Trajectory Auto-Corrected image reconstruction.

Julianna D Ianni1,2, William A Grissom1,2,3,4.   

Abstract

PURPOSE: To estimate k-space trajectory errors in non-Cartesian acquisitions and reconstruct distortion-free images, without trajectory measurements or gradient calibrations. THEORY AND METHODS: The Trajectory Auto-Corrected image Reconstruction method jointly estimates k-space trajectory errors and images, based on SENSE and SPIRiT parallel imaging reconstruction. The underlying idea is that parallel imaging and oversampling in the center of k-space provides data redundancy that can be exploited to simultaneously reconstruct images and correct trajectory errors. Trajectory errors are represented as weighted sums of trajectory-dependent error basis functions, the coefficients of which are estimated using gradient-based optimization.
RESULTS: Trajectory Auto-Corrected image Reconstruction was applied to reconstruct images and errors in golden angle radial, center-out radial, and spiral in vivo 7 Tesla brain acquisitions in five subjects. Compared to reconstructions using nominal trajectories, Trajectory auto-corrected image reconstructions contained considerably less blurring and streaking and were of similar quality to images reconstructed using measured k-space trajectories in the center-out radial and spiral cases. Reconstruction cost function reductions and improvements in normalized image gradient squared were also similar to those for images reconstructed using measured trajectories.
CONCLUSION: Trajectory Auto-Corrected image Reconstruction enables non-Cartesian image reconstructions free from trajectory errors without the need for separate gradient calibrations or trajectory measurements. Magn Reson Med 76:757-768, 2016.
© 2015 Wiley Periodicals, Inc. © 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  center-out radial; eddy currents; golden angle radial; image reconstruction; non-Cartesian imaging; parallel imaging; radial; spiral

Mesh:

Year:  2015        PMID: 26362967      PMCID: PMC4788989          DOI: 10.1002/mrm.25916

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  24 in total

1.  Image metric-based correction (autocorrection) of motion effects: analysis of image metrics.

Authors:  K P McGee; A Manduca; J P Felmlee; S J Riederer; R L Ehman
Journal:  J Magn Reson Imaging       Date:  2000-02       Impact factor: 4.813

2.  Real-time cardiac MRI at 3 tesla.

Authors:  Krishna S Nayak; Charles H Cunningham; Juan M Santos; John M Pauly
Journal:  Magn Reson Med       Date:  2004-04       Impact factor: 4.668

3.  Spatiotemporal magnetic field monitoring for MR.

Authors:  Christoph Barmet; Nicola De Zanche; Klaas P Pruessmann
Journal:  Magn Reson Med       Date:  2008-07       Impact factor: 4.668

4.  Spiral K-space MR imaging of cortical activation.

Authors:  D C Noll; J D Cohen; C H Meyer; W Schneider
Journal:  J Magn Reson Imaging       Date:  1995 Jan-Feb       Impact factor: 4.813

5.  Efficient sample density estimation by combining gridding and an optimized kernel.

Authors:  Nicholas R Zwart; Kenneth O Johnson; James G Pipe
Journal:  Magn Reson Med       Date:  2011-06-17       Impact factor: 4.668

6.  Fast three dimensional sodium imaging.

Authors:  F E Boada; J S Gillen; G X Shen; S Y Chang; K R Thulborn
Journal:  Magn Reson Med       Date:  1997-05       Impact factor: 4.668

7.  A robust ultrashort TE (UTE) imaging method with corrected k-space trajectory by using parametric multiple function model of gradient waveform.

Authors:  Masahiro Takizawa; Hikaru Hanada; Kuniharu Oka; Tetsuhiko Takahashi; Etsuji Yamamoto; Masahiko Fujii
Journal:  IEEE Trans Med Imaging       Date:  2012-10-24       Impact factor: 10.048

8.  Fast l₁-SPIRiT compressed sensing parallel imaging MRI: scalable parallel implementation and clinically feasible runtime.

Authors:  Mark Murphy; Marcus Alley; James Demmel; Kurt Keutzer; Shreyas Vasanawala; Michael Lustig
Journal:  IEEE Trans Med Imaging       Date:  2012-02-15       Impact factor: 10.048

9.  Simple method for MR gradient system characterization and k-space trajectory estimation.

Authors:  Nii Okai Addy; Holden H Wu; Dwight G Nishimura
Journal:  Magn Reson Med       Date:  2011-12-21       Impact factor: 4.668

10.  Self-calibrated trajectory estimation and signal correction method for robust radial imaging using GRAPPA operator gridding.

Authors:  Anagha Deshmane; Martin Blaimer; Felix Breuer; Peter Jakob; Jeffrey Duerk; Nicole Seiberlich; Mark Griswold
Journal:  Magn Reson Med       Date:  2015-03-11       Impact factor: 4.668

View more
  5 in total

1.  Autocalibrated wave-CAIPI reconstruction; Joint optimization of k-space trajectory and parallel imaging reconstruction.

Authors:  Stephen F Cauley; Kawin Setsompop; Berkin Bilgic; Himanshu Bhat; Borjan Gagoski; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2016-10-21       Impact factor: 4.668

Review 2.  Compressed sensing for body MRI.

Authors:  Li Feng; Thomas Benkert; Kai Tobias Block; Daniel K Sodickson; Ricardo Otazo; Hersh Chandarana
Journal:  J Magn Reson Imaging       Date:  2016-12-16       Impact factor: 4.813

3.  Simultaneous auto-calibration and gradient delays estimation (SAGE) in non-Cartesian parallel MRI using low-rank constraints.

Authors:  Wenwen Jiang; Peder E Z Larson; Michael Lustig
Journal:  Magn Reson Med       Date:  2018-03-09       Impact factor: 4.668

4.  Volumetric MRI thermometry using a three-dimensional stack-of-stars echo-planar imaging pulse sequence.

Authors:  Sumeeth V Jonathan; William A Grissom
Journal:  Magn Reson Med       Date:  2017-08-07       Impact factor: 4.668

5.  Ghost reduction in echo-planar imaging by joint reconstruction of images and line-to-line delays and phase errors.

Authors:  Julianna D Ianni; E Brian Welch; William A Grissom
Journal:  Magn Reson Med       Date:  2017-10-16       Impact factor: 4.668

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.