| Literature DB >> 25207409 |
Huiliang Liu1, Zhigeng Jin, Yunpeng Deng, Limin Jing.
Abstract
The purpose of this study was to evaluate peak skin dose received by the patient and scattered dose to the operator during dual-axis rotational coronary angiography (DARCA), and to compare with those of standard coronary angiography (SA). An anthropomorphic phantom was used to simulate a patient undergoing diagnostic coronary angiography. Cine imaging was applied on the phantom for 2 s, 3 s, and 5 s in SA projections to mimic clinical situations with normal vessels, and uncomplicated and complicated coronary lesions. DARCA was performed in two curved trajectories around the phantom. During both SA and DARCA, peak skin dose was measured with thermoluminescent dosimeter arrays and scattered dose with a dosimeter at predefined height (approximately at the level of left eye) at the operator's location. Compared to SA, DARCA was found lower in both peak skin dose (range: 44%-82%, p < 0.001) and scattered dose (range: 40%-70%, p < 0.001). The maximal reductions were observed in the set mimicking complicated lesion examinations (82% reduction for peak skin dose, p < 0.001; 70% reduction for scattered dose, p < 0.001). DARCA reduces both peak skin dose and scattered dose in comparison to SA. The benefi t of radiation dose reduction could be especially signifi cant in complicated lesion examinations due to large reduction in X-ray exposure time. The use of DARCA could, therefore, be recommended in clinical practice to minimize radiation dose.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25207409 PMCID: PMC5875506 DOI: 10.1120/jacmp.v15i4.4805
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Coronary angiography performed on an anthropomorphic phantom. For scattered dose measurements, the dosimeter was clamped to a drip stand at the height of 170 cm (white arrow), which was positioned at a distance of 91 cm from the exposure center of phantom. No radiation protections were used during the measurements.
X‐ray exposure factors for different coronary angiographic protocol
|
|
|
|
|
|
|---|---|---|---|---|
| Fluoroscopy time per run (s) | 3 | 3 | 3 | 5 |
| Cine time per run (s) | 2 | 3 | 5 | 6.8 (4.5) |
| Total fluoroscopy time (s) | 21 | 21 | 21 | 15 |
| Total cine time (s) | 14 | 21 | 35 | 11.3 |
| Field of view (cm) | 25 | 25 | 25 | 25 |
| Source‐image detector distance (cm) | 100 | 100 | 100 | 120 |
| Source‐skin distance (cm) | 60 | 60 | 60 | 60 |
| Fluoroscopy mode | Low | Low | Low | Low |
| Cine mode (fr/s) | 15 | 15 | 15 | 15 |
Number in the bracket indicates cine time for the right coronary spin; out of the bracket for the left coronary spin.
SA = standard coronary angiography; DARCA = dual‐axis rotational coronary angiography.
Figure 2The thermoluminescent dosimeters (TLDs) positions used in the skin dose measurement. The TLD array was placed under the upper back of the anthropomorphic phantom, beneath the heart. TLDs were spaced by 9 cm from each other and numbered by 1 to 9. TLD = thermoluminescent dosimeter.
Radiation dose to the phantom patient during coronary angiography.
|
|
|
|
| |||
|---|---|---|---|---|---|---|
|
|
|
|
| |||
| 2 s SA |
|
|
|
|
|
|
| 3 s SA |
|
|
|
|
|
|
| 5 s SA |
|
|
|
|
|
|
| DARCA |
|
|
|
|
|
|
Values are presented as mean .
; ; ; ; .
Figure 3Skin dose distribution in the thermoluminescent dosimeter arrays separated by different angiographic protocol. 2 s, 3 s, and 5 s cine time was applied in each projection during the 2 s, 3 s, and 5 s SA, which represents evaluating normal vessels, uncomplicated, and complicated coronary lesions in clinical practice, respectively. TLD = thermoluminescent dosimeter; SA = standard coronary angiography; DARCA = dual‐axis rotational coronary angiography.
Scattered dose to the operator during coronary angiography.
|
|
|
|
|
|---|---|---|---|
| 2 s SA |
|
|
|
| 3 s SA |
|
|
|
| 5 s SA |
|
|
|
| DARCA |
|
|
|
Values are presented as mean .
; .
Entrance dose rate during fluoroscopy and dose per frame during cine for different source‐image detector distances.
|
|
|
|
|---|---|---|
| 100 |
|
|
| 110 |
|
|
| 120 |
|
|
Entrance dose rate and dose per frame were measured at the patient entrance reference point in the anterior‐posterior projection (in the low fluoroscopy mode with 25 cm field of view and for both fluoroscopy and cine)