Acceleration and motion-correction techniques for high-resolution intravascular MRI.

Shashank Sathyanarayana Hegde1, Yi Zhang1,2, Paul A Bottomley1,2.   

Abstract

PURPOSE: High-resolution intravascular (IV) MRI is susceptible to degradation from physiological motion and requires high frame-rates for true endoscopy. Traditional cardiac-gating techniques compromise efficiency by reducing the effective scan rate. Here we test whether compressed sensing (CS) reconstruction and ungated motion-compensation using projection shifting, could provide faster motion-suppressed, IVMRI. THEORY AND METHODS: CS reconstruction is developed for undersampled Cartesian and radial imaging using a new IVMRI-specific cost function to effectively increase imaging speed. A new motion correction method is presented wherein individual IVMRI projections are shifted based on the IVMRI detector's intrinsic amplitude and phase properties. The methods are tested at 3 Tesla (T) in fruit, human vessel specimens, and a rabbit aorta in vivo. Images are compared using structural-similarity and "spokal variation" indices.
RESULTS: Although some residual artifacts persisted, CS acceleration and radial motion compensation strategies reduced motion artifact in vitro and in vivo, allowing effective accelerations of up to eight-fold at 200-300 µm resolution.
CONCLUSION: The 3T IVMRI detectors are well-suited to CS and motion correction strategies based on their intrinsic radially-sparse sensitivity profiles and high signal-to-noise ratios.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  compressed sensing; intravascular MRI; motion correction; projection reconstruction

Mesh:

Year:  2014        PMID: 25163750      PMCID: PMC4344423          DOI: 10.1002/mrm.25436

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  30 in total

1.  Undersampled projection reconstruction for active catheter imaging with adaptable temporal resolution and catheter-only views.

Authors:  Dana C Peters; Robert J Lederman; Alexander J Dick; Venkatesh K Raman; Michael A Guttman; J Andrew Derbyshire; Elliot R McVeigh
Journal:  Magn Reson Med       Date:  2003-02       Impact factor: 4.668

2.  Projection reconstruction techniques for reduction of motion effects in MRI.

Authors:  G H Glover; J M Pauly
Journal:  Magn Reson Med       Date:  1992-12       Impact factor: 4.668

3.  Image quality assessment: from error visibility to structural similarity.

Authors:  Zhou Wang; Alan Conrad Bovik; Hamid Rahim Sheikh; Eero P Simoncelli
Journal:  IEEE Trans Image Process       Date:  2004-04       Impact factor: 10.856

4.  Real-time MRI at a resolution of 20 ms.

Authors:  Martin Uecker; Shuo Zhang; Dirk Voit; Alexander Karaus; Klaus-Dietmar Merboldt; Jens Frahm
Journal:  NMR Biomed       Date:  2010-10       Impact factor: 4.044

5.  An expandable catheter loop coil for intravascular MRI in larger blood vessels.

Authors:  Ann-Kathrin Homagk; Reiner Umathum; Matthias Korn; Marc-André Weber; Peter Hallscheidt; Wolfhard Semmler; Michael Bock
Journal:  Magn Reson Med       Date:  2010-02       Impact factor: 4.668

6.  Intravascular magnetic resonance imaging using a loopless catheter antenna.

Authors:  O Ocali; E Atalar
Journal:  Magn Reson Med       Date:  1997-01       Impact factor: 4.668

7.  Accelerated, motion-corrected high-resolution intravascular MRI at 3T.

Authors:  Shashank Sathyanarayana Hegde; Yi Zhang; Paul A Bottomley
Journal:  Proc Int Soc Magn Reson Med Sci Meet Exhib Int Soc Magn Reson Med Sci Meet Exhib       Date:  2013

8.  Consistent projection reconstruction (CPR) techniques for MRI.

Authors:  G H Glover; D C Noll
Journal:  Magn Reson Med       Date:  1993-03       Impact factor: 4.668

9.  Parallel imaging with nonlinear reconstruction using variational penalties.

Authors:  Florian Knoll; Christian Clason; Kristian Bredies; Martin Uecker; Rudolf Stollberger
Journal:  Magn Reson Med       Date:  2011-06-27       Impact factor: 4.668

10.  Prospective motion correction using inductively coupled wireless RF coils.

Authors:  Melvyn B Ooi; Murat Aksoy; Julian Maclaren; Ronald D Watkins; Roland Bammer
Journal:  Magn Reson Med       Date:  2013-06-27       Impact factor: 4.668

View more
  4 in total

1.  High-resolution intravascular MRI-guided perivascular ultrasound ablation.

Authors:  Xiaoyang Liu; Nicholas Ellens; Emery Williams; Everette C Burdette; Parag Karmarkar; Clifford R Weiss; Dara Kraitchman; Paul A Bottomley
Journal:  Magn Reson Med       Date:  2019-08-11       Impact factor: 4.668

2.  Radiofrequency Ablation, MR Thermometry, and High-Spatial-Resolution MR Parametric Imaging with a Single, Minimally Invasive Device.

Authors:  M Arcan Ertürk; Shashank Sathyanarayana Hegde; Paul A Bottomley
Journal:  Radiology       Date:  2016-05-26       Impact factor: 11.105

3.  A Combined Intravascular MRI Endoscope and Intravascular Ultrasound (IVUS) Transducer for High-Resolution Image-Guided Ablation.

Authors:  Xiaoyang Liu; Nicholas Ellens; Emery Williams; Everette Clif Burdette; Parag Karmarkar; Paul Bottomley
Journal:  Proc Int Soc Magn Reson Med Sci Meet Exhib Int Soc Magn Reson Med Sci Meet Exhib       Date:  2017-04

4.  High-resolution and accelerated multi-parametric mapping with automated characterization of vessel disease using intravascular MRI.

Authors:  Guan Wang; Yi Zhang; Shashank Sathyanarayana Hegde; Paul A Bottomley
Journal:  J Cardiovasc Magn Reson       Date:  2017-11-20       Impact factor: 5.364

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.