Sinogram Affirmed Iterative Reconstruction (SAFIRE) versus weighted filtered back projection (WFBP) effects on quantitative measure in the COPDGene 2 test object.

J P Sieren1, E A Hoffman2, M K Fuld3, K S Chan4, J Guo1, J D Newell5.   

Abstract

PURPOSE: Assessing pulmonary emphysema using Quantitative CT of the lung depends on accurate measures of CT density. Sinogram-Affirmed-Iterative-Reconstruction (SAFIRE) is a new approach for reconstructing CT data acquired at significantly lower doses. However, quantitative effects of this method remain unexplored. The authors investigated the effects on the median values of materials in the COPDGene2 test-object as a function of the reconstruction method [weighted filtered back projection (WFBP) versus SAFIRE], test-object size, dose, and material composition using a Siemens SOMATOM Definition FLASH CT scanner.
METHODS: The COPDGene2 test-object contains eight materials; acrylic, water, four foams (20 lb, 12 lb, lung-equivalent, and 4 lb emphysema-equivalent), internal and external-air. The test-object was scanned with three different outer ring sizes, simulating three different body habitus. There is an average size (36 cm) Ring A, large size (40 cm) Ring B, and small size Ring C (30 cm). The CT protocol used 120 kVp, 0.5 s rotation, 1.0 pitch, and a 0.6 slice collimation with progressively decreasing x-ray exposure values, 11.94-0.74 mGy. With a thorax length of 30 cm, the corresponding effective doses would be 5.01-0.31 mSv. The effects of using SAFIRE versus WFBP were assessed using a two tailed t-test for each ring size, material, and dose. Multivariable linear regression was used to evaluate the relative effects of ring size, material composition, dose, and reconstruction method on the measured median value in HU.
RESULTS: SAFIRE versus WFBP, at the largest ring size and two lowest doses there was a significant difference in median values of 4 lb-foam, p<0.01. Using the smallest ring size at the lowest dose level there was a significant difference in the median value of 4 lb-foam, but the effect size was small, 1 HU. There is a significant difference in median values of both internal and external air using both the small and medium size rings at the three lowest dose levels, p<0.05. There are significant differences noted at both high and low dose levels when using the large ring size in the median values of internal and external air when, p<0.05. These effects on 4 lb-foam, inside and outside air are shown to be in part due to truncation effects on the median value since the lowest HU value in the CT scale used is -1024 HU. Multivariable linear regression results demonstrated significant effects on the measured material median value and standard deviation due to ring size, material composition, dose level, and reconstruction method, p<0.05.
CONCLUSIONS: The authors have shown that there is no significant effect on the median values obtained when using WFBP versus SAFIRE in materials with CT density between 120 and -856 HU using three different test-object sizes and CT doses that vary from 11.94 to 0.74 mGy. The authors have demonstrated there are significant effects on median values obtained when using WFBP versus SAFIRE in materials with CT density values between -937 and -1000 HU depending on the ring size and dose used. As expected, there is considerable reduction in image noise (lower standard deviation) using SAFIRE versus WFBP with all ring sizes, doses, and materials in the COPDGene2 test-object.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25186397      PMCID: PMC4149690          DOI: 10.1118/1.4893498

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  44 in total

1.  Chest computed tomography using iterative reconstruction vs filtered back projection (Part 1): Evaluation of image noise reduction in 32 patients.

Authors:  François Pontana; Julien Pagniez; Thomas Flohr; Jean-Baptiste Faivre; Alain Duhamel; Jacques Remy; Martine Remy-Jardin
Journal:  Eur Radiol       Date:  2010-11-05       Impact factor: 5.315

2.  Diffuse lung disease: CT of the chest with adaptive statistical iterative reconstruction technique.

Authors:  Priyanka Prakash; Mannudeep K Kalra; Jeanne B Ackman; Subba R Digumarthy; Jiang Hsieh; Synho Do; Jo-Anne O Shepard; Matthew D Gilman
Journal:  Radiology       Date:  2010-07       Impact factor: 11.105

3.  Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients.

Authors:  François Pontana; Alain Duhamel; Julien Pagniez; Thomas Flohr; Jean-Baptiste Faivre; Anne-Lise Hachulla; Jacques Remy; Martine Remy-Jardin
Journal:  Eur Radiol       Date:  2010-11-16       Impact factor: 5.315

Review 4.  CT technology overview: 64-slice and beyond.

Authors:  Patrik Rogalla; Christian Kloeters; Patrick A Hein
Journal:  Radiol Clin North Am       Date:  2009-01       Impact factor: 2.303

5.  The generalisation of student's problems when several different population variances are involved.

Authors:  B L WELCH
Journal:  Biometrika       Date:  1947       Impact factor: 2.445

Review 6.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm.

Authors:  Alvin C Silva; Holly J Lawder; Amy Hara; Jennifer Kujak; William Pavlicek
Journal:  AJR Am J Roentgenol       Date:  2010-01       Impact factor: 3.959

7.  Genetic epidemiology of COPD (COPDGene) study design.

Authors:  Elizabeth A Regan; John E Hokanson; James R Murphy; Barry Make; David A Lynch; Terri H Beaty; Douglas Curran-Everett; Edwin K Silverman; James D Crapo
Journal:  COPD       Date:  2010-02       Impact factor: 2.409

8.  Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience.

