Literature DB >> 25654023

Evaluation of Correlation Between apical Diameter and File Size Using Propex Pixi Apex Locator.

Sreeha Kaluva Kolanu1, Nagesh Bolla2, Sujana Varri3, Jayaprakash Thummu3, Sayesh Vemuri3, Pragna Mandava4.   

Abstract

AIM: AIM of this study is to evaluate the influence of critical diameter of apical foramen and file size using propex pixi apex locator in working length determination.
MATERIALS AND METHODS: In this study, ten single rooted teeth were selected. They were decoronated at cemento enamel junction. After determining the actual working length, they were embedded in alginate mold. Foramina were widened from 0.6mm to 0.8mm. The measurements were taken with electronic apex locator propex pixi with files from sizes 10 K to respective sizes. Statistical accuracy of propex pixi was calculated by using Anova test for different diameters and for the influence of file size.
RESULTS: RESULTS showed that propex pixi apex locator was accurate when foramen diameter is 0.6 (60k file size), its accuracy diminished with increased foramen diameter
CONCLUSION: Propex pixi is accurate for foramen diameter of 0.6mm, independent of file size. Its accuracy decreases as apical foramen widens, so care should be taken when using clinically.

Entities:  

Keywords:  Apex locators; Apical widening; File size; Root length determination

Year:  2014        PMID: 25654023      PMCID: PMC4316329          DOI: 10.7860/JCDR/2014/9307.5247

Source DB:  PubMed          Journal:  J Clin Diagn Res        ISSN: 0973-709X


  11 in total

1.  Evaluation of the accuracy of three electronic apex locators using glass tubules.

Authors:  W Fan; B Fan; J L Gutmann; Z Bian; M W Fan
Journal:  Int Endod J       Date:  2006-02       Impact factor: 5.264

2.  A comparison between two electronic apex locators: an ex vivo investigation.

Authors:  M Venturi; L Breschi
Journal:  Int Endod J       Date:  2007-02-19       Impact factor: 5.264

3.  Influence of apical constriction diameter on Root ZX apex locator precision.

Authors:  Manuela Herrera; Camilo Abalos; Amparo Jimenez Planas; Rafael Llamas
Journal:  J Endod       Date:  2007-08       Impact factor: 4.171

4.  Critical diameter of apical foramen and of file size using the Root ZX apex locator: an in vitro study.

Authors:  Manuela Herrera; Camilo Ábalos; Cristina Lucena; Amparo Jiménez-Planas; Rafael Llamas
Journal:  J Endod       Date:  2011-07-16       Impact factor: 4.171

5.  Clinical reproducibility of three electronic apex locators.

Authors:  V Miletic; K Beljic-Ivanovic; V Ivanovic
Journal:  Int Endod J       Date:  2011-05-21       Impact factor: 5.264

6.  Electronic apex locator: a useful tool for root canal treatment in the primary dentition.

Authors:  A Katz; E Mass; A Y Kaufman
Journal:  ASDC J Dent Child       Date:  1996 Nov-Dec

7.  An experimental study of the principle of electronic root canal measurement.

Authors:  L Huang
Journal:  J Endod       Date:  1987-02       Impact factor: 4.171

8.  Accuracy of the Endex with variations in canal irrigants and foramen size.

Authors:  A F Fouad; E M Rivera; K V Krell
Journal:  J Endod       Date:  1993-02       Impact factor: 4.171

9.  In vitro evaluation of the accuracy of two electronic apex locators.

Authors:  Viresh Chopra; Shibani Grover; S Datta Prasad
Journal:  J Conserv Dent       Date:  2008-04

10.  The effect of file size on the accuracy of the raypex 5 apex locator: an in vitro study.

Authors:  Shiva Sadeghi; Masoomeh Abolghasemi
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2008-05-15
View more
  3 in total

1.  The Effect of File Size and Type and Irrigation Solutions on the Accuracy of Electronic Apex Locators: An In Vitro Study on Canine Teeth.

Authors:  Maciej Janeczek; Piotr Kosior; Dagmara Piesiak-Pańczyszyn; Krzysztof Dudek; Aleksander Chrószcz; Agnieszka Czajczyńska-Waszkiewicz; Małgorzata Kowalczyk-Zając; Aleksandra Gabren-Syller; Karol Kirstein; Aleksandra Skalec; Ewelina Bryła; Maciej Dobrzyński
Journal:  Biomed Res Int       Date:  2016-09-26       Impact factor: 3.411

2.  The Precision of Propex Pixi with Different Instruments and Coronal Preflaring Procedures.

Authors:  Inês Ferreira; Ana Cristina Braga; Irene Pina-Vaz
Journal:  Eur Endod J       Date:  2019-07-10

3.  Accuracy and stability of electronic apex locator length measurements in root canals with wide apical foramen: an ex vivo study.

Authors:  Maayan Shacham; Avi Levin; Avi Shemesh; Alex Lvovsky; Joe Ben Itzhak; Michael Solomonov
Journal:  BDJ Open       Date:  2020-11-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.