| Literature DB >> 31922349 |
Aaron K Jones1, Bruno C Odisio2.
Abstract
The purpose of this study was to compare, using the same radiation dose and image quality metrics, flat panel computed tomography (FPCT) to multidetector CT (MDCT) in interventional radiology. A single robotic angiography system with FPCT was compared to a single MDCT system, both installed in a hybrid CT-angiography laboratory and both operating under automatic exposure control. Radiation dose was measured on the central axis (Dc ) of a CT dosimetry phantom 30 cm in diameter and 60 cm in length using default protocols for FPCT and MDCT with the imaged length in MDCT matched to the field of view of FPCT. The noise power spectrum (NPS), modulation transfer function (MTF), and z-axis resolution were measured using the same phantom. Iodine contrast to noise ratio (CNR) was also measured. Radiation dose (Dc ) was 41%-69% lower in MDCT compared to FPCT when default protocols and automatic exposure control were used. While spatial resolution could generally be matched with appropriate choice of kernel in MDCT, MTF dropped more quickly at higher spatial frequency for MDCT than FPCT. Image noise was 49%-120% higher for MDCT compared to FPCT for comparable in-plane spatial resolution. Z-axis resolution was slightly better for MDCT than FPCT, while iodine CNR depended on protocol selection. Radiation dose was much lower for MDCT compared to FPCT, but image noise was much higher. Matching image noise in MDCT to FPCT would result in similar radiation doses. Iodine contrast depended on dose modulation settings for MDCT.Entities:
Keywords: cone beam CT; flat panel CT; multidetector CT
Mesh:
Substances:
Year: 2020 PMID: 31922349 PMCID: PMC7020994 DOI: 10.1002/acm2.12808
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Acquisition parameters for FPCT.
| Organ program | kV plateau | Pulse width (ms) | Min. filtration (mm Cu) | Max. filtration (mm Cu) | Focus | Dose per frame (µGy) | Extended pixel depth | Resolution | Bit depth | 3D angle (°) | 3D step (°/frame) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 5sDCT CARE Body | 90 | 5 | 0.0 | 0.0 | Large | 0.36 | On | Low | 16 | 200 | 0.8 |
| 6sDCT Body | 90 | 5 | 0.0 | 0.0 | Large | 0.36 | On | Low | 16 | 200 | 0.5 |
FPCT, flat panel computed tomography.
Low resolution means detector pixels were binned 4 × 4 during acquisition.
Acquisition parameters for MDCT.
| Protocol | kV | Rot. time (s) | Pitch | Tube current (mA) | Detector configuration | Adaptive dose shield |
|---|---|---|---|---|---|---|
| Abdomen Routine | CarekV (ref. kV = 120, Slider Position = 7 or 9). | 0.5 | 0.6 | CareDose 4D, varied | 128 × 0.6 mm | On |
MDCT, multidetector CT.
Reconstruction parameters for image quality characterization.
| Technology | Organ program | Matrix size | Voxel size (mm) for MTF | Voxel size (mm) for NPS | Reconstruction algorithm/kernel |
|---|---|---|---|---|---|
| FPCT | 5sDCT Body CARE | 512 × 512 | 0.1483 | 0.4763 | HU/Normal |
| FPCT | 6sDCT Body | 512 × 512 | 0.1483 | 0.4763 | HU/Normal |
| MDCT | Abdomen Routine | 512 × 512 | 0.1482 × 0.6 mm | 0.4692 × 0.6 mm | B30f, B45f, B70f |
FPCT, flat panel computed tomography; MDCT, multidetector CT; MTF, modulation transfer function.
Pixel size x image thickness
Measured 0.5 and 0.1 MTF values for FPCT and MDCT.
| 5sDCT Body CARE (cm‐1) |
6sDCT Body (cm−1) | Abdomen Routine B30f (cm−1) | Abdomen Routine B45f (cm−1) | Abdomen Routine B70f (cm−1) | |
|---|---|---|---|---|---|
| 0.5 MTF | 4.68 | 4.75 | 3.54 | 4.92 | 8.01 |
| 0.1 MTF | 8.91 | 9.00 | 5.80 | 7.22 | 14.9 |
FPCT, flat panel computed tomography; MDCT, multidetector CT; MTF, modulation transfer function.
Comparison of iodine contrast and CNR, listed as contrast (HU)/ CNR.
| 5sDCT Body CARE | 6sDCT Body | Abdomen Routine (Slider Position 7) | Abdomen Routine (Slider Position 9) | |
|---|---|---|---|---|
| 4 mg/mL blood/iodine | 212.9/7.35 | 207.7/9.62 | 210.0/3.37 | 167.4/2.91 |
| 2 mg/mL blood/iodine (central) | 113.4/3.01 | 118.8/3.84 | 129.2/1.67 | 105.5/1.56 |
| 2 mg/mL blood/iodine (peripheral) | 125.8/3.67 | 123.3/4.93 | 132.6/1.93 | 107.8/1.70 |
MDCT, multidetector CT.
Comparison of radiation dose and image quality between flat panel computed tomography and multidetector computed tomography in a hybrid CT‐angiography suite.
Slider Position 7 resulted in selection of 80 kV by the MDCT, Slider Position 9 resulted in selection of 100 kV by the MDCT.
Designed to mimic blood/iodine mixture at the specified iodine concentration.
Figure 1Measured MTF for FPCT and MDCT. MDCT, multidetector CT; MTF, modulation transfer function.
Figure 2a) Radially averaged NPS; b) image of uniform area of ACR phantom acquired using 6sDCT FPCT protocol; same image acquired using MDCT c) Slider Position 7: 80 kV, B45 kernel; d) Slider Position 7: 80 kV, B30 kernel; e) Slider Position 7: 80 kV, B70 kernel. Note the differences in noise texture among FPCT and MDCT using different reconstruction kernels, and compare to the shapes of the NPS in part a). MDCT, multidetector CT.