Improving safety in the electrophysiology laboratory using a simple radiation dose reduction strategy: a study of 1007 radiofrequency ablation procedures.

Dominic P S Rogers1, Fiona England, Konstantin Lozhkin, Martin D Lowe, Pier D Lambiase, Anthony W C Chow.   

Abstract

BACKGROUND: The use of fluoroscopic screening involves exposure to ionising radiation for both patients and operators.
OBJECTIVE: To assess the effects of radiation dose reduction manoeuvres (DRM) during radiofrequency ablation (RFA) procedures.
DESIGN: Prospective study of DRM.
SETTING: Tertiary cardiac centre. Interventions Two DRM were combined: removal of the secondary radiation grid and programming an ultra-low pulsed fluoroscopy rate. These methods were assessed using an anthropomorphic phantom model to measure skin entrance dose rates. Procedures were classified as complex (ablation of atrial fibrillation, ventricular tachycardia or complex congenital heart disease arrhythmias) or simple (all other RFA). MAIN OUTCOME MEASURES: Dose area product and screening times were compared for ablations performed before and after DRM. Equivalent doses to organs and malignancy risk were determined by computer modelling.
RESULTS: Over a 39-month period, 1007 ablation procedures were performed (631 simple, 376 complex). Radiation dose was significantly reduced after DRM for both simple (20.4±26.9 Gycm(2) vs 8.0±10.3 Gycm(2), p<0.00001) and complex ablations (63.3±50.1 Gycm(2) vs 32.8±31.7 Gycm(2), p<0.00001) with no difference in screening times. The mean lifetime risk of fatal cancer attributable to radiation exposure per million procedures was reduced from 182 to 68 for simple ablations and from 440 to 155 for complex ablations.
CONCLUSIONS: Significant reductions in radiation exposure during RFA were achieved using simple DRM, corresponding to a two-thirds reduction of the risk of excess fatal malignancy.

Entities:  

Mesh:

Year:  2010        PMID: 21036800     DOI: 10.1136/hrt.2010.204222

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  11 in total

Review 1.  Radiation exposure and safety for the electrophysiologist.

Authors:  Sabine Ernst; Isabel Castellano
Journal:  Curr Cardiol Rep       Date:  2013-10       Impact factor: 2.931

2.  Reduction of radiation exposure during ablation of atrial fibrillation.

Authors:  Ralph Schneider; Jörg Lauschke; Cindy Schneider; Tina Tischer; Aenne Glass; Dietmar Bänsch
Journal:  Herz       Date:  2015-05-20       Impact factor: 1.443

3.  Radiation safety protocol using real-time dose reporting reduces patient exposure in pediatric electrophysiology procedures.

Authors:  Akash R Patel; Jamie Ganley; Xiaowei Zhu; Jonathan J Rome; Maully Shah; Andrew C Glatz
Journal:  Pediatr Cardiol       Date:  2014-05-07       Impact factor: 1.655

4.  Reduction of Fluoroscopy Time and Radiation Dosage During Catheter Ablation for Atrial Fibrillation.

Authors:  Kenichiro Yamagata; Bashar Aldhoon; Josef Kautzner
Journal:  Arrhythm Electrophysiol Rev       Date:  2016-08

Review 5.  Anatomic Challenges In Pediatric Catheter Ablation.

Authors:  Thomas A Pilcher Md; Elizabeth V Saarel Md
Journal:  J Atr Fibrillation       Date:  2014-08-31

6.  Optimized Fluoroscopy Setting and Appropriate Project Position Can Reduce X-ray Radiation Doses Rates during Electrophysiology Procedures.

Authors:  Bing-Bo Hou; Yan Yao; Ling-Min Wu; Yu Qiao; Li-Hui Zheng; Li-Gang Ding; Gang Chen; Shu Zhang
Journal:  Chin Med J (Engl)       Date:  2015-05-05       Impact factor: 2.628

7.  Radiation-associated plasmacytoma following catheter ablation for atrial fibrillation.

Authors:  Justin Chow; Theresa Trotter; Peter Duggan; Stephen B Wilton
Journal:  HeartRhythm Case Rep       Date:  2016-12-31

8.  Comparison of pediatric radiation dose and vessel visibility on angiographic systems using piglets as a surrogate: antiscatter grid removal vs. lower detector air kerma settings with a grid - a preclinical investigation.

Authors:  Keith J Strauss; John M Racadio; Todd A Abruzzo; Neil D Johnson; Manish N Patel; Kamlesh U Kukreja; Mark J H den Hartog; Bart P A Hoonaert; Rami A Nachabe
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

9.  Extremely low-frame-rate digital fluoroscopy in catheter ablation of atrial fibrillation: A comparison of 2 versus 4 frame rate.

Authors:  Ji Hyun Lee; Jun Kim; Minsu Kim; Jongmin Hwang; You Mi Hwang; Joon-Won Kang; Gi-Byoung Nam; Kee-Joon Choi; You-Ho Kim
Journal:  Medicine (Baltimore)       Date:  2017-06       Impact factor: 1.817

10.  Evaluation of a new very low dose imaging protocol: feasibility and impact on X-ray dose levels in electrophysiology procedures.

Authors:  Felix Bourier; Tilko Reents; Sonia Ammar-Busch; Alessandra Buiatti; Marc Kottmaier; Verena Semmler; Marta Telishevska; Amir Brkic; Christian Grebmer; Carsten Lennerz; Christof Kolb; Gabriele Hessling; Isabel Deisenhofer
Journal:  Europace       Date:  2015-11-20       Impact factor: 5.214

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.