Literature DB >> 8509737

Accuracy of the Endex with variations in canal irrigants and foramen size.

A F Fouad1, E M Rivera, K V Krell.   

Abstract

Several electronic apex locators (EALs) are currently available. The manufacturer of a new device, the Endex, claims that it is accurate regardless of canal conditions. This study compared the accuracy of the Endex with that of the Exact-a-pex, the Sono-Explorer Mark III, and the Neosono-D SE as to the effects of fluids in the canal and variation in foramen size. Sixty extracted single-canaled teeth were divided into two groups (narrow and wide foramina), depending on whether the apical foramina permitted the tip of a #30 K file to pass through. An in vitro model was used, in which teeth were fitted in test tubes with the roots immersed in 1% agar in phosphate-buffered saline. Root canal lengths were measured in dry canals, then with ethanol, Xylocaine, and sodium hypochlorite in the canals by each electronic apex locator. These lengths were compared with those of the actual root canal lengths. No significant differences were noted among the instruments in dry canals regardless of the foramen size. The endex was generally superior to the other instruments examined in canals containing conductive fluids, especially where the apical foramen was widened.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8509737     DOI: 10.1016/s0099-2399(06)81196-9

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  14 in total

1.  Root canal length measurement in teeth with electrolyte compensation.

Authors:  K C Nam; S C Kim; S J Lee; Y J Kim; N G Kim; D W Kim
Journal:  Med Biol Eng Comput       Date:  2002-03       Impact factor: 2.602

2.  Evaluation of Correlation Between apical Diameter and File Size Using Propex Pixi Apex Locator.

Authors:  Sreeha Kaluva Kolanu; Nagesh Bolla; Sujana Varri; Jayaprakash Thummu; Sayesh Vemuri; Pragna Mandava
Journal:  J Clin Diagn Res       Date:  2014-12-05

3.  Comparison of Accuracy of Apex Locator with Tactile and Conventional Radiographic Method for Working Length Determination in Primary and Permanent Teeth.

Authors:  Khushboo Rathore; Sandeep Tandon; Meenakshi Sharma; Garima Kalia; Tejshree Shekhawat; Yogendra Chundawat
Journal:  Int J Clin Pediatr Dent       Date:  2020 May-Jun

4.  Comparative evaluation of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study.

Authors:  Saru Jain; Ravi Kapur
Journal:  Contemp Clin Dent       Date:  2012-09

5.  Accuracy of electronic apex locators to detect root canal perforations with inserted metallic posts: an ex vivo study.

Authors:  Benjamín Briseño Marroquín; Claudia Cortazar Fernández; Irene Schmidtmann; Brita Willershausen; Fernando Goldberg
Journal:  Head Face Med       Date:  2014-12-23       Impact factor: 2.151

6.  In vitro comparison of working length determination using three different electronic apex locators.

Authors:  Alper Kuştarci; Dilara Arslan; Demet Altunbaş
Journal:  Dent Res J (Isfahan)       Date:  2014-09

7.  The Effect of File Size and Type and Irrigation Solutions on the Accuracy of Electronic Apex Locators: An In Vitro Study on Canine Teeth.

Authors:  Maciej Janeczek; Piotr Kosior; Dagmara Piesiak-Pańczyszyn; Krzysztof Dudek; Aleksander Chrószcz; Agnieszka Czajczyńska-Waszkiewicz; Małgorzata Kowalczyk-Zając; Aleksandra Gabren-Syller; Karol Kirstein; Aleksandra Skalec; Ewelina Bryła; Maciej Dobrzyński
Journal:  Biomed Res Int       Date:  2016-09-26       Impact factor: 3.411

8.  Use of artificial primary teeth for endodontic laboratory research: experiments related to canal length determination.

Authors:  Anna Carolina V Mello-Moura; Carmela R Bresolin; Cacio Moura-Netto; André Ito; Angela T Araki; José Carlos P Imparato; Fausto M Mendes
Journal:  BMC Oral Health       Date:  2017-11-17       Impact factor: 2.757

9.  The effect of file size on the accuracy of the raypex 5 apex locator: an in vitro study.

Authors:  Shiva Sadeghi; Masoomeh Abolghasemi
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2008-05-15

10.  Evaluation of Two New Electronic Apex-Locator-Controlled Handpieces Using a NiTi Rotary File: An In Vitro Study.

Authors:  Kiumarz Nazarimoghadam; Hossein Labaf
Journal:  J Dent (Tehran)       Date:  2013-11-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.