A comparison between two electronic apex locators: an ex vivo investigation.

M Venturi1, L Breschi.   

Abstract

AIM: To compare ex vivo the performance of the Apex Finder and the Root ZX apex locators, with and without irrigant, in canals having different diameters.
METHODOLOGY: Sixty canals in 60 teeth were prepared using stainless steel hand files and 0.04 taper NiTi rotary instruments. During preparation the narrowest diameter of the canal was transported to the apical root surface. The canals were irrigated with RC-Prep and 5% NaOCl solution. Six groups were obtained, each with 10 canals having the same diameter of foramen, either 0.15, 0.20, 0.25, 0.40, 0.60 and 0.80 mm. A size 15 K-file was advanced into each canal until its tip was observed under x10 magnification to reach the foramen and the corresponding length recorded. The measurements were performed to an accuracy of 0.25 mm as a base unit of length. The teeth were then fixed to a plastic bar suspended over a glass container filled with 0.9% NaCl solution. Each apex locator was tested when the K-file was at the foramen, or 0.5, 1.0, 1.5 and 2.0 mm short; with the root apex immersed into the solution; with the canal dry or irrigated with NaCl. To evaluate the accuracy of both electronic apex locators (EALs) each electronically determined distance was compared with the actual length and the data analysed using the General Linear Model and the Student t-test.
RESULTS: Out of 2400 measurements 100 were electrically unstable, all with the Root ZX. In total, 521 measurements located the position of the file tip beyond the apex, in general, in high conductive conditions with the Root ZX and in low conductive conditions with the Apex Finder. No significant difference in terms of accuracy was found between the two EALs when the file tip was at the foramen (Root ZX mean +0.12 mm, SD 1.22 mm; Apex Finder mean +0.57 mm, SD 1.16 mm). Comparing all the measurements performed with the file tip within 2 mm of the foramen, in all the different conditions tested, the accuracy was affected (P<0.025) by diameter of the foramen, type of EAL, distance to the apex, and by several interactions.
CONCLUSIONS: Under the different ex vivo conditions both EALs provided accurate measurements when the file tip was at the foramen. The accuracy of the Apex Finder was negatively influenced by high conductive conditions, whilst the Root ZX provided inaccurate and unstable measurements mostly in low conductive conditions.

Entities:  

Mesh:

Year:  2007        PMID: 17309741     DOI: 10.1111/j.1365-2591.2006.01229.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  8 in total

1.  Comparison of two electronic apex locators on human cadavers.

Authors:  Luca Marigo; Gian Luca Gervasi; Francesco Somma; Giuseppe Squeo; Raffaella Castagnola
Journal:  Clin Oral Investig       Date:  2015-11-04       Impact factor: 3.573

2.  Evaluation of Correlation Between apical Diameter and File Size Using Propex Pixi Apex Locator.

Authors:  Sreeha Kaluva Kolanu; Nagesh Bolla; Sujana Varri; Jayaprakash Thummu; Sayesh Vemuri; Pragna Mandava
Journal:  J Clin Diagn Res       Date:  2014-12-05

3.  Effect of Sodium Hypochlorite Concentration on Electronic Apex Locator Reliability.

Authors:  Franck Diemer; Emma Plews; Marie Georgelin-Gurgel; Lora Mishra; Hyeon-Cheol Kim
Journal:  Materials (Basel)       Date:  2022-01-23       Impact factor: 3.623

4.  Efficiency of 2 electronic apex locators on working length determination: A clinical study.

Authors:  Sibel Koçak; Mustafa Murat Koçak; Baran Can Sağlam
Journal:  J Conserv Dent       Date:  2013-05

5.  An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files.

Authors:  Paras Mull Gehlot; Vinutha Manjunath; Mysore Krishnaswamy Manjunath
Journal:  Restor Dent Endod       Date:  2016-01-04

6.  A comparative evaluation of accuracy of three electronic apex locators using histological section as gold standard: An ex vivo study.

Authors:  Divya Saxena; Suparna Ganguly Saha; Anuj Bharadwaj; Neelam Vijaywargiya; Sandeep Dubey; Shubham Kala
Journal:  J Conserv Dent       Date:  2017 Jul-Aug

7.  Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants.

Authors:  Sakshi Bilaiya; Pallav Mahesh Patni; Pradeep Jain; Sanket Hans Pandey; Swadhin Raghuwanshi; Bhupesh Bagulkar
Journal:  Eur Endod J       Date:  2020-03-19

8.  In vitro detection of simulated apical root perforation with two electronic apex locators.

Authors:  Kiumars Nazari Moghaddam; Shahrzad Nazari; Leila Shakeri; Kiamars Honardar; Farshid Mirmotalebi
Journal:  Iran Endod J       Date:  2010-02-20
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.