Bing-Bo Hou, Yan Yao1, Ling-Min Wu, Yu Qiao, Li-Hui Zheng, Li-Gang Ding, Gang Chen, Shu Zhang.
Abstract
BACKGROUND: Nonfluoroscopic three-dimensional electroanatomical system is widely used nowadays, but X-ray remains indispensable for complex electrophysiology procedures. This study aimed to evaluate the value of optimized parameter setting and different projection position to reduce X-ray radiation dose rates.Entities:
Mesh:
Year: 2015 PMID: 25947395 PMCID: PMC4831539 DOI: 10.4103/0366-6999.156079
Source DB: PubMed Journal: Chin Med J (Engl) ISSN: 0366-6999 Impact factor: 2.628
The X-ray parameter of the default and optimal setting
| Parameter item | Default setting | Optimal setting |
|---|---|---|
| Preset voltage (kV) | 81 | 73 |
| Pulse width (ms) | 8.0 | 6.4 |
| Critical voltage (kV) | 90 | 102 |
| Peak voltage (kV) | 109 | 109 |
| Noise reduction | Normal | Smooth |
| Edge enhancement (%) | 10 | 20 |
| Fluoro type | Pulsed | Pulsed |
| Pulse rate (P/s) | 15 | 6 |
| Minimum CU-filter (mm) | 0.6 | 0.6 |
| Maximum CU-filter (mm) | 0.9 | 0.9 |
| Dose (nGy/p) | 29 | 6 |
The baseline characteristics of the patients
| Characteristics ( | Results |
|---|---|
| Age (years) | 54.7 ± 11.1 |
| Males, | 85 (81.0) |
| Weight (kg) | 75.1 ± 11.3 |
| Height (cm) | 171.0 ± 8.1 |
| Body surface area (m2) | 1.85 ± 0.18 |
| BMI (kg/m2) | 25.6 ± 2.8 |
| Diagnosis, | |
| PAF | 54 (51.4) |
| PeAF | 32 (30.5) |
| VT | 7 (6.7) |
| AAFL | 12 (11.4) |
PAF: Paroxysmal atrial fibrillation; PeAF: Persistent Atrial Fibrillation; VT: Ventricular Tachycardia; AAFL: Atypical atrial flutter; BMI: Body mass index.
X-ray radiation dose rate in different projection position and setting
| Items | Default setting (Gy/min) | Optimized setting (Gy/min) | |
|---|---|---|---|
| AP | 13.6 ± 5.7 | 1.7 ± 0.8 | <0.001 |
| LAO 30° | 19.7 ± 6.3 | 2.5 ± 1.0 | <0.001 |
| LAO 45° | 22.1 ± 6.6 | 3.1 ± 1.4 | <0.001 |
| 51.26 | 39.5 | ||
| <0.001 | <0.001 | ||
LAO: Left anterior oblique; AP: Anterior-posterior.