Literature DB >> 16454793

Evaluation of the accuracy of three electronic apex locators using glass tubules.

W Fan1, B Fan, J L Gutmann, Z Bian, M W Fan.   

Abstract

AIM: To evaluate the accuracy of three different electronic apex locators (EALs) using glass tubules.
METHODOLOGY: Forty-eight glass tubules with different diameters and an agar model were used to mimic root canals. A size 15 stainless steel K-file was used as the measuring electrode. The Root ZX, Propex and Neosono Ultima EZ were used to measure the tubule length with tubules dry, or filled with 0.9% NaCl, 3% H(2)O(2), 2.5% NaOCl or 17% EDTA. The distance between the real length (RL) and measured length (ML) of the tubules was recorded. The range of RL +/- 0.5 mm and RL +/- 1 mm was used to evaluate the accuracy of the EALs. Results were subject to correlation analysis and Friedman's test.
RESULTS: In dry tubules, the accuracy of Root ZX was 75-91.7% for RL +/- 0.5 mm and 100% for RL +/- 1 mm, whilst the measurements of the other two EALs were all within the RL +/- 0.5 mm. No influence from the increase in tubule diameter on the accuracy of all three EALs was observed in dry tubules. In tubules filled with electrolyte, the accuracy of the Root ZX decreased as tubule diameter increased (R(d) > 0, P < 0.05). The RL-ML distance recorded by Propex was inversely related to the tubule diameter (R(d) < 0, P < 0.05). The accuracy of Propex was 75-100% for RL +/- 0.5 mm and 100% for RL +/- 0.5 mm when the tubule diameter was not more than 0.80 mm, but decreased in tubules with diameter over 0.80 mm and filled with 2.5% NaOCl or 17% EDTA. Nearly, all the measurements (except for six tubules) using Neosono Ultima EZ were within 1 mm shorter than RL despite the contents in tubules and the increase of tubule diameter.
CONCLUSIONS: The accuracy of the Root ZX decreased as the tubule diameter increased when tubules were filled with electrolytes. The electrolytes in the tubules decreased the accuracy of Propex when the tubule diameter was large. The electrolytes in tubules and tubule diameter had no influence on the accuracy of Neosono Ultima EZ. The Propex and Neosono Ultima EZ were more accurate than the Root ZX under various conditions in this laboratory study.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16454793     DOI: 10.1111/j.1365-2591.2006.01056.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  8 in total

1.  Evaluation of Correlation Between apical Diameter and File Size Using Propex Pixi Apex Locator.

Authors:  Sreeha Kaluva Kolanu; Nagesh Bolla; Sujana Varri; Jayaprakash Thummu; Sayesh Vemuri; Pragna Mandava
Journal:  J Clin Diagn Res       Date:  2014-12-05

2.  Precision of multi-frequency electronic apex locators.

Authors:  Roy George
Journal:  Evid Based Dent       Date:  2016-09

3.  Effect of Sodium Hypochlorite Concentration on Electronic Apex Locator Reliability.

Authors:  Franck Diemer; Emma Plews; Marie Georgelin-Gurgel; Lora Mishra; Hyeon-Cheol Kim
Journal:  Materials (Basel)       Date:  2022-01-23       Impact factor: 3.623

4.  Efficiency of an Integrated Apex Locator in Determining Working Length in Various Irrigating Solutions: An In Vivo Study.

Authors:  Rakesh Reddy Chukka; Malati Devi Bellam; Narender Reddy Marukala; Sainath Dinapadu; Naresh Kumar Konda; Jithender Nagilla
Journal:  J Pharm Bioallied Sci       Date:  2020-08-28

5.  A comparative evaluation of working length with digital radiography and third generation apex locator (ProPex) in the presence of various intracanal irrigants: An in vivo/ex vivo study.

Authors:  Irfana Khursheed; Ruchika Bansal; Tajinder Bansal; Harkanwal Preet Singh; Madhulika Yadav; K Jetender Reddy
Journal:  Dent Res J (Isfahan)       Date:  2014-01

6.  A comparative evaluation of accuracy of three electronic apex locators using histological section as gold standard: An ex vivo study.

Authors:  Divya Saxena; Suparna Ganguly Saha; Anuj Bharadwaj; Neelam Vijaywargiya; Sandeep Dubey; Shubham Kala
Journal:  J Conserv Dent       Date:  2017 Jul-Aug

7.  Ex Vivo Evaluation of the Accuracy of 3 Electronic Apex Locators in Different Environments: A Micro-Computed Tomography Study.

Authors:  Firdevs Çınar; Yakup Üstün
Journal:  Eur Endod J       Date:  2020-12

8.  An in vitro stereomicroscopic comparative evaluation of a combination of apex locator and endodontic motor with an integrated endodontic motor.

Authors:  Ch Swarupa; Girija S Sajjan; Yv Sashi Kanth
Journal:  J Conserv Dent       Date:  2013-09
  8 in total