups: a 2-year randomized controlled trial of 52 hips using radiostereometric analysis.

Authors:  Amanda D Klaassen; Elisabeth A Schäffer; Nienke W Willigenburg; Loes W A H Van Beers; Vanessa A B Scholtes; Victor P M Van der Hulst; Lennard A Koster; Bart L Kaptein; Dirk Jan F Moojen; Rudolf W Poolman
Journal:  Acta Orthop       Date:  2022-04-28       Impact factor: 3.925

2.  Clinical results at 10 years of minimum follow-up with the ABG 2 hip arthroplasty, matched with ceramic-on-ceramic bearings.

Authors:  Remy Coulomb; Jad Mansour; Jérome Essig; Gérard Asencio; Pascal Kouyoumdjian
Journal:  SICOT J       Date:  2022-08-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.