| Literature DB >> 26150687 |
Michał Biegała1, Teresa Jakubowska1, Karolina Markowska2.
Abstract
A comparative analysis of the mean glandular doses was conducted in 100 female patients who underwent screening mammography in 2011 and 2013. Siemens Mammomat Novation with the application of the W/Rh anode/filter combination was used in 2011, whereas in 2013 anode/filter combination was Mo/Mo or Mo/Rh. The functioning of mammography was checked and the effectiveness of the automatic exposure control (AEC) system was verified by measuring compensation of changes in the phantom thickness and measuring tube voltage. On the base of exposure parameters, an average glandular dose for each of 100 female patients was estimated. The images obtained by using AEC system had the acceptable threshold contrast visibility irrespective of the applied anode/filter combination. Mean glandular doses in the females, examined with the application of the W/Rh anode/filter combination, were on average 23.6% lower than that of the Mo/Mo or Mo/Rh anode/filter combinations. It is recommended to use a combination of the W/Rh anode /filter which exhibited lower mean glandular doses.Entities:
Keywords: Automatic exposure control system; average glandular dose; digital mammography; threshold contrast visibility
Year: 2015 PMID: 26150687 PMCID: PMC4471644 DOI: 10.4103/0971-6203.152251
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1CDMAM phantom (image from: www.artinis.com)
Measurements of compensation of changes in the phantom thickness and tube voltage in 2011
Measurements of compensation of changes in the phantom thickness and tube voltage in 2013
Measurements of the threshold contrast visibility for clinical exposure in 2011 and 2013
Figure 2CDMAM phantom with 4 plates PMMA (photo: M. Biegała)
Measured average glandular dose for clinical exposure for the breast thickness simulated by PMMA
Figure 3Relationship between an average glandular dose and miliampere second value (mAs) for particular voltage values in patients examined in 2011
Figure 4Relationship between an average glandular dose and miliampere second value (mAs) for particular voltage values in patients examined in 2013. The average doses can be compared only if the AEC system works properly and the image is of good quality. AEC = Automatic exposure control
Glandular dose D in one examination procedure [mGy]
Exposure parameters in 2011 and 2013
Average glandular doses in various projections in 2011 and 2013