Clinical reproducibility of three electronic apex locators.

V Miletic1, K Beljic-Ivanovic, V Ivanovic.   

Abstract

AIM: To compare the reproducibility of three electronic apex locators (EALs), Dentaport ZX, RomiApex A-15 and Raypex 5, under clinical conditions.
METHODOLOGY: Forty-eight root canals of incisors, canines and premolars with or without radiographically confirmed periapical lesions required root canal treatment in 42 patients. In each root canal, all three EALs were used to determine the working length (WL) that was defined as the zero reading and indicated by 'Apex', '0.0' or 'red square' markings on the EAL display. A new K-file of the same size was used for each measurement. The file length was fixed with a rubber stop and measured to an accuracy of 0.01 mm. Measurements were undertaken by two calibrated operators. Differences in zero readings between the three EALs in the same root canal were statistically analysed using paired t-tests with the Bonferroni correction, Bland-Altman plot and Linn's concordance correlation coefficients at α = 0.05.
RESULTS: Mean and standard deviation values measured by the three EALs showed no statistically significant differences. Identical readings by all three EALs were found in 10.4% of root canals. Forty-three per cent of readings differed by less than ± 0.5 mm and 31.3% exceeded a difference of ± 1 mm.
CONCLUSIONS: The clinical reproducibility of Dentaport ZX, RomiApex A-15 and Raypex 5 was confirmed with the majority of readings within the ± 1.0 mm range. However, the small number of identical zero readings suggests that EALs are not reliable as the sole means of WL determination under clinical conditions.
© 2011 International Endodontic Journal.

Entities:  

Mesh:

Year:  2011        PMID: 21599708     DOI: 10.1111/j.1365-2591.2011.01897.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  5 in total

1.  The use of cone beam computed tomography to predetermine root canal lengths in molar teeth: a comparison between two-dimensional and three-dimensional measurements.

Authors:  J P Tchorz; P J Poxleitner; S Stampf; S B M Patzelt; D Rottke; E Hellwig; M J Altenburger
Journal:  Clin Oral Investig       Date:  2013-07-30       Impact factor: 3.573

2.  Evaluation of Correlation Between apical Diameter and File Size Using Propex Pixi Apex Locator.

Authors:  Sreeha Kaluva Kolanu; Nagesh Bolla; Sujana Varri; Jayaprakash Thummu; Sayesh Vemuri; Pragna Mandava
Journal:  J Clin Diagn Res       Date:  2014-12-05

3.  The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.

Authors:  Carlos G Adorno; Sandra M Solaeche; Ileana E Ferreira; Andrea Pedrozo; Patricia M Escobar; Vicente R Fretes
Journal:  Clin Oral Investig       Date:  2021-02-13       Impact factor: 3.573

4.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

5.  Accuracy of Software-Based Three-Dimensional Root Canal Length Measurements Using Cone-Beam Computed Tomography.

Authors:  J P Tchorz; K T Wrbas; C Von See; K Vach; Sbm Patzelt
Journal:  Eur Endod J       Date:  2018-11-13
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.