Accuracy of segmented measurement of axial length in ultra-high myopia filled with silicone oil using immersion B-scan ultrasonography.

Qing-Hua Yang1, Hong-Tao Zhang1, Xiao-Qi Li1, Bing Chen1, Zhao-Hui Li1, Yi-Fei Huang1, Xin Jin1, Ying Zhang1, Li-Qiang Wang1.   

Abstract

AIM: To evaluate the accuracy of segmented measurement of axial length (AL) in high myopia filled with silicone oil by immersion B-scan ultrasonography (immersion B-scan).
METHODS: From June 2016 to June 2020, a total of 67 ultra-high myopia inpatients (67 eyes) who underwent silicone oil removal combined with cataract extraction and intraocular lens (IOL) implantation were retrospectively enrolled. The preoperative axial length (AL) of 31 patients with severe cataract were segmented measured using immersion B-scan (B-scan group) and another 36 patients with mild or moderate cataract were measured using IOLMaster 500 (IOLMaster group). The post-operative ALs in two groups were both measured using IOLMaster 500. The IOL power was calculated with Haigis formula. The differences in ALs between pre- and post-surgery, as well as the postoperative refractive spherical equivalent, absolute refractive error, the prediction deviation of postoperative refraction and best corrected visual acuity (BCVA) were compared.
RESULTS: The pre- and post-operative ALs were 30.46±1.63 mm (range 28.09-33.51 mm) and 30.42±1.70 mm (range 28.03-33.90 mm) in B-scan group (t=0.644, P=0.542) and 30.51±1.21 mm (range 28.03-33.90 mm) and 30.43±1.27mm (range 28.54-33.50 mm) in IOLMaster group (t=1.843, P=0.074), respectively. Three months after surgery, BCVA were 0.45±0.13 (range 0.3-0.9) and 0.44±0.20 (range 0.2-1.0) in B-scan and IOLMaster group respectively (t=0.086, P=0.932). There was no significant difference of the postoperative spherical equivalent (-3.11±0.65 D vs -3.21±0.51 D, t=0.671, P=0.505) and the absolute refractive error (0.589±0.340 vs 0.470±0.245 D, t=1.615, P=0.112) between two groups. In B-scan group, absolute refractive error within ±0.50 D was found in 18 eyes (58.1%), within ±1.00 D in 26 eyes (83.9%), and within ±1.50 D in 31 eyes (100%). In IOLMaster group, absolute refractive error within ±0.50 D was found in 23 eyes (63.9%), within ±1.00 D in 34 eyes (94.4%), and within ±1.50 D in 36 eyes (Z=0.757, P=0.449).
CONCLUSION: The segmented measurement of ALs by immersion B-scan shows comparable measurement accuracy with that of IOLMaster 500 in ultra-high myopia patients with severe cataract secondary to silicone oil filling and can obtain an ideal postoperative refractive state. International Journal of Ophthalmology Press.

Entities:  

Keywords:  IOLMaster; axial length; cataract; high myopia; immersion B-ultrasound; silicone oil

Year:  2022        PMID: 35601170      PMCID: PMC9091887          DOI: 10.18240/ijo.2022.05.16

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.645


  40 in total

Review 1.  IOL Power Calculation in Short and Long Eyes.

Authors:  Kenneth J Hoffer; Giacomo Savini
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2017 Jul-Aug

2.  Refractive outcomes after silicone oil removal and intraocular lens implantation.

Authors:  Kevin K Suk; William E Smiddy; Wei Shi
Journal:  Retina       Date:  2013-03       Impact factor: 4.256

3.  Accuracy of user-adjusted axial length measurements with optical biometry in eyes having combined phacovitrectomy for macular-off rhegmatogenous retinal detachment.

Authors:  Rubina Rahman; Simon Kolb; Chun Xiao Bong; John Stephenson
Journal:  J Cataract Refract Surg       Date:  2016-07       Impact factor: 3.351

Review 4.  Advances in the treatment of rhegmatogenous retinal detachment.

Authors:  Li Liao; Xiao-Hua Zhu
Journal:  Int J Ophthalmol       Date:  2019-04-18       Impact factor: 1.779

Review 5.  Global Prevalence of Myopia and High Myopia and Temporal Trends from 2000 through 2050.

Authors:  Brien A Holden; Timothy R Fricke; David A Wilson; Monica Jong; Kovin S Naidoo; Padmaja Sankaridurg; Tien Y Wong; Thomas J Naduvilath; Serge Resnikoff
Journal:  Ophthalmology       Date:  2016-02-11       Impact factor: 12.079

6.  The Correlation of Pars Plana Incision and Transient Hypotony After Silicone Oil Removal.

Authors:  Xian Zhang; Bo Chen; Huajing Yang; Yinwei Song; Di Zhang; Brian T Soetikno; Xufang Sun
Journal:  Ophthalmic Surg Lasers Imaging Retina       Date:  2018-09-01       Impact factor: 1.300

7.  Need for Retinal Detachment Reoperation Based on Primary Repair Method Among Commercially Insured Patients, 2003-2016.

Authors:  Mary-Grace R Reeves; Armin R Afshar; Suzann Pershing
Journal:  Am J Ophthalmol       Date:  2021-04-22       Impact factor: 5.488

8.  Comparison of Ocular Biometry Using New Swept-source Optical Coherence Tomography-based Optical Biometer with Other Devices.

Authors:  Young Joo Cho; Tae Hyung Lim; Kee Yong Choi; Beom Jin Cho
Journal:  Korean J Ophthalmol       Date:  2018-08

9.  Postoperative Refractive Prediction Error Measured by Optical and Acoustic Biometry after Phacovitrectomy for Rhegmatogenous Retinal Detachment without Macular Involvement.

Authors:  Masashi Sakamoto; Izumi Yoshida; Takahiro Sodeno; Asao Sakai; Hidetaka Masahara; Takatoshi Maeno
Journal:  J Ophthalmol       Date:  2019-05-02       Impact factor: 1.909

10.  Accuracy of Intraocular Lens Power Calculation Formulas for Highly Myopic Eyes.

Authors:  Yichi Zhang; Xiao Ying Liang; Shu Liu; Jacky W Y Lee; Srinivasan Bhaskar; Dennis S C Lam
Journal:  J Ophthalmol       Date:  2016-03-29       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.