Authors:  Priyanka Prakash; Mannudeep K Kalra; Subba R Digumarthy; Jiang Hsieh; Homer Pien; Sarabjeet Singh; Matthew D Gilman; Jo-Anne O Shepard
Journal:  J Comput Assist Tomogr       Date:  2010-01       Impact factor: 1.826

Review 9.  Pushing the envelope: new computed tomography techniques for cardiothoracic imaging.

Authors:  Thomas G Flohr; Ernst Klotz; Thomas Allmendinger; Rainer Raupach; Herbert Bruder; Bernhard Schmidt
Journal:  J Thorac Imaging       Date:  2010-05       Impact factor: 3.000

10.  A multivariate analysis of risk factors for the air-trapping asthmatic phenotype as measured by quantitative CT analysis.

Authors:  Ashley Busacker; John D Newell; Thomas Keefe; Eric A Hoffman; Janice Cook Granroth; Mario Castro; Sean Fain; Sally Wenzel
Journal:  Chest       Date:  2008-08-08       Impact factor: 9.410

View more
  16 in total

1.  Quantitative computed tomographic imaging-based clustering differentiates asthmatic subgroups with distinctive clinical phenotypes.

Authors:  Sanghun Choi; Eric A Hoffman; Sally E Wenzel; Mario Castro; Sean Fain; Nizar Jarjour; Mark L Schiebler; Kun Chen; Ching-Long Lin
Journal:  J Allergy Clin Immunol       Date:  2017-01-29       Impact factor: 10.793

2.  CT-derived Biomechanical Metrics Improve Agreement Between Spirometry and Emphysema.

Authors:  Surya P Bhatt; Sandeep Bodduluri; John D Newell; Eric A Hoffman; Jessica C Sieren; Meilan K Han; Mark T Dransfield; Joseph M Reinhardt
Journal:  Acad Radiol       Date:  2016-04-04       Impact factor: 3.173

3.  SPIROMICS Protocol for Multicenter Quantitative Computed Tomography to Phenotype the Lungs.

Authors:  Jered P Sieren; John D Newell; R Graham Barr; Eugene R Bleecker; Nathan Burnette; Elizabeth E Carretta; David Couper; Jonathan Goldin; Junfeng Guo; MeiLan K Han; Nadia N Hansel; Richard E Kanner; Ella A Kazerooni; Fernando J Martinez; Stephen Rennard; Prescott G Woodruff; Eric A Hoffman
Journal:  Am J Respir Crit Care Med       Date:  2016-10-01       Impact factor: 21.405

4.  Lumen area change (Delta Lumen) between inspiratory and expiratory multidetector computed tomography as a measure of severe outcomes in asthmatic patients.

Authors:  Sung Shine Shim; Mark L Schiebler; Michael D Evans; Nizar Jarjour; Ron L Sorkness; Loren C Denlinger; Alfonso Rodriguez; Sally Wenzel; Eric A Hoffman; Ching-Long Lin; David S Gierada; Mario Castro; Sean B Fain
Journal:  J Allergy Clin Immunol       Date:  2018-02-10       Impact factor: 10.793

5.  Impact of advanced detector technology and iterative reconstruction on low-dose quantitative assessment of lung computed tomography density in a biological lung model.

Authors:  E Hammond; K S Chan; J C Ames; N Stoyles; C M Sloan; J Guo; J D Newell; E A Hoffman; J C Sieren
Journal:  Med Phys       Date:  2018-06-21       Impact factor: 4.071

6.  Comparison of low- and ultralow-dose computed tomography protocols for quantitative lung and airway assessment.

Authors:  Emily Hammond; Chelsea Sloan; John D Newell; Jered P Sieren; Melissa Saylor; Craig Vidal; Shayna Hogue; Frank De Stefano; Alexa Sieren; Eric A Hoffman; Jessica C Sieren
Journal:  Med Phys       Date:  2017-08-02       Impact factor: 4.071

7.  The effects of iterative reconstruction and kernel selection on quantitative computed tomography measures of lung density.

Authors:  Alfonso Rodriguez; Frank N Ranallo; Philip F Judy; Sean B Fain
Journal:  Med Phys       Date:  2017-05-12       Impact factor: 4.071

8.  Low-dose lung cancer screening with photon-counting CT: a feasibility study.

Authors:  Rolf Symons; Tyler E Cork; Pooyan Sahbaee; Matthew K Fuld; Steffen Kappler; Les R Folio; David A Bluemke; Amir Pourmorteza
Journal:  Phys Med Biol       Date:  2016-12-17       Impact factor: 3.609

Review 9.  Pulmonary CT and MRI phenotypes that help explain chronic pulmonary obstruction disease pathophysiology and outcomes.

Authors:  Eric A Hoffman; David A Lynch; R Graham Barr; Edwin J R van Beek; Grace Parraga
Journal:  J Magn Reson Imaging       Date:  2015-07-22       Impact factor: 4.813

10.  Very low-dose (0.15 mGy) chest CT protocols using the COPDGene 2 test object and a third-generation dual-source CT scanner with corresponding third-generation iterative reconstruction software.

Authors:  John D Newell; Matthew K Fuld; Thomas Allmendinger; Jered P Sieren; Kung-Sik Chan; Junfeng Guo; Eric A Hoffman
Journal:  Invest Radiol       Date:  2015-01       Impact factor: 6.016

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.