Model-based CT performance assessment and optimization for iodinated and noniodinated imaging tasks as a function of kVp and body size.

Ehsan Samei1, Samuel Richard1, Lynne Lurwitz1.   

Abstract

PURPOSE: The goal of this study was to assess the comparative performance of iterative reconstruction in space (IRIS) and filtered back projection (FBP) reconstruction algorithms in terms of image quality and dose across kVps and phantom sizes.
METHODS: The ACR CT phantom (model 464) was supplemented with the addition of an iodinated spherical capsule (1.5 mm diameter, 3.4 mg iodine per ml) to simulate the contrast filled structures and with an additional circular attachment consisting of an array of 500 um brass beads for spatial resolution measurements. A larger sized phantom was also created by wrapping the original phantom with additional tissue equivalent material of 4 cm thickness. The phantoms were imaged on a 64 detector array multidetector computed tomography scanner (Somatom Definition, Siemens, Germany) using clinically applicable protocols (0.5 s rotation time; 80, 100, 120, and 140 kVp; 64 to 640 mA; 220 to 250 mm field of view). Images were reconstructed using the FBP and the IRIS algorithms. Combining measurements of image noise and spatial resolution with a task function, a figure of merit (FOM) for image quality was generated taking into account the type of visualization required from the image for the detection of either large or small image features with and without iodine content. The FOM was further reported in terms of area under the receiver operating characteristic (ROC) curve (AZ) to predict the comparative diagnostic performance of the two algorithms at different dose levels.
RESULTS: For a given dose level, the predicted AZ for IRIS consistently outperformed that of FBP. At comparative AZ, depending on protocol and task, the dose requirement for the optimal technique (optimized kVp with IRIS) was 2-3 times lower than that for standard technique (120 kVp with FBP). The potential for dose reduction was found to be higher when performing small feature detection tasks in comparison to larger feature detection tasks. The optimal kVp was from 80 to 100 kVp for the small phantom, 100 to 120 kVp for the larger phantom.
CONCLUSIONS: Overall, greater dose reduction may be achieved with IRIS compared to FBP, with enhanced advantage at thinner slice reconstructions. The results highlight how IRIS may offer a superior balance between image quality and dose across a range of imaging tasks, thus enabling dose reduction at constant quality or image quality improvement at constant dose. The prediction of the investigation can be used toward effective design of subsequent clinical studies.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25086541      PMCID: PMC4111837          DOI: 10.1118/1.4890082

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Detection of subtle lung nodules: relative influence of quantum and anatomic noise on chest radiographs.

Authors:  E Samei; M J Flynn; W R Eyler
Journal:  Radiology       Date:  1999-12       Impact factor: 11.105

2.  Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: the ERASIR study.

Authors:  Jonathon Leipsic; Troy M Labounty; Brett Heilbron; James K Min; G B John Mancini; Fay Y Lin; Carolyn Taylor; Allison Dunning; James P Earls
Journal:  AJR Am J Roentgenol       Date:  2010-09       Impact factor: 3.959

3.  Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images.

Authors:  Samuel Richard; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

4.  Objective assessment of image quality: effects of quantum noise and object variability.

Authors:  H H Barrett
Journal:  J Opt Soc Am A       Date:  1990-07       Impact factor: 2.129

5.  Quantitative comparison of noise texture across CT scanners from different manufacturers.

Authors:  Justin B Solomon; Olav Christianson; Ehsan Samei
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

6.  Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms.

Authors:  Samuel Richard; Daniela B Husarik; Girijesh Yadava; Simon N Murphy; Ehsan Samei
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

7.  Assessment of volumetric noise and resolution performance for linear and nonlinear CT reconstruction methods.

Authors:  Baiyu Chen; Olav Christianson; Joshua M Wilson; Ehsan Samei
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

8.  Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience.

Authors:  Márcio Sommer Bittencourt; Bernhard Schmidt; Martin Seltmann; Gerd Muschiol; Dieter Ropers; Werner Günther Daniel; Stephan Achenbach
Journal:  Int J Cardiovasc Imaging       Date:  2010-12-01       Impact factor: 2.357

9.  Patient-specific radiation dose and cancer risk for pediatric chest CT.

Authors:  Xiang Li; Ehsan Samei; W Paul Segars; Gregory M Sturgeon; James G Colsher; Donald P Frush
Journal:  Radiology       Date:  2011-04-05       Impact factor: 11.105

10.  Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience.

Authors:  Daniele Marin; Rendon C Nelson; Sebastian T Schindera; Samuel Richard; Richard S Youngblood; Terry T Yoshizumi; Ehsan Samei
Journal:  Radiology       Date:  2010-01       Impact factor: 11.105

View more
  5 in total

1.  Task-driven image acquisition and reconstruction in cone-beam CT.

Authors:  Grace J Gang; J Webster Stayman; Tina Ehtiati; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-03-24       Impact factor: 3.609

2.  Statistical model based iterative reconstruction in clinical CT systems. Part III. Task-based kV/mAs optimization for radiation dose reduction.

Authors:  Ke Li; Daniel Gomez-Cardona; Jiang Hsieh; Meghan G Lubner; Perry J Pickhardt; Guang-Hong Chen
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

3.  Can conclusions drawn from phantom-based image noise assessments be generalized to in vivo studies for the nonlinear model-based iterative reconstruction method?

Authors:  Daniel Gomez-Cardona; Ke Li; Jiang Hsieh; Meghan G Lubner; Perry J Pickhardt; Guang-Hong Chen
Journal:  Med Phys       Date:  2016-02       Impact factor: 4.071

4.  Comparison of radiation dose and image quality between flat panel computed tomography and multidetector computed tomography in a hybrid CT-angiography suite.

Authors:  Aaron K Jones; Bruno C Odisio
Journal:  J Appl Clin Med Phys       Date:  2020-01-10       Impact factor: 2.102

5.  A pitfall of using the circular-edge technique with image averaging for spatial resolution measurement in iteratively reconstructed CT images.

Authors:  Akihiro Narita; Masaki Ohkubo
Journal:  J Appl Clin Med Phys       Date:  2020-01-20       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.