Early loosening of a press-fit cup with ceramic-on-ceramic articulation: our early results.

Daniël Haverkamp1, Stijn Westerbos, Martin M Campo, Ritsert H Boonstra, G H Rob Albers, Harm M van der Vis.   

Abstract

INTRODUCTION: In this study, we present the short-term results of the Selexys TH+ cup with the Ceramys inlay which is a press-fit cup with a ceramic-on-ceramic articulation. (Mathys, Bettlach, Switzerland). We compared the results with a retrospective-matched control group with a Delta PF cup (Lima, Udine, Italy), which is also a press-fit cup with a ceramic-on-ceramic articulation.
MATERIALS AND METHODS: 257 elective hip arthroplasties with the Selexys TH+ cup in 250 patients placed in 2009 and 2010 were analyzed and compared with a control group retrospective analysis of the uncemented Delta PF cup (Lima, Udine, Italy) placed in 2007 and 2008 in 208 patients (222 hips). Surgical technique and surgeons were identical in both groups.
RESULTS: During a follow-up period of 3-21 months, 19 aseptic loosenings (7.4 %) were found for the Selexys TH+ cup. The survival plotted by a Kaplan-Meier curve shows a 1-year survival of 87.4 %. The Lima Delta PF cup showed a 1-year survival of 99.5 %. Failure analysis showed no clear explanation for this early loosening.
CONCLUSION: The Selexys TH+ cup combined with the Ceramys ceramic-on-ceramic inlay coupling show an unacceptable high early revision rate. Therefore, we advice against using this combination.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24085556     DOI: 10.1007/s00402-013-1861-z

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  4 in total

1.  Comment to "Ceramic-on-ceramic articulation in press-fit total hip arthroplasty as a potential reason for early failure, what about the survivors: a ten year follow-up".

Authors:  Zhi-Qiang Fan; Xin Xu
Journal:  Int Orthop       Date:  2021-03-16       Impact factor: 3.075

2.  High failure rate of a new pressfit cup in mid-term follow-up.

Authors:  Steffen Brodt; Georg Matziolis; Christoph Windisch; Andreas Gosse; Matthias Spalteholz; Ralf Herbert Gahr
Journal:  Int Orthop       Date:  2015-07-08       Impact factor: 3.075

3.  Midterm analysis of the seleXys cup with ceramic inlay.

Authors:  Tamas Halasi; David Kieser; Christopher M A Frampton; Gary J Hooper
Journal:  Arthroplast Today       Date:  2016-11-09

4.  Results of 198 primary total hip arthroplasties using the Delta PF-FIT system with ceramic-on-ceramic articulating surfaces with average seven years follow up.

Authors:  Petr Fulin; David Pokorny; Jan Hert; Antonin Sosna
Journal:  BMC Musculoskelet Disord       Date:  2020-05-19       Impact factor: 2.362