| Literature DB >> 25878896 |
Anna I Dastiridou1, Xiaojing Pan2, ZhouYuan Zhang1, Kenneth M Marion1, Brian A Francis3, Srinivas R Sadda3, Vikas Chopra3.
Abstract
Purpose. To compare the effects of physiologic versus pharmacologic pupil dilation on anterior chamber angle (ACA) measurements obtained with spectral domain optical coherence tomography (SD-OCT). Methods. Forty eyes from 20 healthy, phakic individuals with open angles underwent anterior segment OCT imaging under 3 pupillary states: (1) pupil constricted under standard room lighting, (2) physiologic mydriasis in a darkened room, and (3) postpharmacologic mydriasis. Inferior angle Schwalbe's line-angle opening distance (SL-AOD) and SL-trabecular-iris-space area (SL-TISA) were computed for each eye and pupillary condition by masked, certified Reading Center graders using customized grading software. Results. SL-AOD and SL-TISA under pupillary constriction to room light were 0.87 ± 0.31 mm and 0.33 ± 0.14 mm(2), respectively; decreased to 0.75 ± 0.29 mm (P < 0.01) and 0.29 ± 0.13 mm(2) (P < 0.01), respectively, under physiologic mydriasis; and increased to 0.90 ± 0.38 mm (P < 0.01) and 0.34 ± 0.17 mm(2) (P = 0.06) under pharmacologic mydriasis compared to baseline. Conclusions. Using SD-OCT imaging, pharmacologic mydriasis yielded the widest angle opening, whereas physiologic mydriasis yielded the most angle narrowing in normal individuals with open iridocorneal angles. Accounting for the state of the pupil and standardizing the lighting condition would appear to be of importance for future studies of the angle.Entities:
Year: 2015 PMID: 25878896 PMCID: PMC4386293 DOI: 10.1155/2015/845643
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Inferior angle scanning under the three pupillary conditions (ACA: anterior chamber angle).
Effect of differing lighting conditions and pharmacologic dilation on pupil size and angle metrics obtained with Cirrus. Results are presented as mean ± standard deviation.
| Light | Dark | Pharmacologic dilation |
| |
|---|---|---|---|---|
| Pupil size (mm) | 4.00 ± 0.81 | 5.27 ± 0.77 | 6.54 ± 0.76 | overall: <0.0011
|
|
| ||||
| SL-AOD (mm) | 0.87 ± 0.31 | 0.75 ± 0.29 | 0.9 ± 0.38 | overall: <0.0011
|
|
| ||||
| SL-TISA (mm2) | 0.33 ± 0.14 | 0.29 ± 0.13 | 0.34 ± 0.17 | overall: 0.0031
|
1Repeated measures, ANOVA.
2Post hoc pairwise comparisons.
SL-AOD: Schwalbe's line-angle opening distance.
SL-TISA: Schwalbe's line-trabecular-iris space area.