Maciej Janeczek1, Piotr Kosior2, Dagmara Piesiak-Pańczyszyn2, Krzysztof Dudek3, Aleksander Chrószcz1, Agnieszka Czajczyńska-Waszkiewicz2, Małgorzata Kowalczyk-Zając2, Aleksandra Gabren-Syller4, Karol Kirstein1, Aleksandra Skalec1, Ewelina Bryła5, Maciej Dobrzyński6.
Abstract
Measurements of the root canal during endodontic treatment have a significant influence on the course of the therapeutic process as well as on its final result in both human and veterinary medicine. The apical constriction should be the termination point for the preparation and filling of the root canal. This research was conducted with the use of a Septodont kit consisting of a small chamber filled with the examined solution in which a healthy second incisor was placed. The step back method was applied for the root canal preparation and master apical file of 30 was used. The working length was 22 mm. The examination was conducted with the use of steel as well as nickel titanium hand instruments. Different irrigation solutions and two types of apex locators were used. Measurements of the working length of the root canal showed dependence on the size of the instrument. Examinations carried out in various environments showed that analogical measurements were obtained only for sodium hypochlorite solutions. In other environments the measured sections were shortened. Comparative examinations with the use of steel instruments demonstrated insignificant measurement differences. Compared to these results, the measurements in nickel titanium group were characterized by more considerable deviations.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27747242 PMCID: PMC5056237 DOI: 10.1155/2016/8594087
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Working length of the instrument (master apical file (MAF)) was 22 mm and it was established with the use of VDW endodontic ruler.
Figure 2Septodont measuring kit consisting of a chamber filled with a selected solution, a passive electrode connected with the chamber, and an active electrode joined with the endodontic instrument placed in a root canal of the tooth and connected with the Endopilot apex locator (Schlumbohm, Brokstedt, Germany).
Figure 3Septodont measuring kit consisting of a chamber filled with a selected solution, a passive electrode connected with the chamber, and an active electrode joined with the endodontic instrument placed in a root canal of the tooth and connected with the iPex apex locator (NSK, Tochigi, Japan).
Basic statistics of the measurement results of the distance between the instrument and the apical constriction performed in the 5.25% NaOCl environment with the use of steel S-files (Poldent) with the size ranging from 08 to 30 in accordance with ISO and with the use of two apex locators: Endopilot and iPex NSK as well as the results of comparison made with the use of Student's t-test for dependent variables.
| Instrument size | Endopilot | iPex NSK |
|
|
|---|---|---|---|---|
|
| 3.23 ± 0.03 | 2.58 ± 0.03 |
|
|
|
| 2.93 ± 0.12 | 2.22 ± 0.10 |
|
|
|
| 3.02 ± 0.12 | 2.07 ± 0.15 |
|
|
|
| 2.58 ± 0.10 | 2.00 ± 0.05 |
|
|
|
| 2.48 ± 0.06 | 1.98 ± 0.03 |
|
|
|
| 2.08 ± 0.06 | 1.88 ± 0.13 | 2.370 | 0.077 |
|
| 1.82 ± 0.08 | 1.70 ± 0.10 | 1.606 | 0.184 |
|
| 1.73 ± 0.15 | 1.67 ± 0.03 | 0.499 | 0.499 |
|
| 1.52 ± 0.06 | 1.38 ± 0.03 |
|
|
|
| 1.37 ± 0.03 | 1.30 ± 0.05 | 2.000 | 0.116 |
|
| 1.25 ± 0.00 | 1.23 ± 0.03 | 1.000 | 0.374 |
|
| 1.12 ± 0.08 | 1.22 ± 0.03 | −2.121 | 0.101 |
|
| 1.48 ± 0.08 | 1.38 ± 0.03 | 2.121 | 0.101 |
|
| 1.38 ± 0.08 | 1.32 ± 0.03 | 1.414 | 0.230 |
|
| 1.15 ± 0.05 | 1.03 ± 0.03 |
|
|
|
| 0.93 ± 0.03 | 0.83 ± 0.03 |
|
|
|
| 0.78 ± 0.03 | 0.62 ± 0.03 |
|
|
|
| 0.53 ± 0.03 | 0.52 ± 0.03 | 0.707 | 0.519 |
Basic statistics of the distance between the endodontic device and the apical constriction measured with the use of a steel file with the size of 30/22 mm for different irrigation solutions.
| Irrigation solution | Endopilot | iPex NSK |
|
|
|---|---|---|---|---|
| 2.0% NaOCl | 0.51 ± 0.02 | 0.49 ± 0.02 | 1.604 | 0.184 |
| 5.25% NaOCl | 0.50 ± 0.05 | 0.52 ± 0.01 | −0.611 | 0.574 |
| 40.0% citric acid | 2.10 ± 0.04 | 1.51 ± 0.08 |
|
|
| 2.0% Gluco-chex | 2.11 ± 0.05 | 1.99 ± 0.06 | 2.668 | 0.056 |
| Isopropyl alcohol | 0 | 0 | — | — |
| 3.0% H2O2 | 2.13 ± 0.03 | 2.02 ± 0.08 | 2.220 | 0.091 |
| 15.0% Endosal | 1.37 ± 0.10 | 1.49 ± 0.06 | −1.886 | 0.132 |
| 0.9% NaCl | 2.07 ± 0.03 | 1.23 ± 0.05 |
|
|
Statistically significant differences between the measurement results obtained with the use of the Endopilot and iPex NSK apex locators were observed in the 40.0% citric acid and 0.9% NaCl solutions (p < 0.001).
Measurements of the distance from the apical constriction performed with the use of steel and NiTi instruments with the size of 30/22 mm and with the use of Endopilot apex locator in the 5.25% NaOCl environment.
| Material | ANOVA | |||||
|---|---|---|---|---|---|---|
| (1) H-file Mani steel | (2) H-file Polintech NiTi Flexible | (3) K-file Medin steel | (4) K-reamer Mani steel | (5) S-file Poldent steel | ||
| Apex locator reading | 44.7 ± 0.5 | 40.0 ± 0.8 | 45 ± 1 | 46 ± 1 | 48 ± 1 |
|
| Distance | 0.58 ± 0.06 | 1.13 ± 0.12 | 0.50 ± 0.07 | 0.46 ± 0.07 | 0.14 ± 0.07 |
|
Results of multiple comparisons with the use of a post hoc test.
| Material | (1) | (2) | (3) | (4) | (5) | |
|---|---|---|---|---|---|---|
| 1 | (1) |
| 0.773 | 0.442 |
| |
| 2 | (2) |
|
|
|
| |
| 3 | (3) | 0.773 |
| 0.971 |
| |
| 4 | (4) | 0.442 |
| 0.971 |
| |
| 5 | (5) |
|
|
|
| |
Tukey HSD test. Variable: d (mm).
Bold differences are significant at p < 0.05.
Figure 4Correlation diagram of the distance between the instrument and the apical constriction (mm) with the indication of the Endopilot apex locator and the regression equation.
Figure 5Comparison of measurement results of the distance from the apical constriction d (mm) obtained with the use of the Endopilot apex locator and the results of the analysis of variance, where 1 stands for H-file Mani steel, 2 stands for H-file Polintech NiTi Flexible, 3 represents K-file Medin steel, 4 represents K-reamer Mani steel, and 5 represents S-file Poldent steel.