Self-calibrated trajectory estimation and signal correction method for robust radial imaging using GRAPPA operator gridding.

Anagha Deshmane1, Martin Blaimer2, Felix Breuer2, Peter Jakob2, Jeffrey Duerk1,3, Nicole Seiberlich1, Mark Griswold1,3.   

Abstract

PURPOSE: In radial imaging, projections may become "miscentered" due to gradient errors such as delays and eddy currents. These errors may result in image artifacts and can disrupt the reliability of direct current (DC) navigation. The proposed parallel imaging-based technique retrospectively estimates trajectory error from miscentered radial data without extra acquisitions, hardware, or sequence modification. THEORY AND METHODS: After phase correction, self-calibrated GRAPPA operator gridding (GROG) weights are iteratively applied to shift-miscentered projections toward the center of k-space. A search algorithm identifies the shift that aligns the peak k-space signals by maximizing the sum-of-squares DC signal estimate of each projection. The algorithm returns a trajectory estimate and a corrected radial k-space signal.
RESULTS: Data from a spherical phantom, the head, and the heart demonstrate that image reconstruction with the estimated trajectory restores image quality and reduces artifacts such as streaks and signal voids. The DC signal level is increased and variability is reduced.
CONCLUSION: Retrospective phase correction and iterative application of GROG can be used to successfully estimate the trajectory error in two-dimensional radial acquisitions for improved image reconstruction without requiring extra data acquisition or sequence modification.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  DC signal; parallel imaging; radial; trajectory estimation

Mesh:

Year:  2015        PMID: 25765372      PMCID: PMC4567526          DOI: 10.1002/mrm.25648

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  25 in total

1.  The NMR phased array.

Authors:  P B Roemer; W A Edelstein; C E Hayes; S P Souza; O M Mueller
Journal:  Magn Reson Med       Date:  1990-11       Impact factor: 4.668

2.  Rapid fat-suppressed isotropic steady-state free precession imaging using true 3D multiple-half-echo projection reconstruction.

Authors:  Aiming Lu; Ethan Brodsky; Thomas M Grist; Walter F Block
Journal:  Magn Reson Med       Date:  2005-03       Impact factor: 4.668

3.  Calibration of gradient propagation delays for accurate two-dimensional radiofrequency pulses.

Authors:  Nigel P Davies; Peter Jezzard
Journal:  Magn Reson Med       Date:  2005-01       Impact factor: 4.668

4.  NMR probes for measuring magnetic fields and field dynamics in MR systems.

Authors:  Nicola De Zanche; Christoph Barmet; Jurek A Nordmeyer-Massner; Klaas P Pruessmann
Journal:  Magn Reson Med       Date:  2008-07       Impact factor: 4.668

5.  Self-calibrating GRAPPA operator gridding for radial and spiral trajectories.

Authors:  Nicole Seiberlich; Felix Breuer; Martin Blaimer; Peter Jakob; Mark Griswold
Journal:  Magn Reson Med       Date:  2008-04       Impact factor: 4.668

6.  Characterizing and correcting gradient errors in non-cartesian imaging: Are gradient errors linear time-invariant (LTI)?

Authors:  Ethan K Brodsky; Alexey A Samsonov; Walter F Block
Journal:  Magn Reson Med       Date:  2009-12       Impact factor: 4.668

7.  Effects of gradient anisotropy in MRI.

Authors:  B Aldefeld; P Börnert
Journal:  Magn Reson Med       Date:  1998-04       Impact factor: 4.668

8.  A robust ultrashort TE (UTE) imaging method with corrected k-space trajectory by using parametric multiple function model of gradient waveform.

Authors:  Masahiro Takizawa; Hikaru Hanada; Kuniharu Oka; Tetsuhiko Takahashi; Etsuji Yamamoto; Masahiko Fujii
Journal:  IEEE Trans Med Imaging       Date:  2012-10-24       Impact factor: 10.048

9.  Correction of gradient-induced phase errors in radial MRI.

Authors:  Amir Moussavi; Markus Untenberger; Martin Uecker; Jens Frahm
Journal:  Magn Reson Med       Date:  2013-02-25       Impact factor: 4.668

10.  Trajectory correction for free-breathing radial cine MRI.

Authors:  Guido Buonincontri; Carmen Methner; Thomas Krieg; T Adrian Carpenter; Stephen J Sawiak
Journal:  Magn Reson Imaging       Date:  2014-04-21       Impact factor: 2.546

View more
  6 in total

1.  Trajectory Auto-Corrected image reconstruction.

Authors:  Julianna D Ianni; William A Grissom
Journal:  Magn Reson Med       Date:  2015-09-12       Impact factor: 4.668

Review 2.  Noncontrast MR angiography: An update.

Authors:  Robert R Edelman; Ioannis Koktzoglou
Journal:  J Magn Reson Imaging       Date:  2018-12-19       Impact factor: 4.813

3.  Autocalibrated wave-CAIPI reconstruction; Joint optimization of k-space trajectory and parallel imaging reconstruction.

Authors:  Stephen F Cauley; Kawin Setsompop; Berkin Bilgic; Himanshu Bhat; Borjan Gagoski; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2016-10-21       Impact factor: 4.668

4.  Simultaneous auto-calibration and gradient delays estimation (SAGE) in non-Cartesian parallel MRI using low-rank constraints.

Authors:  Wenwen Jiang; Peder E Z Larson; Michael Lustig
Journal:  Magn Reson Med       Date:  2018-03-09       Impact factor: 4.668

5.  Volumetric MRI thermometry using a three-dimensional stack-of-stars echo-planar imaging pulse sequence.

Authors:  Sumeeth V Jonathan; William A Grissom
Journal:  Magn Reson Med       Date:  2017-08-07       Impact factor: 4.668

6.  Efficient gradient waveform measurements with variable-prephasing.

Authors:  Kevin D Harkins; Mark D Does
Journal:  J Magn Reson       Date:  2021-02-20       Impact factor: 2.734

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.